A comparison of treatment plan quality between Tri-Co-60 intensity modulated radiation therapy and volumetric modulated arc therapy for cervical cancer. (August 2017)
- Record Type:
- Journal Article
- Title:
- A comparison of treatment plan quality between Tri-Co-60 intensity modulated radiation therapy and volumetric modulated arc therapy for cervical cancer. (August 2017)
- Main Title:
- A comparison of treatment plan quality between Tri-Co-60 intensity modulated radiation therapy and volumetric modulated arc therapy for cervical cancer
- Authors:
- Park, Jong Min
Park, So-Yeon
Kim, Jung-in
Kang, Hyun-Cheol
Choi, Chang Heon - Abstract:
- Highlights: Margin reduction capability of MR-IGRT is beneficial for cervical cancer treatment. MR-IGRT could reduce dose to OARs. MR-IGRT could reduce normal tissue irradiations. Abstract: Purpose: To investigate the plan quality of tri-Co-60 intensity-modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT) for cervical cancer. Methods: A total of 20 patients who received postoperative radiotherapy for cervical cancer were selected. For each patient, a tri-Co-60 IMRT plan for which the target volume was the planning target volume (PTV) generated by adding 1 mm isotropic margins from the clinical target volume (CTV) and a VMAT plan for which the target volume was the PTV generated by adding 7 mm and 10 mm margins from the CTV were generated. The tri-Co-60 IMRT plans were generated with the ViewRay™ system while the VMAT plans were generated with 15-MV photon beams from a linear accelerator (prescription dose = 50.4 Gy in 28 fractions). Results: The average volumes of the PTVs and CTVs were 704.9 cc ± 87.8 cc and 271.6 cc ± 51.6 cc, respectively. No noticeable differences in the dose-volumetric parameters for the target volumes were observed between the tri-Co-60 IMRT and VMAT plans. The values of V40Gy for the small bowel and rectal wall, V45Gy of the bladder, and V35Gy of the femoral heads for the VMAT plans were 14.6% ± 7.8%, 54.4% ± 4.2%, 30.0% ± 4.7%, and 8.9% ± 3.3%, respectively. Those of the tri-Co-60 IMRT plans were 2.8% ± 2.1%, 23.0% ± 8.9%,Highlights: Margin reduction capability of MR-IGRT is beneficial for cervical cancer treatment. MR-IGRT could reduce dose to OARs. MR-IGRT could reduce normal tissue irradiations. Abstract: Purpose: To investigate the plan quality of tri-Co-60 intensity-modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT) for cervical cancer. Methods: A total of 20 patients who received postoperative radiotherapy for cervical cancer were selected. For each patient, a tri-Co-60 IMRT plan for which the target volume was the planning target volume (PTV) generated by adding 1 mm isotropic margins from the clinical target volume (CTV) and a VMAT plan for which the target volume was the PTV generated by adding 7 mm and 10 mm margins from the CTV were generated. The tri-Co-60 IMRT plans were generated with the ViewRay™ system while the VMAT plans were generated with 15-MV photon beams from a linear accelerator (prescription dose = 50.4 Gy in 28 fractions). Results: The average volumes of the PTVs and CTVs were 704.9 cc ± 87.8 cc and 271.6 cc ± 51.6 cc, respectively. No noticeable differences in the dose-volumetric parameters for the target volumes were observed between the tri-Co-60 IMRT and VMAT plans. The values of V40Gy for the small bowel and rectal wall, V45Gy of the bladder, and V35Gy of the femoral heads for the VMAT plans were 14.6% ± 7.8%, 54.4% ± 4.2%, 30.0% ± 4.7%, and 8.9% ± 3.3%, respectively. Those of the tri-Co-60 IMRT plans were 2.8% ± 2.1%, 23.0% ± 8.9%, 17.1% ± 6.1%, and 0.3% ± 0.4%, respectively. Conclusions: Owing to the target margin reduction capability, the tri-Co-60 IMRT plans were more favorable than the VMAT plans for cervical cancer. … (more)
- Is Part Of:
- Physica medica. Volume 40(2017)
- Journal:
- Physica medica
- Issue:
- Volume 40(2017)
- Issue Display:
- Volume 40, Issue 2017 (2017)
- Year:
- 2017
- Volume:
- 40
- Issue:
- 2017
- Issue Sort Value:
- 2017-0040-2017-0000
- Page Start:
- 11
- Page End:
- 16
- Publication Date:
- 2017-08
- Subjects:
- Volumetric modulated arc therapy -- Intensity-modulated radiation therapy -- Cervical cancer -- MRI -- Image guided radiotherapy
Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2017.06.018 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 4643.xml