Comparison of two fiber‐optical temperature measurement systems in magnetic fields up to 9.4 Tesla. Issue 5 (19th June 2014)
- Record Type:
- Journal Article
- Title:
- Comparison of two fiber‐optical temperature measurement systems in magnetic fields up to 9.4 Tesla. Issue 5 (19th June 2014)
- Main Title:
- Comparison of two fiber‐optical temperature measurement systems in magnetic fields up to 9.4 Tesla
- Authors:
- Buchenberg, Waltraud B.
Dadakova, Tetiana
Groebner, Jens
Bock, Michael
Jung, Bernd - Abstract:
- Abstract : Purpose: Precise temperature measurements in the magnetic field are indispensable for MR safety studies and for temperature calibration during MR‐guided thermotherapy. In this work, the interference of two commonly used fiber‐optical temperature measurement systems with the static magnetic field B 0 was determined. Methods: Two fiber‐optical temperature measurement systems, a GaAs‐semiconductor and a phosphorescent phosphor ceramic, were compared for temperature measurements in B 0 . The probes and a glass thermometer for reference were placed in an MR‐compatible tube phantom within a water bath. Temperature measurements were carried out at three different MR systems covering static magnetic fields up to B 0 = 9.4T, and water temperatures were changed between 25°C and 65°C. Results: The GaAs‐probe significantly underestimated absolute temperatures by an amount related to the square of B 0 . A maximum difference of Δ T = −4.6°C was seen at 9.4T. No systematic temperature difference was found with the phosphor ceramic probe. For both systems, the measurements were not dependent on the orientation of the sensor to B 0 . Conclusion: Temperature measurements with the phosphor ceramic probe are immune to magnetic fields up to 9.4T, whereas the GaAs‐probes either require a recalibration inside the MR system or a correction based on the square of B 0 . Magn Reson Med 73:2047–2051, 2015. © 2014 Wiley Periodicals, Inc.
- Is Part Of:
- Magnetic resonance in medicine. Volume 73:Issue 5(2015:May)
- Journal:
- Magnetic resonance in medicine
- Issue:
- Volume 73:Issue 5(2015:May)
- Issue Display:
- Volume 73, Issue 5 (2015)
- Year:
- 2015
- Volume:
- 73
- Issue:
- 5
- Issue Sort Value:
- 2015-0073-0005-0000
- Page Start:
- 2047
- Page End:
- 2051
- Publication Date:
- 2014-06-19
- Subjects:
- fiber‐optical temperature measurement -- temperature measurements in high magnetic fields -- MR safety -- MR thermometry
Nuclear magnetic resonance -- Periodicals
Electron paramagnetic resonance -- Periodicals
616.07548 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1522-2594 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/mrm.25314 ↗
- Languages:
- English
- ISSNs:
- 0740-3194
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5337.798000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 4726.xml