Reproducibility of Anterior Segment Angle Metrics Measurements Derived From Cirrus Spectral Domain Optical Coherence Tomography. Issue 5 (June 2015)
- Record Type:
- Journal Article
- Title:
- Reproducibility of Anterior Segment Angle Metrics Measurements Derived From Cirrus Spectral Domain Optical Coherence Tomography. Issue 5 (June 2015)
- Main Title:
- Reproducibility of Anterior Segment Angle Metrics Measurements Derived From Cirrus Spectral Domain Optical Coherence Tomography
- Authors:
- Pan, Xiaojing
Marion, Ken
Maram, Jyotsna
Zhang, Zhou Yuan
Francis, Brian A.
Nittala, Muneeswar G.
Sadda, Srinivas R.
Chopra, Vikas - Abstract:
- Abstract : Purpose: To investigate the reproducibility of anterior segment angle (ACA) metrics measurements in normal subjects on Cirrus spectral domain optical coherence tomography (SD-OCT). Methods: 40 eyes from 20 healthy, normal subjects underwent anterior segment imaging using a Cirrus SD-OCT. For each eye, 2 acquisitions of 5-line raster scans were performed perpendicularly on the inferior (270 degrees) angle. The Schwalbe's line-angle opening distance (SL-AOD) and Schwalbe's line-trabecular-iris space area (SL-TISA) measurements were performed by masked certified reading center graders using customized grading software. Intra-acquisition, intergrader, and intragrader reproducibility of SL-AOD and SL-TISA measurements were evaluated by intraclass correlation coefficients (ICC), Bland-Altman plots, and computation of mean percent difference (MPD) and coefficient of variability (CV). Results: The mean SL-AOD (average of 2 acquisitions) was 0.75 mm (range, 0.32 mm to 1.39 mm); SL-TISA was 0.28 mm 2 (range, 0.082 mm 2 to 0.569 mm 2 ). The repeatability of Cirrus SD-OCT was excellent for both SL-AOD (MPD 4.74%, CV=0.92, ICC=0.99) and SL-TISA (MPD 9.4%, CV=0.8, ICC=0.99). The intragrader reproducibility was high for SL-AOD (MPD 4.28%, CV=0.94, ICC=0.995) and SL-TISA (MPD 6.05%, CV=0.89, ICC=0.993). The inter-grader reproducibility was not as good but still excellent for both SL-AOD (MPD 15.47%, CV=0.95, ICC=0.94) and SL-TISA (MPD 19.43%, CV=0.99, ICC=0.93). Bland-AltmanAbstract : Purpose: To investigate the reproducibility of anterior segment angle (ACA) metrics measurements in normal subjects on Cirrus spectral domain optical coherence tomography (SD-OCT). Methods: 40 eyes from 20 healthy, normal subjects underwent anterior segment imaging using a Cirrus SD-OCT. For each eye, 2 acquisitions of 5-line raster scans were performed perpendicularly on the inferior (270 degrees) angle. The Schwalbe's line-angle opening distance (SL-AOD) and Schwalbe's line-trabecular-iris space area (SL-TISA) measurements were performed by masked certified reading center graders using customized grading software. Intra-acquisition, intergrader, and intragrader reproducibility of SL-AOD and SL-TISA measurements were evaluated by intraclass correlation coefficients (ICC), Bland-Altman plots, and computation of mean percent difference (MPD) and coefficient of variability (CV). Results: The mean SL-AOD (average of 2 acquisitions) was 0.75 mm (range, 0.32 mm to 1.39 mm); SL-TISA was 0.28 mm 2 (range, 0.082 mm 2 to 0.569 mm 2 ). The repeatability of Cirrus SD-OCT was excellent for both SL-AOD (MPD 4.74%, CV=0.92, ICC=0.99) and SL-TISA (MPD 9.4%, CV=0.8, ICC=0.99). The intragrader reproducibility was high for SL-AOD (MPD 4.28%, CV=0.94, ICC=0.995) and SL-TISA (MPD 6.05%, CV=0.89, ICC=0.993). The inter-grader reproducibility was not as good but still excellent for both SL-AOD (MPD 15.47%, CV=0.95, ICC=0.94) and SL-TISA (MPD 19.43%, CV=0.99, ICC=0.93). Bland-Altman plots of all comparisons did not demonstrate any apparent bias, with similar repeatability at various SL-AOD and SL-TISA values. Conclusions: In our population of young healthy adults with normal eyes, there was excellent intra-acquisition, intragrader, and intergrader reproducibility for Schwalbe's line-based ACA metrics obtained from Cirrus SD-OCT images. These SD-OCT-derived measures may serve as reliable descriptors of angle morphometry for use in clinical trials and clinical practice. … (more)
- Is Part Of:
- Journal of glaucoma. Volume 24:Issue 5(2015:Jun./Jul.)
- Journal:
- Journal of glaucoma
- Issue:
- Volume 24:Issue 5(2015:Jun./Jul.)
- Issue Display:
- Volume 24, Issue 5 (2015)
- Year:
- 2015
- Volume:
- 24
- Issue:
- 5
- Issue Sort Value:
- 2015-0024-0005-0000
- Page Start:
- Page End:
- Publication Date:
- 2015-06
- Subjects:
- anterior chamber angle -- anterior-segment OCT -- reproducibility -- metrics -- glaucoma
Glaucoma -- Periodicals
617.741005 - Journal URLs:
- http://ovidsp.ovid.com/ovidweb.cgi?T=JS&NEWS=n&CSC=Y&PAGE=toc&D=yrovft&AN=00061198-000000000-00000 ↗
http://www.glaucomajournal.com ↗
http://journals.lww.com/glaucomajournal/pages/default.aspx ↗
http://journals.lww.com ↗ - DOI:
- 10.1097/IJG.0000000000000136 ↗
- Languages:
- English
- ISSNs:
- 1057-0829
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4996.230000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 4951.xml