Evaluation of Mindray BC‐5000 hematology analyzer: a new miniature 5‐part WBC differential instrument. (5th May 2015)
- Record Type:
- Journal Article
- Title:
- Evaluation of Mindray BC‐5000 hematology analyzer: a new miniature 5‐part WBC differential instrument. (5th May 2015)
- Main Title:
- Evaluation of Mindray BC‐5000 hematology analyzer: a new miniature 5‐part WBC differential instrument
- Authors:
- Xiang, D.
Yue, J.
Lan, Y.
Sha, C.
Ren, S.
Li, Y.
Li, M.
Wang, C. - Abstract:
- Summary: Introduction: The Mindray BC‐5000 automated hematology analyzer is a miniature, automated hematology analyzer, and 5‐part leukocyte differential counter for in vitro diagnostic use in clinical laboratories. Methods: Precision, linearity, carryover, and method comparison were carried out. The analyzer was evaluated and compared with the Sysmex XE‐2100 hematology analyzer and manual microscopic in the hematology laboratory of a tertiary hospital in Chinese. Results: There were minimal carryover (<0.25%), and excellent linearity for white blood cell, and platelet counts ( r > 0.99). Within‐run precision was good at all levels for the routine cell blood count parameters (CV < 3.5%). Between‐run precision was acceptable at all levels for the analysis parameters (CV < 5%) except for eosinophil and basophil (CV% >10%). BC‐5000 displayed very good correlation ( r > 0.94) with the XE‐2100 for cell blood count and cell differential parameters except for basophil ( r = 0.72). The comparison of 258 leukocyte differential count results analyzed in parallel with manual microscopy, BC‐5000 analyzer showed perfect specificity (93%) and negative predictive value (90%) on abnormal blood samples. Conclusion: It is concluded that the overall performance of the BC‐5000 is acceptable. The miniature analyzer is suitable for use in small‐ to medium‐size laboratories.
- Is Part Of:
- International journal of laboratory hematology. Volume 37:Number 5(2015:Oct.)
- Journal:
- International journal of laboratory hematology
- Issue:
- Volume 37:Number 5(2015:Oct.)
- Issue Display:
- Volume 37, Issue 5 (2015)
- Year:
- 2015
- Volume:
- 37
- Issue:
- 5
- Issue Sort Value:
- 2015-0037-0005-0000
- Page Start:
- 597
- Page End:
- 605
- Publication Date:
- 2015-05-05
- Subjects:
- Automated hematology analyzer -- Mindray BC‐5000 -- performance evaluation -- leukocyte differential flag
Hematology -- Periodicals
Blood -- Diseases -- Periodicals
Hematology -- Periodicals
616.15005 - Journal URLs:
- http://firstsearch.oclc.org/FSIP?db=ECO&journal=1751-5521&screen=info&done=referer ↗
http://www.blackwell-synergy.com/loi/clh ↗
http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1751-553X ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1111/ijlh.12370 ↗
- Languages:
- English
- ISSNs:
- 1751-5521
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4542.312220
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 4576.xml