Repeatability of estimated fetal weight: Comparison between MR imaging versus 2D ultrasound in at- and near-term patients. Issue 91 (June 2017)
- Record Type:
- Journal Article
- Title:
- Repeatability of estimated fetal weight: Comparison between MR imaging versus 2D ultrasound in at- and near-term patients. Issue 91 (June 2017)
- Main Title:
- Repeatability of estimated fetal weight: Comparison between MR imaging versus 2D ultrasound in at- and near-term patients
- Authors:
- Kadji, Caroline
Camus, Margaux F.
Bevilacqua, Elisa
Cannie, Mieke M.
Sanchez, Teresa Cos
Jani, Jacques C. - Abstract:
- Highlights: MR imaging for the estimation of fetal weight is more reliable than 2D-ultrasound. Operator experience has a marginal impact on the variability of US estimated weight. Operator experience has no impact on the variability of MR estimated weight. Abstract: Introduction: Our aim was to evaluate the intra- and inter-observer variability and the impact of operator experience on the estimation of fetal weight (EFW) as measured by 2-dimensional ultrasound (2D-US) and magnetic resonance (MR) imaging. Material and methods: We estimated fetal weight in 46 singleton pregnancies at 35.6–41.4 weeks gestation using 2D-US according to the Hadlock formula and using MR imaging according to the equation developed by Baker. Each examination was performed twice, once by an inexperienced operator and once by an experienced operator. The MR-EFW was derived from the planimetric measurement of fetal body volume (FBV) using an assisted semi-automated method. Intra- and inter-observer variability was evaluated by Bland-Altman analysis. Regression analysis was used to investigate the effect of maternal BMI, delivery weight, diabetes and fetal gender on the differences in US-EFW between the inexperienced and experienced operators. Results: US-EFW showed higher intra-observer variability than MR-EFW, irrespective of operator experience. The 95% limits of agreement of MR were narrower compared with those of the US measurements. Similarly, US-EFW showed higher inter-observer variability thanHighlights: MR imaging for the estimation of fetal weight is more reliable than 2D-ultrasound. Operator experience has a marginal impact on the variability of US estimated weight. Operator experience has no impact on the variability of MR estimated weight. Abstract: Introduction: Our aim was to evaluate the intra- and inter-observer variability and the impact of operator experience on the estimation of fetal weight (EFW) as measured by 2-dimensional ultrasound (2D-US) and magnetic resonance (MR) imaging. Material and methods: We estimated fetal weight in 46 singleton pregnancies at 35.6–41.4 weeks gestation using 2D-US according to the Hadlock formula and using MR imaging according to the equation developed by Baker. Each examination was performed twice, once by an inexperienced operator and once by an experienced operator. The MR-EFW was derived from the planimetric measurement of fetal body volume (FBV) using an assisted semi-automated method. Intra- and inter-observer variability was evaluated by Bland-Altman analysis. Regression analysis was used to investigate the effect of maternal BMI, delivery weight, diabetes and fetal gender on the differences in US-EFW between the inexperienced and experienced operators. Results: US-EFW showed higher intra-observer variability than MR-EFW, irrespective of operator experience. The 95% limits of agreement of MR were narrower compared with those of the US measurements. Similarly, US-EFW showed higher inter-observer variability than MR-EFW. MR-EFW improvement over 2D-US for the limits of agreement was 77.9% for intra-observer variability and 74.5% for inter-observer variability. Regression analysis showed that the differences between US-EFW measurements were not related to any of the tested variables. Conclusions: Operator experience has a marginal impact on the variability of US-EFW and no impact on MR-EFW variability. The variability in US-EFW measurements is unpredictable. … (more)
- Is Part Of:
- European journal of radiology. Issue 91(2017)
- Journal:
- European journal of radiology
- Issue:
- Issue 91(2017)
- Issue Display:
- Volume 91, Issue 91 (2017)
- Year:
- 2017
- Volume:
- 91
- Issue:
- 91
- Issue Sort Value:
- 2017-0091-0091-0000
- Page Start:
- 35
- Page End:
- 40
- Publication Date:
- 2017-06
- Subjects:
- EFW estimated fetal weight -- 2D-US two-dimensional ultrasound -- MR magnetic resonance -- FBV fetal body volume
Estimation of fetal weight -- Fetal body volume -- Magnetic resonance imaging -- 2D ultrasound -- Repeatability -- Operator experience
Medical radiology -- Periodicals
Radiology -- Periodicals
Radiologie médicale -- Périodiques
Medical radiology
Periodicals
616.075705 - Journal URLs:
- http://www.sciencedirect.com/science/journal/0720048X ↗
http://www.elsevier.com/homepage/elecserv.htt ↗
http://www.clinicalkey.com/dura/browse/journalIssue/0720048X ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/0720048X ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.ejrad.2017.03.005 ↗
- Languages:
- English
- ISSNs:
- 0720-048X
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3829.738050
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 2881.xml