Risk of Implant Loosening After Cyclic Loading of Fusionless Growth Modulation Techniques: Nitinol Staples Versus Flexible Tether. Issue 7 (1st April 2017)
- Record Type:
- Journal Article
- Title:
- Risk of Implant Loosening After Cyclic Loading of Fusionless Growth Modulation Techniques: Nitinol Staples Versus Flexible Tether. Issue 7 (1st April 2017)
- Main Title:
- Risk of Implant Loosening After Cyclic Loading of Fusionless Growth Modulation Techniques
- Authors:
- Yaszay, Burt
Doan, Joshua D.
Parvaresh, Kevin C.
Farnsworth, Christine L. - Abstract:
- Abstract : Study Design: Biomechanical evaluation using porcine spines. Objective: Compare the fixation strength of two currently used fusionless adolescent idiopathic scoliosis correction techniques following cyclic loading using porcine spines. Summary of Background Data: The ability of fusionless implants to control or correct scoliosis in a growing patient requires such implants to maintain spinal fixation. Because they cross the disc, motion may weaken fixation over time. Methods: Eight pig spines were divided into cycled segments (T10-T13) and uncycled segments (T7-T8, L2-L3). Initial range of motion (ROM) was determined in torsion, flexion-extension, and lateral bending (0.5°/s to 1.75 N·m). Staple group (n = 4): 6 mm parallel staples were inserted on the right anterolateral spine across each intervertebral disc. Cycled segments received six staples (three adjacent discs) and uncycled segments received four staples (two separate discs). Tether group (n = 4): 5.35 × 35 mm right anterolateral vertebral body screws were placed into each vertebra. Cycled segments received four screws and uncycled segments received four screws. Screws in cycled segments were connected with a flexible tether tensioned to straight alignment. ROM of instrumented cycled segments was measured, and then segments were loaded to the measured ROM in flexion-extension (2000 cycles), lateral bending (1000 cycles), and axial rotation (2000 cycles). Implants were axially loaded to failure. ParametricAbstract : Study Design: Biomechanical evaluation using porcine spines. Objective: Compare the fixation strength of two currently used fusionless adolescent idiopathic scoliosis correction techniques following cyclic loading using porcine spines. Summary of Background Data: The ability of fusionless implants to control or correct scoliosis in a growing patient requires such implants to maintain spinal fixation. Because they cross the disc, motion may weaken fixation over time. Methods: Eight pig spines were divided into cycled segments (T10-T13) and uncycled segments (T7-T8, L2-L3). Initial range of motion (ROM) was determined in torsion, flexion-extension, and lateral bending (0.5°/s to 1.75 N·m). Staple group (n = 4): 6 mm parallel staples were inserted on the right anterolateral spine across each intervertebral disc. Cycled segments received six staples (three adjacent discs) and uncycled segments received four staples (two separate discs). Tether group (n = 4): 5.35 × 35 mm right anterolateral vertebral body screws were placed into each vertebra. Cycled segments received four screws and uncycled segments received four screws. Screws in cycled segments were connected with a flexible tether tensioned to straight alignment. ROM of instrumented cycled segments was measured, and then segments were loaded to the measured ROM in flexion-extension (2000 cycles), lateral bending (1000 cycles), and axial rotation (2000 cycles). Implants were axially loaded to failure. Parametric tests compared pre- to postimplant ROM; nonparametric tests compared staple to screw pullout strength; P < 0.05 was significant. Results: There were no differences in ROM before instrumentation between groups. ROM was not changed except tethers decreased left lateral bending (–6.2°). Although staple pullout was less than screw pullout for cycled and uncycled segments ( P < 0.05 and P = 0.057, respectively), there was no difference in pullout strength with and without cyclic loading for either group ( P = 0.4). Conclusion: Tethers decreased lateral bending away from the tether. Screws had almost five times greater load to failure than staples. Five thousand cycles did not result in loosening of either staple or tether screws. Level of Evidence: N/A … (more)
- Is Part Of:
- Spine. Volume 42:Issue 7(2017)
- Journal:
- Spine
- Issue:
- Volume 42:Issue 7(2017)
- Issue Display:
- Volume 42, Issue 7 (2017)
- Year:
- 2017
- Volume:
- 42
- Issue:
- 7
- Issue Sort Value:
- 2017-0042-0007-0000
- Page Start:
- Page End:
- Publication Date:
- 2017-04-01
- Subjects:
- anterolateral spinal implant cyclic loading -- anterolateral spinal staples -- anterolateral spinal tether -- anterolateral spinal tether pull-put -- anterolateral vertebral staple pull-out -- fusionless scoliosis correction -- fusionless spinal implant biomechanics -- SMA staples -- spinal growth modulation -- staple spine range of motion -- tether spine range of motion
Spine -- Abnormalities -- Periodicals
Spine -- Diseases -- Periodicals
Spine -- Surgery -- Periodicals
616.73005 - Journal URLs:
- http://gateway.ovid.com/ovidweb.cgi?T=JS&MODE=ovid&NEWS=n&PAGE=toc&D=ovft&AN=00007632-000000000-00000 ↗
http://journals.lww.com/spinejournal/pages/default.aspx ↗
http://www.spinejournal.com/ ↗
http://journals.lww.com ↗ - DOI:
- 10.1097/BRS.0000000000001809 ↗
- Languages:
- English
- ISSNs:
- 0362-2436
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 8413.903000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 5033.xml