Comparative experiments on phantom and ex vivo liver tissue in microwave ablation. (21st January 2014)
- Record Type:
- Journal Article
- Title:
- Comparative experiments on phantom and ex vivo liver tissue in microwave ablation. (21st January 2014)
- Main Title:
- Comparative experiments on phantom and ex vivo liver tissue in microwave ablation
- Authors:
- Zhai, Fei
Nan, Qun
Ding, Jinli
Xu, Dehao
Zhang, Huijuan
Liu, Youjun
Bai, Fan - Abstract:
- Abstract: Purpose : The aim of this study is to investigate the thermal field distribution of phantom and ex vivo liver tissue in microwave ablation. We intent to verify if the phantom can be used in future studies in lieu of actual tissue. Methods : This experiment was divided into two groups of phantom and ex vivo porcine liver tissue. 2450 MHz is set. The tests last up to 240 s in 60 W. The velocity of the circulating water pumps were adjusted to 40 rounds/min. Twenty-five copper-constantan thermocouples (TCs) were inserted at the specified position to record temperature data. Result : For the cooling water, the temperature field was non-symmetric distribution at the gap before ( z > z < 0 mm) of two groups of experiments. At the part without cooling water ( z > 0 mm), effective ablation areas were larger; near the microwave antenna, the temperature curves showed good consistency for both materials. Far away from the microwave antenna, the value difference increased between phantom and liver tissue. Moreover, the effect of cooling water in phantom is more obvious than it in liver tissue. The shapes of ablation areas from two groups are not same. Conclusion : The result of the present work implied that heating patterns of liver tissue and phantom are comparable. But the difference of temperature field between two kinds of materials cannot be ignored. In cases of using phantom to verify temperature field in lieu of actual tissue, the researchers should pay fullAbstract: Purpose : The aim of this study is to investigate the thermal field distribution of phantom and ex vivo liver tissue in microwave ablation. We intent to verify if the phantom can be used in future studies in lieu of actual tissue. Methods : This experiment was divided into two groups of phantom and ex vivo porcine liver tissue. 2450 MHz is set. The tests last up to 240 s in 60 W. The velocity of the circulating water pumps were adjusted to 40 rounds/min. Twenty-five copper-constantan thermocouples (TCs) were inserted at the specified position to record temperature data. Result : For the cooling water, the temperature field was non-symmetric distribution at the gap before ( z > z < 0 mm) of two groups of experiments. At the part without cooling water ( z > 0 mm), effective ablation areas were larger; near the microwave antenna, the temperature curves showed good consistency for both materials. Far away from the microwave antenna, the value difference increased between phantom and liver tissue. Moreover, the effect of cooling water in phantom is more obvious than it in liver tissue. The shapes of ablation areas from two groups are not same. Conclusion : The result of the present work implied that heating patterns of liver tissue and phantom are comparable. But the difference of temperature field between two kinds of materials cannot be ignored. In cases of using phantom to verify temperature field in lieu of actual tissue, the researchers should pay full attention to these difference points. … (more)
- Is Part Of:
- Electromagnetic biology and medicine. Volume 34:Number 1(2015:Jan.)
- Journal:
- Electromagnetic biology and medicine
- Issue:
- Volume 34:Number 1(2015:Jan.)
- Issue Display:
- Volume 34, Issue 1 (2015)
- Year:
- 2015
- Volume:
- 34
- Issue:
- 1
- Issue Sort Value:
- 2015-0034-0001-0000
- Page Start:
- 29
- Page End:
- 36
- Publication Date:
- 2014-01-21
- Subjects:
- Microwave ablation -- phantom -- porcine liver tissue -- temperature field
Electrophysiology -- Periodicals
Electromagnetic Fields -- Periodicals
Electrophysiology -- Periodicals
Radiation, Nonionizing -- Periodicals
574 - Journal URLs:
- http://informahealthcare.com/loi/ebm ↗
http://informahealthcare.com ↗ - DOI:
- 10.3109/15368378.2013.868813 ↗
- Languages:
- English
- ISSNs:
- 1536-8378
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3699.478500
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 4452.xml