AirSPACE™ device for real-time positioning during laryngoscopy. (August 2017)
- Record Type:
- Journal Article
- Title:
- AirSPACE™ device for real-time positioning during laryngoscopy. (August 2017)
- Main Title:
- AirSPACE™ device for real-time positioning during laryngoscopy
- Authors:
- Burnett, Tyrone
Mancillas, Omar L.
Cai, Chunyan
Ghebremichael, Semhar
Gumbert, Sam D.
Vanga, Naveen
Hagberg, Carin A. - Abstract:
- Abstract: Background: Patient positioning is critical in establishing an optimal laryngeal view during various airway management techniques, including direct laryngoscopy. The "sniffing" position, in which all 3 axes are aligned, has been the traditional method for direct laryngoscopy. [1, 2] The AirSPACE™ is a new positioning device that was developed to facilitate patient positioning during laryngoscopy via mechanical head and neck manipulation, potentially improving the visualization of laryngeal structures. Aim: The purpose of this study was to evaluate the performance of the AirSPACE™ device, as measured by the percentage of the Modified Cormack-Lehane (C-L) classification system, Grade's III or IV specifically, for glottic visualization during the first attempt of laryngoscopy and intubation. Methods: Following IRB approval, 30 adult (≥18 y/o) patients with ASA status classifications I-III, requiring tracheal intubation and general anesthesia, participated in this study. In order to assess the performance of the device, various measurements were obtained, including the initial C-L airway grade view during the first attempt, the C-L grade view after repositioning (if necessary), the time required for patient repositioning, the time required to obtain an optimal view of the glottis and CO2 detection, the time required to fully set up the AirSPACE™ device onto the designated OR table, the number of attempts required for a successful intubation, the minimum oxygenAbstract: Background: Patient positioning is critical in establishing an optimal laryngeal view during various airway management techniques, including direct laryngoscopy. The "sniffing" position, in which all 3 axes are aligned, has been the traditional method for direct laryngoscopy. [1, 2] The AirSPACE™ is a new positioning device that was developed to facilitate patient positioning during laryngoscopy via mechanical head and neck manipulation, potentially improving the visualization of laryngeal structures. Aim: The purpose of this study was to evaluate the performance of the AirSPACE™ device, as measured by the percentage of the Modified Cormack-Lehane (C-L) classification system, Grade's III or IV specifically, for glottic visualization during the first attempt of laryngoscopy and intubation. Methods: Following IRB approval, 30 adult (≥18 y/o) patients with ASA status classifications I-III, requiring tracheal intubation and general anesthesia, participated in this study. In order to assess the performance of the device, various measurements were obtained, including the initial C-L airway grade view during the first attempt, the C-L grade view after repositioning (if necessary), the time required for patient repositioning, the time required to obtain an optimal view of the glottis and CO2 detection, the time required to fully set up the AirSPACE™ device onto the designated OR table, the number of attempts required for a successful intubation, the minimum oxygen saturation (SpO2 ) while the airway was being secured, the method of laryngoscopy performed, subjective assessment evaluations from utilizing the AirSPACE™, and safety reporting of any adverse events. Results: Of the 30 patients enrolled in the study, only 3 (10.0%, 95% confidence interval: 2.1%–26.5%) were recorded with a C-L Grade III or IV, with an average repositioning time of 29.3 ± 8.4s. The frequency of initial C-L grade views was: I, 50.0%; IIa, 16.7%; IIb, 23.3%; III, 6.7%; and IV, 3.3%. For the 3 patients who required repositioning, 2 demonstrated improvement in the glottic view after repositioning, while one demonstrated no improvement. The time required to obtain an optimal view of the glottis and that for CO2 detection was 19.7 ± 13.8s and 42.6 ± 22.0s, respectively. The average time required to set up the AirSPACE™ device was 4.29 ± 1.81 min. The average minimum oxygenation during intubation was 99.7± 0.6%. Direct laryngoscopy (90%) was the preferred method when performing laryngoscopy and intubation, when compared to indirect (10%) during the study period. Nearly half of the laryngoscopists assessed the difficulty of laryngoscopy and difficulty of ETT delivery as 1 ( Very Easy ), 43.3% and 46.7%, respectively, whereas 68.8% assessed AirSPACE™ device ease of use as 1 ( Very Easy ). Conclusion: This study demonstrated that the AirSPACE™ device is an effective positioning mechanism that provided the laryngoscopist with a highly favorable glottic view initially. Yet, further research is warranted in evaluating the AirSPACE™'s effectiveness in improving glottic visualization of less favorable and more difficult airways on a much larger spectrum. Highlights: Patient positioning is critical in establishing a patent airway. Proper head and neck positioning can optimize one's laryngeal view. The "sniffing" position has been the traditional method for direct laryngoscopy. The AirSPACE™ device facilitates patient positioning via mechanical capabilities. The AirSPACE™ device provides 'real-time' airway manipulation during laryngoscopy. … (more)
- Is Part Of:
- Trends in anaesthesia and critical care. Volume 15(2017)
- Journal:
- Trends in anaesthesia and critical care
- Issue:
- Volume 15(2017)
- Issue Display:
- Volume 15, Issue 2017 (2017)
- Year:
- 2017
- Volume:
- 15
- Issue:
- 2017
- Issue Sort Value:
- 2017-0015-2017-0000
- Page Start:
- 29
- Page End:
- 36
- Publication Date:
- 2017-08
- Subjects:
- Airway management -- Airway equipment -- Laryngoscopy -- Tracheal intubation -- Patient positioning -- Positioning device
Anesthesia -- Periodicals
Critical care medicine -- Periodicals
Anesthesia in traumatology -- Periodicals
Anesthesia -- Case studies -- Periodicals
617.9605 - Journal URLs:
- http://www.sciencedirect.com/ ↗
- DOI:
- 10.1016/j.tacc.2017.06.004 ↗
- Languages:
- English
- ISSNs:
- 2210-8440
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File: