Development of an acoustic measurement protocol to monitor acetabular implant fixation in cementless total hip Arthroplasty: A preliminary study. (November 2017)
- Record Type:
- Journal Article
- Title:
- Development of an acoustic measurement protocol to monitor acetabular implant fixation in cementless total hip Arthroplasty: A preliminary study. (November 2017)
- Main Title:
- Development of an acoustic measurement protocol to monitor acetabular implant fixation in cementless total hip Arthroplasty: A preliminary study
- Authors:
- Goossens, Quentin
Leuridan, Steven
Henyš, Petr
Roosen, Jorg
Pastrav, Leonard
Mulier, Michiel
Desmet, Wim
Denis, Kathleen
Vander Sloten, Jos - Abstract:
- Highlights: A real-time method to monitor acetabular implant stability intra-operatively is presented. The method relies on a non-destructive acoustic measurement. This study is performed on artificial bone models and a cadaveric hemi pelvis. Abstract: In cementless total hip arthroplasty (THA), the initial stability is obtained by press-fitting the implant in the bone to allow osseointegration for a long term secondary stability. However, finding the insertion endpoint that corresponds to a proper initial stability is currently based on the tactile and auditory experiences of the orthopedic surgeon, which can be challenging. This study presents a novel real-time method based on acoustic signals to monitor the acetabular implant fixation in cementless total hip arthroplasty. Twelve acoustic in vitro experiments were performed on three types of bone models; a simple bone block model, an artificial pelvic model and a cadaveric model. A custom made beam was screwed onto the implant which functioned as a sound enhancer and insertor. At each insertion step an acoustic measurement was performed. A significant acoustic resonance frequency shift was observed during the insertion process for the different bone models; 250 Hz (35%, second bending mode) to 180 Hz (13%, fourth bending mode) for the artificial bone block models and 120 Hz (11%, eighth bending mode) for the artificial pelvis model. No significant frequency shift was observed during the cadaveric experiment due to a lackHighlights: A real-time method to monitor acetabular implant stability intra-operatively is presented. The method relies on a non-destructive acoustic measurement. This study is performed on artificial bone models and a cadaveric hemi pelvis. Abstract: In cementless total hip arthroplasty (THA), the initial stability is obtained by press-fitting the implant in the bone to allow osseointegration for a long term secondary stability. However, finding the insertion endpoint that corresponds to a proper initial stability is currently based on the tactile and auditory experiences of the orthopedic surgeon, which can be challenging. This study presents a novel real-time method based on acoustic signals to monitor the acetabular implant fixation in cementless total hip arthroplasty. Twelve acoustic in vitro experiments were performed on three types of bone models; a simple bone block model, an artificial pelvic model and a cadaveric model. A custom made beam was screwed onto the implant which functioned as a sound enhancer and insertor. At each insertion step an acoustic measurement was performed. A significant acoustic resonance frequency shift was observed during the insertion process for the different bone models; 250 Hz (35%, second bending mode) to 180 Hz (13%, fourth bending mode) for the artificial bone block models and 120 Hz (11%, eighth bending mode) for the artificial pelvis model. No significant frequency shift was observed during the cadaveric experiment due to a lack of implant fixation in this model. This novel diagnostic method shows the potential of using acoustic signals to monitor the implant seating during insertion. … (more)
- Is Part Of:
- Medical engineering & physics. Volume 49(2017)
- Journal:
- Medical engineering & physics
- Issue:
- Volume 49(2017)
- Issue Display:
- Volume 49, Issue 2017 (2017)
- Year:
- 2017
- Volume:
- 49
- Issue:
- 2017
- Issue Sort Value:
- 2017-0049-2017-0000
- Page Start:
- 28
- Page End:
- 38
- Publication Date:
- 2017-11
- Subjects:
- Total hip arthroplasty -- Vibro-acoustics -- Cementless prosthesis -- Acetabular cup -- Primary stability -- Resonance frequency
Biomedical engineering -- Periodicals
Biomedical Engineering -- Periodicals
Physics -- Periodicals
Génie biomédical -- Périodiques
Biomedical engineering
Electronic journals
Periodicals
610.28 - Journal URLs:
- http://www.medengphys.com ↗
http://www.sciencedirect.com/science/journal/13504533 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/13504533 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/13504533 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.medengphy.2017.07.006 ↗
- Languages:
- English
- ISSNs:
- 1350-4533
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5527.323000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File: