Aortic annulus eccentricity before and after transcatheter aortic valve implantation: Comparison of balloon-expandable and self-expanding prostheses. Issue 7 (July 2015)
- Record Type:
- Journal Article
- Title:
- Aortic annulus eccentricity before and after transcatheter aortic valve implantation: Comparison of balloon-expandable and self-expanding prostheses. Issue 7 (July 2015)
- Main Title:
- Aortic annulus eccentricity before and after transcatheter aortic valve implantation: Comparison of balloon-expandable and self-expanding prostheses
- Authors:
- Schuhbaeck, Annika
Weingartner, Christina
Arnold, Martin
Schmid, Jasmin
Pflederer, Tobias
Marwan, Mohamed
Rixe, Johannes
Nef, Holger
Schneider, Christian
Lell, Michael
Uder, Michael
Ensminger, Stephan
Feyrer, Richard
Weyand, Michael
Achenbach, Stephan - Abstract:
- Highlights: Post-implant geometry of catheter-based aortic valve prostheses is influenced by aortic valve calcification. Balloon-expandable prostheses are more circular as compared to self-expanding prostheses. The impact of post-implant geometry on valve function needs to be investigated. Abstract: Introduction: The geometry of the aortic annulus and implanted transcatheter aortic valve prosthesis might influence valve function. We investigated the influence of valve type and aortic valve calcification on post-implant geometry of catheter-based aortic valve prostheses. Methods: Eighty consecutive patients with severe aortic valve stenosis (mean age 82 ± 6 years) underwent computed tomography before and after TAVI. Aortic annulus diameters were determined. Influence of prosthesis type and degree of aortic valve calcification on post-implant eccentricity were analysed. Results: Aortic annulus eccentricity was reduced in patients after TAVI (0.21 ± 0.06 vs. 0.08 ± 0.06, p < 0.0001). Post-TAVI eccentricity was significantly lower in 65 patients following implantation of a balloon-expandable prosthesis as compared to 15 patients who received a self-expanding prosthesis (0.06 ± 0.05 vs. 0.15 ± 0.07, p < 0.0001), even though the extent of aortic valve calcification was not different. After TAVI, patients with a higher calcium amount retained a significantly higher eccentricity compared to patients with lower amounts of calcium. Conclusions: Patients undergoing TAVI with aHighlights: Post-implant geometry of catheter-based aortic valve prostheses is influenced by aortic valve calcification. Balloon-expandable prostheses are more circular as compared to self-expanding prostheses. The impact of post-implant geometry on valve function needs to be investigated. Abstract: Introduction: The geometry of the aortic annulus and implanted transcatheter aortic valve prosthesis might influence valve function. We investigated the influence of valve type and aortic valve calcification on post-implant geometry of catheter-based aortic valve prostheses. Methods: Eighty consecutive patients with severe aortic valve stenosis (mean age 82 ± 6 years) underwent computed tomography before and after TAVI. Aortic annulus diameters were determined. Influence of prosthesis type and degree of aortic valve calcification on post-implant eccentricity were analysed. Results: Aortic annulus eccentricity was reduced in patients after TAVI (0.21 ± 0.06 vs. 0.08 ± 0.06, p < 0.0001). Post-TAVI eccentricity was significantly lower in 65 patients following implantation of a balloon-expandable prosthesis as compared to 15 patients who received a self-expanding prosthesis (0.06 ± 0.05 vs. 0.15 ± 0.07, p < 0.0001), even though the extent of aortic valve calcification was not different. After TAVI, patients with a higher calcium amount retained a significantly higher eccentricity compared to patients with lower amounts of calcium. Conclusions: Patients undergoing TAVI with a balloon-expandable prosthesis show a more circular shape of the implanted prosthesis as compared to patients with a self-expanding prosthesis. Eccentricity of the deployed prosthesis is affected by the extent of aortic valve calcification. … (more)
- Is Part Of:
- European journal of radiology. Volume 84:Issue 7(2015)
- Journal:
- European journal of radiology
- Issue:
- Volume 84:Issue 7(2015)
- Issue Display:
- Volume 84, Issue 7 (2015)
- Year:
- 2015
- Volume:
- 84
- Issue:
- 7
- Issue Sort Value:
- 2015-0084-0007-0000
- Page Start:
- 1242
- Page End:
- 1248
- Publication Date:
- 2015-07
- Subjects:
- Aortic valve stenosis -- Heart valve prosthesis implantation -- Multidetector computed tomography -- Cardiac imaging techniques -- Three-dimensional imaging of the aortic root
Medical radiology -- Periodicals
Radiology -- Periodicals
Radiologie médicale -- Périodiques
Medical radiology
Periodicals
616.075705 - Journal URLs:
- http://www.sciencedirect.com/science/journal/0720048X ↗
http://www.elsevier.com/homepage/elecserv.htt ↗
http://www.clinicalkey.com/dura/browse/journalIssue/0720048X ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/0720048X ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.ejrad.2015.04.003 ↗
- Languages:
- English
- ISSNs:
- 0720-048X
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3829.738050
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 4876.xml