Accuracy validation of the Microlife 3AS1-2 blood pressure device in a pregnant population with low blood pressure. Issue 5 (October 2015)
- Record Type:
- Journal Article
- Title:
- Accuracy validation of the Microlife 3AS1-2 blood pressure device in a pregnant population with low blood pressure. Issue 5 (October 2015)
- Main Title:
- Accuracy validation of the Microlife 3AS1-2 blood pressure device in a pregnant population with low blood pressure
- Authors:
- Nathan, Hannah L.
de Greeff, Annemarie
Hezelgrave, Natasha L.
Chappell, Lucy C.
Shennan, Andrew H. - Abstract:
- Abstract : Objective: To assess the accuracy of the Microlife 3AS1-2 blood pressure (BP) device in pregnant women with low BP to investigate suitability for hypotensive detection in low-income and middle-income countries. Methods: A prospective observational study was carried out evaluating the Microlife 3AS1-2, a hand-held, upper-arm, semiautomated BP device, according to British Hypertension Society (BHS) protocol methods. Thirty (stable) pregnant women with a clinical systolic BP less than 100 mmHg and/or diastolic BP less than 60 mmHg were recruited from antenatal wards and clinics and their BP was measured by three trained observers at a district-level hospital in South Africa. Accuracy was assessed according to the BHS grading criteria (A/B=pass) and the ANSI/AAMI/ISO standard for mean difference and SD (⩽5±8 mmHg). Results: The device achieved an A/A grade according to the BHS grading criteria. The mean difference±SD between the observer and the test device was 0.5±6.2 and 1.3±5.4 mmHg for systolic and diastolic BP, respectively, fulfilling the standard required by the ANSI/AAMI/ISO protocol. All observer differences were within 4 mmHg. Conclusion: According to the BHS protocol, the Microlife 3AS1-2 BP device is accurate in pregnant women with low BP. The device has been validated previously in pregnancy and pre-eclampsia and also fulfils the criteria of the WHO for use in a low-resource setting. Although unstable women were not included in this validation (for safetyAbstract : Objective: To assess the accuracy of the Microlife 3AS1-2 blood pressure (BP) device in pregnant women with low BP to investigate suitability for hypotensive detection in low-income and middle-income countries. Methods: A prospective observational study was carried out evaluating the Microlife 3AS1-2, a hand-held, upper-arm, semiautomated BP device, according to British Hypertension Society (BHS) protocol methods. Thirty (stable) pregnant women with a clinical systolic BP less than 100 mmHg and/or diastolic BP less than 60 mmHg were recruited from antenatal wards and clinics and their BP was measured by three trained observers at a district-level hospital in South Africa. Accuracy was assessed according to the BHS grading criteria (A/B=pass) and the ANSI/AAMI/ISO standard for mean difference and SD (⩽5±8 mmHg). Results: The device achieved an A/A grade according to the BHS grading criteria. The mean difference±SD between the observer and the test device was 0.5±6.2 and 1.3±5.4 mmHg for systolic and diastolic BP, respectively, fulfilling the standard required by the ANSI/AAMI/ISO protocol. All observer differences were within 4 mmHg. Conclusion: According to the BHS protocol, the Microlife 3AS1-2 BP device is accurate in pregnant women with low BP. The device has been validated previously in pregnancy and pre-eclampsia and also fulfils the criteria of the WHO for use in a low-resource setting. Although unstable women were not included in this validation (for safety and pragmatic reasons), this device could potentially improve the detection of shock secondary to obstetric haemorrhage or sepsis, as well as being used in pre-eclampsia, particularly in low-income and middle-income countries. Abstract : Supplemental Digital Content is available in the text. … (more)
- Is Part Of:
- Blood pressure monitoring. Volume 20:Issue 5(2015)
- Journal:
- Blood pressure monitoring
- Issue:
- Volume 20:Issue 5(2015)
- Issue Display:
- Volume 20, Issue 5 (2015)
- Year:
- 2015
- Volume:
- 20
- Issue:
- 5
- Issue Sort Value:
- 2015-0020-0005-0000
- Page Start:
- Page End:
- Publication Date:
- 2015-10
- Subjects:
- blood pressure measurement -- hypotension -- low-resource setting -- pregnancy -- validation
Blood pressure -- Measurement -- Periodicals
Blood Pressure Determination -- Periodicals
Blood Pressure Monitoring, Ambulatory -- Periodicals
Blood Pressure Monitors -- Periodicals
Hypertension -- Periodicals
612.140287 - Journal URLs:
- http://gateway.ovid.com/ovidweb.cgi?T=JS&MODE=ovid&PAGE=toc&D=ovft&AN=00126097-000000000-00000&NEWS=N ↗
http://ovidsp.ovid.com/ovidweb.cgi?T=JS&NEWS=n&CSC=Y&PAGE=toc&D=yrovft&AN=00126097-000000000-00000 ↗
http://www.bpmonitoring.com/ ↗
http://journals.lww.com/pages/default.aspx ↗
http://firstsearch.oclc.org ↗ - DOI:
- 10.1097/MBP.0000000000000134 ↗
- Languages:
- English
- ISSNs:
- 1359-5237
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 2113.035500
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 5074.xml