Effects of Needle and Catheter Size on Commercially Available Ultrasound Contrast Agents. (18th September 2015)
- Record Type:
- Journal Article
- Title:
- Effects of Needle and Catheter Size on Commercially Available Ultrasound Contrast Agents. (18th September 2015)
- Main Title:
- Effects of Needle and Catheter Size on Commercially Available Ultrasound Contrast Agents
- Authors:
- Eisenbrey, John R.
Daecher, Annemarie
Kramer, Michael R.
Forsberg, Flemming - Abstract:
- Abstract : Objectives: To investigate effects of needle and catheter size on in vitro ultrasound contrast agent (UCA) enhancement and concentrations using 4 commercially available UCAs. Methods: Definity (Lantheus Medical Imaging, North Billerica, MA), Optison (GE Healthcare, Princeton, NJ), SonoVue (Bracco SA, Geneva, Switzerland), and Sonazoid (GE Healthcare, Oslo, Norway) were investigated. The UCA was injected via a 1‐mL syringe (BD, Franklin Lakes, NJ) into a 3‐way stopcock (Smith Medical, Dublin, OH) and flushed with 10 mL of saline through an 18‐cm infusion extension tube connected to either a 16‐, 18‐, 20‐, 22‐, or 24‐gauge catheter (BD) or an 18‐, 20‐, 21‐, or 25‐gauge needle (BD). In vitro enhancement was determined in a flow phantom (ATS Laboratories, Bridgeport, CT), and microbubble concentrations were determined using an LSRII flow cytometer (BD Biosciences, San Jose, CA). Results: Significant decreases in enhancement and microbubble concentrations were observed for all 4 UCAs ( P < .001) when administration was performed through a 25‐gauge needle. No statistically significant differences in enhancement or concentrations were observed between all catheter sizes and 18‐ to 21‐gauge needles for SonoVue and Sonazoid. Definity and Optison administration through a 24‐gauge catheter resulted in a significant loss of enhancement (P < .02), although these differences were not significant on flow cytometry. Conclusions: Administration of commercial UCAs in a clinicalAbstract : Objectives: To investigate effects of needle and catheter size on in vitro ultrasound contrast agent (UCA) enhancement and concentrations using 4 commercially available UCAs. Methods: Definity (Lantheus Medical Imaging, North Billerica, MA), Optison (GE Healthcare, Princeton, NJ), SonoVue (Bracco SA, Geneva, Switzerland), and Sonazoid (GE Healthcare, Oslo, Norway) were investigated. The UCA was injected via a 1‐mL syringe (BD, Franklin Lakes, NJ) into a 3‐way stopcock (Smith Medical, Dublin, OH) and flushed with 10 mL of saline through an 18‐cm infusion extension tube connected to either a 16‐, 18‐, 20‐, 22‐, or 24‐gauge catheter (BD) or an 18‐, 20‐, 21‐, or 25‐gauge needle (BD). In vitro enhancement was determined in a flow phantom (ATS Laboratories, Bridgeport, CT), and microbubble concentrations were determined using an LSRII flow cytometer (BD Biosciences, San Jose, CA). Results: Significant decreases in enhancement and microbubble concentrations were observed for all 4 UCAs ( P < .001) when administration was performed through a 25‐gauge needle. No statistically significant differences in enhancement or concentrations were observed between all catheter sizes and 18‐ to 21‐gauge needles for SonoVue and Sonazoid. Definity and Optison administration through a 24‐gauge catheter resulted in a significant loss of enhancement (P < .02), although these differences were not significant on flow cytometry. Conclusions: Administration of commercial UCAs in a clinical scenario is possible with catheters or needles smaller than 20 gauge, although the minimal allowable size appears to be UCA specific. … (more)
- Is Part Of:
- Journal of ultrasound in medicine. Volume 34:Number 11(2015)
- Journal:
- Journal of ultrasound in medicine
- Issue:
- Volume 34:Number 11(2015)
- Issue Display:
- Volume 34, Issue 11 (2015)
- Year:
- 2015
- Volume:
- 34
- Issue:
- 11
- Issue Sort Value:
- 2015-0034-0011-0000
- Page Start:
- 1961
- Page End:
- 1968
- Publication Date:
- 2015-09-18
- Subjects:
- catheter size -- Definity -- needle size -- Optison -- Sonazoid -- SonoVue
Ultrasonics in medicine -- Periodicals
Ultrasonics
Ultrasonography
Ultrasonics in medicine
Electronic journals
Periodicals
Periodicals
616.07543 - Journal URLs:
- http://www.jultrasoundmed.org/ ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.7863/ultra.14.11008 ↗
- Languages:
- English
- ISSNs:
- 0278-4297
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5071.455000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 2824.xml