Towards ultrasound‐guided adaptive radiotherapy for cervical cancer: Evaluation of Elekta's semiautomated uterine segmentation method on 3D ultrasound images. Issue 7 (16th June 2017)
- Record Type:
- Journal Article
- Title:
- Towards ultrasound‐guided adaptive radiotherapy for cervical cancer: Evaluation of Elekta's semiautomated uterine segmentation method on 3D ultrasound images. Issue 7 (16th June 2017)
- Main Title:
- Towards ultrasound‐guided adaptive radiotherapy for cervical cancer: Evaluation of Elekta's semiautomated uterine segmentation method on 3D ultrasound images
- Authors:
- Mason, Sarah A.
O'Shea, Tuathan P.
White, Ingrid M.
Lalondrelle, Susan
Downey, Kate
Baker, Mariwan
Behrens, Claus F.
Bamber, Jeffrey C.
Harris, Emma J. - Abstract:
- Abstract : Purpose: 3D ultrasound (US) images of the uterus may be used to adapt radiotherapy (RT) for cervical cancer patients based on changes in daily anatomy. This requires accurate on‐line segmentation of the uterus. The aim of this work was to assess the accuracy of Elekta's "Assisted Gyne Segmentation" (AGS) algorithm in semi‐automatically segmenting the uterus on 3D transabdominal ultrasound images by comparison with manual contours. Materials & methods: Nine patients receiving RT for cervical cancer were imaged with the 3D Clarity ® transabdominal probe at RT planning, and 1 to 7 times during treatment. Image quality was rated from unusable (0)–excellent (3). Four experts segmented the uterus (defined as the uterine body and cervix) manually and using AGS on images with a ranking > 0. Pairwise analysis between manual contours was evaluated to determine interobserver variability. The accuracy of the AGS method was assessed by measuring its agreement with manual contours via pairwise analysis. Results: 35/44 images acquired (79.5%) received a ranking > 0. For the manual contour variation, the median [interquartile range (IQR)] distance between centroids (DC) was 5.41 [5.0] mm, the Dice similarity coefficient (DSC) was 0.78 [0.11], the mean surface‐to‐surface distance (MSSD) was 3.20 [1.8] mm, and the uniform margin of 95% (UM95) was 4.04 [5.8] mm. There was no correlation between image quality and manual contour agreement. AGS failed to give a result in 19.3% ofAbstract : Purpose: 3D ultrasound (US) images of the uterus may be used to adapt radiotherapy (RT) for cervical cancer patients based on changes in daily anatomy. This requires accurate on‐line segmentation of the uterus. The aim of this work was to assess the accuracy of Elekta's "Assisted Gyne Segmentation" (AGS) algorithm in semi‐automatically segmenting the uterus on 3D transabdominal ultrasound images by comparison with manual contours. Materials & methods: Nine patients receiving RT for cervical cancer were imaged with the 3D Clarity ® transabdominal probe at RT planning, and 1 to 7 times during treatment. Image quality was rated from unusable (0)–excellent (3). Four experts segmented the uterus (defined as the uterine body and cervix) manually and using AGS on images with a ranking > 0. Pairwise analysis between manual contours was evaluated to determine interobserver variability. The accuracy of the AGS method was assessed by measuring its agreement with manual contours via pairwise analysis. Results: 35/44 images acquired (79.5%) received a ranking > 0. For the manual contour variation, the median [interquartile range (IQR)] distance between centroids (DC) was 5.41 [5.0] mm, the Dice similarity coefficient (DSC) was 0.78 [0.11], the mean surface‐to‐surface distance (MSSD) was 3.20 [1.8] mm, and the uniform margin of 95% (UM95) was 4.04 [5.8] mm. There was no correlation between image quality and manual contour agreement. AGS failed to give a result in 19.3% of cases. For the remaining cases, the level of agreement between AGS contours and manual contours depended on image quality. There were no significant differences between the AGS segmentations and the manual segmentations on the images that received a quality rating of 3. However, the AGS algorithm had significantly worse agreement with manual contours on images with quality ratings of 1 and 2 compared with the corresponding interobserver manual variation. The overall median [IQR] DC, DSC, MSSD, and UM95 between AGS and manual contours was 5.48 [5.45] mm, 0.77 [0.14], 3.62 [2.7] mm, and 5.19 [8.1] mm, respectively. Conclusions: The AGS tool was able to represent uterine shape of cervical cancer patients in agreement with manual contouring in cases where the image quality was excellent, but not in cases where image quality was degraded by common artifacts such as shadowing and signal attenuation. The AGS tool should be used with caution for adaptive RT purposes, as it is not reliable in accurately segmenting the uterus on 'good' or 'poor' quality images. The interobserver agreement between manual contours of the uterus drawn on 3D US was consistent with results of similar studies performed on CT and MRI images. … (more)
- Is Part Of:
- Medical physics. Volume 44:Issue 7(2017)
- Journal:
- Medical physics
- Issue:
- Volume 44:Issue 7(2017)
- Issue Display:
- Volume 44, Issue 7 (2017)
- Year:
- 2017
- Volume:
- 44
- Issue:
- 7
- Issue Sort Value:
- 2017-0044-0007-0000
- Page Start:
- 3630
- Page End:
- 3638
- Publication Date:
- 2017-06-16
- Subjects:
- adaptive radiotherapy -- cervical cancer -- segmentation -- ultrasound -- uterus
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1002/mp.12325 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 2911.xml