Inappropriate shock in a subcutaneous cardiac defibrillator due to residual air. Issue 8 (6th June 2017)
- Record Type:
- Journal Article
- Title:
- Inappropriate shock in a subcutaneous cardiac defibrillator due to residual air. Issue 8 (6th June 2017)
- Main Title:
- Inappropriate shock in a subcutaneous cardiac defibrillator due to residual air
- Authors:
- Lee, Shawn
Souvaliotis, Nektarios
Mehta, Davendra
Suri, Ranjit - Abstract:
- Key Clinical Message: Inappropriate shock due to residual air in subcutaneous implantable cardiac defibrillators is not a well‐known complication. Obtaining overpenetrated X‐rays, recognizing electrocardiogram findings, limiting blunt finger dissection, and switching to sense at another vector are techniques which might lead to avoidance of unnecessary wound exploration or device removal. Abstract : Lateral chest film showing air surrounding subxiphoid node.
- Is Part Of:
- Clinical case reports. Volume 5:Issue 8(2017)
- Journal:
- Clinical case reports
- Issue:
- Volume 5:Issue 8(2017)
- Issue Display:
- Volume 5, Issue 8 (2017)
- Year:
- 2017
- Volume:
- 5
- Issue:
- 8
- Issue Sort Value:
- 2017-0005-0008-0000
- Page Start:
- 1203
- Page End:
- 1206
- Publication Date:
- 2017-06-06
- Subjects:
- Defibrillator -- ICD -- inappropriate shock -- subcutaneous air -- subcutaneous ICD
Medicine -- Periodicals
616.09 - Journal URLs:
- http://onlinelibrary.wiley.com/ ↗
http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)2050-0904 ↗ - DOI:
- 10.1002/ccr3.1009 ↗
- Languages:
- English