Noninvasive microwave ablation zone radii estimation using x‐ray CT image analysis. Issue 8 (14th July 2016)
- Record Type:
- Journal Article
- Title:
- Noninvasive microwave ablation zone radii estimation using x‐ray CT image analysis. Issue 8 (14th July 2016)
- Main Title:
- Noninvasive microwave ablation zone radii estimation using x‐ray CT image analysis
- Authors:
- Weiss, Noam
Goldberg, S. Nahum
Nissenbaum, Yitzhak
Sosna, Jacob
Azhari, Haim - Abstract:
- Abstract : Purpose: The aims of this study were to noninvasively and automatically estimate both the radius of the ablated liver tissue and the radius encircling the treated zone, which also defines where the tissue is definitely untreated during a microwave (MW) thermal ablation procedure. Methods: Fourteen ex vivo bovine fresh liver specimens were ablated at 40 W using a 14 G microwave antenna, for durations of 3, 6, 8, and 10 min. The tissues were scanned every 5 s during the ablation using an x‐ray CT scanner. In order to estimate the radius of the ablation zone, the acquired images were transformed into a polar presentation by displaying the Hounsfield units (HU) as a function of angle and radius. From this polar presentation, the average HU radial profile was analyzed at each time point and the ablation zone radius was estimated. In addition, textural analysis was applied to the original CT images. The proposed algorithm identified high entropy regions and estimated the treated zone radius per time. The estimated ablated zone radii as a function of treatment durations were compared, by means of correlation coefficient and root mean square error (RMSE) to gross pathology measurements taken immediately post‐treatment from similarly ablated tissue. Results: Both the estimated ablation radii and the treated zone radii demonstrated strong correlation with the measured gross pathology values ( R 2 ≥ 0.89 and R 2 ≥ 0.86, respectively). The automated ablation radii estimationAbstract : Purpose: The aims of this study were to noninvasively and automatically estimate both the radius of the ablated liver tissue and the radius encircling the treated zone, which also defines where the tissue is definitely untreated during a microwave (MW) thermal ablation procedure. Methods: Fourteen ex vivo bovine fresh liver specimens were ablated at 40 W using a 14 G microwave antenna, for durations of 3, 6, 8, and 10 min. The tissues were scanned every 5 s during the ablation using an x‐ray CT scanner. In order to estimate the radius of the ablation zone, the acquired images were transformed into a polar presentation by displaying the Hounsfield units (HU) as a function of angle and radius. From this polar presentation, the average HU radial profile was analyzed at each time point and the ablation zone radius was estimated. In addition, textural analysis was applied to the original CT images. The proposed algorithm identified high entropy regions and estimated the treated zone radius per time. The estimated ablated zone radii as a function of treatment durations were compared, by means of correlation coefficient and root mean square error (RMSE) to gross pathology measurements taken immediately post‐treatment from similarly ablated tissue. Results: Both the estimated ablation radii and the treated zone radii demonstrated strong correlation with the measured gross pathology values ( R 2 ≥ 0.89 and R 2 ≥ 0.86, respectively). The automated ablation radii estimation had an average discrepancy of less than 1 mm (RMSE = 0.65 mm) from the gross pathology measured values, while the treated zone radii showed a slight overestimation of approximately 1.5 mm (RMSE = 1.6 mm). Conclusions: Noninvasive monitoring of MW ablation using x‐ray CT and image analysis is feasible. Automatic estimations of the ablation zone radius and the radius encompassing the treated zone that highly correlate with actual ablation measured values can be obtained. This technique can therefore potentially be used to obtain real time monitoring and improve the clinical outcome. … (more)
- Is Part Of:
- Medical physics. Volume 43:Issue 8(2016)Part 1
- Journal:
- Medical physics
- Issue:
- Volume 43:Issue 8(2016)Part 1
- Issue Display:
- Volume 43, Issue 8, Part 1 (2016)
- Year:
- 2016
- Volume:
- 43
- Issue:
- 8
- Part:
- 1
- Issue Sort Value:
- 2016-0043-0008-0001
- Page Start:
- 4476
- Page End:
- 4482
- Publication Date:
- 2016-07-14
- Subjects:
- biological tissues -- computerised tomography -- diagnostic radiography -- entropy -- image texture -- liver -- mean square error methods -- medical image processing -- microwave antennas -- patient monitoring -- radiation therapy
Computed tomography -- Computed radiography -- Image analysis -- General statistical methods -- Therapeutic applications
Computerised tomographs -- Radiation therapy -- Biological material, e.g. blood, urine; Haemocytometers -- Digital computing or data processing equipment or methods, specially adapted for specific applications -- Image data processing or generation, in general -- Analysis of texture
microwave -- thermal ablation -- x‐ray CT -- thermal monitoring
Medical imaging -- Tissue ablation -- Computed tomography -- Antennas -- Entropy -- Liver -- Image analysis -- Microwaves -- Pathology -- Laser ablation
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1118/1.4954843 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 4453.xml