Prospective use of Ablation Index targets improves clinical outcomes following ablation for atrial fibrillation. (26th July 2017)
- Record Type:
- Journal Article
- Title:
- Prospective use of Ablation Index targets improves clinical outcomes following ablation for atrial fibrillation. (26th July 2017)
- Main Title:
- Prospective use of Ablation Index targets improves clinical outcomes following ablation for atrial fibrillation
- Authors:
- Hussein, Ahmed
Das, Moloy
Chaturvedi, Vivek
Asfour, Issa Khalil
Daryanani, Niji
Morgan, Maureen
Ronayne, Christina
Shaw, Matthew
Snowdon, Richard
Gupta, Dhiraj - Abstract:
- Abstract: Aims: Late recovery of ablated tissue leading to reconnection of pulmonary veins remains common following radiofrequency catheter ablation for AF. Ablation Index (AI), a novel ablation quality marker, incorporates contact force (CF), time, and power in a weighted formula. We hypothesized that prospective use of our previously published derived AI targets would result in better outcomes when compared to CF‐guided ablation. Methods: Eighty‐nine consecutive drug‐refractory AF patients (49% paroxysmal) underwent AI‐guided ablation (AI‐group). AI targets were 550 for anterior/roof and 400 for posterior/inferior left atrial segments. Procedural and clinical outcomes of these patients were compared to 89 propensity‐matched controls who underwent CF‐guided ablation (CF‐group). All 178 procedures were otherwise similar, and both groups were followed‐up for 12 months. The last 25 patients from each group underwent analysis of all VisiTags™ for ablation duration, CF, Force‐Time Integral, and impedance drop. Results: First‐pass pulmonary vein isolation (PVI) was more frequent in AI‐group than in CF‐group (173 [97%] vs. 149 [84%] circles, P < 0.001), and acute PV reconnection was lower (11 [6%] vs. 24 [13%] circles, P = 0.02). Mean PVI ablation time was similar (AI‐group: 42 ± 9 vs. CF‐group: 45 ± 14 minutes, P = 0.14). Median impedance drop for AI‐group was significantly higher than in CF‐group (13.7 [9‐19] Ω vs. 8.8 [5.2‐13] Ω, P < 0.001). Two major complications occurred inAbstract: Aims: Late recovery of ablated tissue leading to reconnection of pulmonary veins remains common following radiofrequency catheter ablation for AF. Ablation Index (AI), a novel ablation quality marker, incorporates contact force (CF), time, and power in a weighted formula. We hypothesized that prospective use of our previously published derived AI targets would result in better outcomes when compared to CF‐guided ablation. Methods: Eighty‐nine consecutive drug‐refractory AF patients (49% paroxysmal) underwent AI‐guided ablation (AI‐group). AI targets were 550 for anterior/roof and 400 for posterior/inferior left atrial segments. Procedural and clinical outcomes of these patients were compared to 89 propensity‐matched controls who underwent CF‐guided ablation (CF‐group). All 178 procedures were otherwise similar, and both groups were followed‐up for 12 months. The last 25 patients from each group underwent analysis of all VisiTags™ for ablation duration, CF, Force‐Time Integral, and impedance drop. Results: First‐pass pulmonary vein isolation (PVI) was more frequent in AI‐group than in CF‐group (173 [97%] vs. 149 [84%] circles, P < 0.001), and acute PV reconnection was lower (11 [6%] vs. 24 [13%] circles, P = 0.02). Mean PVI ablation time was similar (AI‐group: 42 ± 9 vs. CF‐group: 45 ± 14 minutes, P = 0.14). Median impedance drop for AI‐group was significantly higher than in CF‐group (13.7 [9‐19] Ω vs. 8.8 [5.2‐13] Ω, P < 0.001). Two major complications occurred in CF‐group and none in AI‐group. Atrial tachyarrhythmia recurrence was significantly lower in AI‐group (15 of 89 [17%]) than in CF‐group (33 of 89 [37%], P = 0.002). Conclusion: AI‐guided ablation is associated with significant improvements in the incidence of acute PV reconnection and atrial tachyarrhythmia recurrence rate compared to CF‐guided ablation, potentially due to creation of better quality lesions as suggested by greater impedance drop. … (more)
- Is Part Of:
- Journal of cardiovascular electrophysiology. Volume 28:Number 9(2017)
- Journal:
- Journal of cardiovascular electrophysiology
- Issue:
- Volume 28:Number 9(2017)
- Issue Display:
- Volume 28, Issue 9 (2017)
- Year:
- 2017
- Volume:
- 28
- Issue:
- 9
- Issue Sort Value:
- 2017-0028-0009-0000
- Page Start:
- 1037
- Page End:
- 1047
- Publication Date:
- 2017-07-26
- Subjects:
- ablation index -- atrial fibrillation -- contact force -- impedance drop -- pulmonary vein isolation
Blood vessels -- Physiology -- Periodicals
Electrophysiology -- Periodicals
Heart -- Physiology -- Periodicals
612.1 - Journal URLs:
- http://onlinelibrary.wiley.com/ ↗
- DOI:
- 10.1111/jce.13281 ↗
- Languages:
- English
- ISSNs:
- 1045-3873
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4954.866000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 4686.xml