Measured acoustic intensities for clinical diagnostic ultrasound transducers and correlation with thermal index. (20th June 2017)
- Record Type:
- Journal Article
- Title:
- Measured acoustic intensities for clinical diagnostic ultrasound transducers and correlation with thermal index. (20th June 2017)
- Main Title:
- Measured acoustic intensities for clinical diagnostic ultrasound transducers and correlation with thermal index
- Authors:
- Retz, K.
Kotopoulis, S.
Kiserud, T.
Matre, K.
Eide, G. E.
Sande, R. - Abstract:
- ABSTRACT: Objectives: To investigate if the thermal index for bone (TIB) displayed on screen is an adequate predictor for the derated spatial‐peak temporal‐average (ISPTA .3 ) and spatial‐peak pulse‐average (ISPPA .3 ) acoustic intensities in a selection of clinical diagnostic ultrasound machines and transducers. Methods: We calibrated five clinical diagnostic ultrasound scanners and 10 transducers, using two‐dimensional grayscale, color Doppler and pulsed‐wave Doppler, both close to and far from the transducer, with a TIB between 0.1 and 4.0, recording 103 unique measurements. Acoustic measurements were performed in a bespoke three‐axis computer‐controlled scanning tank, using a 200‐ μ m‐diameter calibrated needle hydrophone. Results: There was significant but poor correlation between the acoustic intensities and the on‐screen TIB. At a TIB of 0.1, the ISPTA .3 range was 0.51–50.49 mW/cm 2 and the ISPPA .3 range was 0.01–207.29 W/cm 2 . At a TIB of 1.1, the ISPTA .3 range was 19.02–309.44 mW/cm 2 and the ISPPA .3 range was 3.87–51.89 W/cm 2 . Conclusions: TIB is a poor predictor for ISPTA .3 and ISPPA .3 and for the potential bioeffects of clinical diagnostic ultrasound scanners. Copyright © 2016 ISUOG. Published by John Wiley & Sons Ltd.
- Is Part Of:
- Ultrasound in obstetrics & gynecology. Volume 50:Number 2(2017)
- Journal:
- Ultrasound in obstetrics & gynecology
- Issue:
- Volume 50:Number 2(2017)
- Issue Display:
- Volume 50, Issue 2 (2017)
- Year:
- 2017
- Volume:
- 50
- Issue:
- 2
- Issue Sort Value:
- 2017-0050-0002-0000
- Page Start:
- 236
- Page End:
- 241
- Publication Date:
- 2017-06-20
- Subjects:
- intensity measurements -- in vitro -- thermal index -- ultrasound safety
Ultrasonics in obstetrics -- Periodicals
Generative organs, Female -- Diseases -- Diagnosis -- Periodicals
Diagnosis, Ultrasonic -- Periodicals
Genital Diseases, Female -- ultrasonography -- Periodicals
Ultrasonography, Prenatal -- Periodicals
618.047543 - Journal URLs:
- http://obgyn.onlinelibrary.wiley.com/hub/journal/10.1002/(ISSN)1469-0705/ ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/uog.17298 ↗
- Languages:
- English
- ISSNs:
- 0960-7692
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 9082.815300
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 4403.xml