Comparison of Intraocular Pressure Measurements With the Reichert Pt100, the Keeler Pulsair Intellipuff Portable Noncontact Tonometers, and Goldmann Applanation Tonometry. Issue 5 (June 2015)
- Record Type:
- Journal Article
- Title:
- Comparison of Intraocular Pressure Measurements With the Reichert Pt100, the Keeler Pulsair Intellipuff Portable Noncontact Tonometers, and Goldmann Applanation Tonometry. Issue 5 (June 2015)
- Main Title:
- Comparison of Intraocular Pressure Measurements With the Reichert Pt100, the Keeler Pulsair Intellipuff Portable Noncontact Tonometers, and Goldmann Applanation Tonometry
- Authors:
- Hubanova, Ralitsa
Aptel, Florent
Zhou, Thierry
Arnol, Nathalie
Romanet, Jean-Paul
Chiquet, Christophe - Abstract:
- Abstract : Purpose: The aim of this study was to compare intraocular pressure (IOP) measurements using 2 portable tonometers, the Keeler Pulsair Intellipuff and the Reichert PT100, with Goldmann applanation tonometry (GAT). Materials and Methods: Prospective cross-sectional study was conducted on normotensive and hypertensive patients recruited from the outpatient clinic of the University Hospital of Grenoble. IOP was measured using 2 portable noncontact tonometers (NCTs) and GAT in one eye of each patient in a random order. Central corneal thickness was measured with the Accutome PachPen Pachymeter. The Wilcoxon and t tests were used to compare the differences between tonometers, the Pearson and the Spearman correlation tests to evaluate the correlation among the methods, and the Bland-Altman plots to evaluate the agreement among the methods. Intraclass correlation coefficient (ICC) was used to evaluate the agreement among the methods in normotensive and hypertensive patients separately. Results: A total of 137 eyes of 137 patients were included, 104 normotensive and 33 hypertensive eyes. We found an excellent agreement between PT100 and GAT (ICC=0.77) and between Intellipuff and GAT (ICC=0.75) in normotensive patients. We found a fair to good agreement (ICC=0.67) between PT100 and GAT in hypertensive patients. The agreement we found between Intellipuff and GAT was also excellent in hypertensive patients (ICC=0.86). Both NCTs significantly overestimated IOP measurementsAbstract : Purpose: The aim of this study was to compare intraocular pressure (IOP) measurements using 2 portable tonometers, the Keeler Pulsair Intellipuff and the Reichert PT100, with Goldmann applanation tonometry (GAT). Materials and Methods: Prospective cross-sectional study was conducted on normotensive and hypertensive patients recruited from the outpatient clinic of the University Hospital of Grenoble. IOP was measured using 2 portable noncontact tonometers (NCTs) and GAT in one eye of each patient in a random order. Central corneal thickness was measured with the Accutome PachPen Pachymeter. The Wilcoxon and t tests were used to compare the differences between tonometers, the Pearson and the Spearman correlation tests to evaluate the correlation among the methods, and the Bland-Altman plots to evaluate the agreement among the methods. Intraclass correlation coefficient (ICC) was used to evaluate the agreement among the methods in normotensive and hypertensive patients separately. Results: A total of 137 eyes of 137 patients were included, 104 normotensive and 33 hypertensive eyes. We found an excellent agreement between PT100 and GAT (ICC=0.77) and between Intellipuff and GAT (ICC=0.75) in normotensive patients. We found a fair to good agreement (ICC=0.67) between PT100 and GAT in hypertensive patients. The agreement we found between Intellipuff and GAT was also excellent in hypertensive patients (ICC=0.86). Both NCTs significantly overestimated IOP measurements compared with GAT in normotensive and hypertensive eyes (mean of the differences between PT100 and GAT: 1.3±2.1 and 8.1±4.6 mm Hg in normotensive and in hypertensive patients, respectively ( P <0.05), and between Pulsair Intellipuff and GAT: 1.5±1.8 and 2.3±4.8 mm Hg in normotensive and in hypertensive patients, respectively ( P <0.05). Conclusions: The 2 NCTs agree well with GAT in normotensive patients. The Pulsair Intellipuff also agrees well with GAT in hypertensive patients. … (more)
- Is Part Of:
- Journal of glaucoma. Volume 24:Issue 5(2015:Jun./Jul.)
- Journal:
- Journal of glaucoma
- Issue:
- Volume 24:Issue 5(2015:Jun./Jul.)
- Issue Display:
- Volume 24, Issue 5 (2015)
- Year:
- 2015
- Volume:
- 24
- Issue:
- 5
- Issue Sort Value:
- 2015-0024-0005-0000
- Page Start:
- Page End:
- Publication Date:
- 2015-06
- Subjects:
- portable noncontact tonometer -- intraocular pressure -- screening -- glaucoma
Glaucoma -- Periodicals
617.741005 - Journal URLs:
- http://ovidsp.ovid.com/ovidweb.cgi?T=JS&NEWS=n&CSC=Y&PAGE=toc&D=yrovft&AN=00061198-000000000-00000 ↗
http://www.glaucomajournal.com ↗
http://journals.lww.com/glaucomajournal/pages/default.aspx ↗
http://journals.lww.com ↗ - DOI:
- 10.1097/01.ijg.0000435776.99193.41 ↗
- Languages:
- English
- ISSNs:
- 1057-0829
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4996.230000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 4950.xml