Initial assessment of a novel delivery system (NuDEL™®) for the covered Cheatham-Platinum stent. (15th June 2017)
- Record Type:
- Journal Article
- Title:
- Initial assessment of a novel delivery system (NuDEL™®) for the covered Cheatham-Platinum stent. (15th June 2017)
- Main Title:
- Initial assessment of a novel delivery system (NuDEL™®) for the covered Cheatham-Platinum stent
- Authors:
- Morgan, Gareth J.
Kenny, Damien
Duke, Christopher
Walsh, Kevin P.
Qureshi, Shakeel A. - Abstract:
- Abstract: Aims: We sought to evaluate the first-in-man use of a new system for implantation of covered stents in patients with complex structural and CHD. Methods and results: Retrospective data were collected of the first 13 NuDEL™ delivery systems used in patients. The NuDEL™ comprises a covered Cheatham-Platinum stent mounted on a balloon-in-balloon and pre-loaded in a long delivery sheath. Data were collected from three centres in the United Kingdom and Ireland. A total of 13 covered stents were delivered via 12 NuDEL™ delivery systems in 12 patients. Among them, six patients had coarctation of the aorta, five patients had right ventricular outflow tract stenosis, and one patient had severe stenosis of a Mustard systemic venous baffle. There were no complications, and all the stents were deployed in the desired position with satisfactory haemodynamic results. Conclusions: The development of a bespoke system of a pre-mounted, pre-loaded covered stent may negate some of the technical challenges that complicate large-calibre stent deployment. Our preliminary results suggest that the NuDEL™ system is a safe and effective means of covered stent deployment in challenging anatomy.
- Is Part Of:
- Cardiology in the young. Volume 27:Number 8(2017)
- Journal:
- Cardiology in the young
- Issue:
- Volume 27:Number 8(2017)
- Issue Display:
- Volume 27, Issue 8 (2017)
- Year:
- 2017
- Volume:
- 27
- Issue:
- 8
- Issue Sort Value:
- 2017-0027-0008-0000
- Page Start:
- 1465
- Page End:
- 1469
- Publication Date:
- 2017-06-15
- Subjects:
- Covered stent, -- coarctation, -- right ventricular outflow tract stent, -- percutaneous pulmonary valve
Pediatric cardiology -- Periodicals
618.9212 - Journal URLs:
- http://journals.cambridge.org/action/displayJournal?jid=CTY ↗
- DOI:
- 10.1017/S1047951117000488 ↗
- Languages:
- English
- ISSNs:
- 1047-9511
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library STI - ELD Digital Store
- Ingest File:
- 4484.xml