Taggart, Nathaniel W. - Abstract:
- Abstract : Objectives: To develop a novel delivery technique to overcome vascular size limitations for device closure of a secundum ASD in the young patient. Background: The efficacy, clinical utility, and safety of transcatheter closure in comparison to surgical closure are well established; however, access vessel size remains a potential limitation to device selection in the young patient. Methods: A retrospective medical record review of all cases of secundum ASD closure between April 2015 and April 2016 utilizing a novel delivery method described herein at two separate academic institutions. The device is disassembled allowing the delivery sheath to serve as the introducer. It is advanced to the RA with re‐introduction of the device allowing for deployment in the standard fashion. Results: Overall, 10 patients underwent secundum ASD closure via this novel delivery technique. There were six females (60%) and the average age at time of the procedure was 4.2 years ± 1.6 years (range 2–6 years). The majority of patients had an isolated secundum ASD (70%) with the primary indication for closure being right ventricular volume overload (90%). All patients had successful placement of a GORE ® CARDIOFORM ® Septal Occluder (GCSO) with no or trivial residual shunt. No patients had vascular complications related to the procedure. Conclusions: We describe a novel technique for ASD closure using the GCSO delivery sheath as the access sheath, which reduces the vascular access sizeAbstract : Objectives: To develop a novel delivery technique to overcome vascular size limitations for device closure of a secundum ASD in the young patient. Background: The efficacy, clinical utility, and safety of transcatheter closure in comparison to surgical closure are well established; however, access vessel size remains a potential limitation to device selection in the young patient. Methods: A retrospective medical record review of all cases of secundum ASD closure between April 2015 and April 2016 utilizing a novel delivery method described herein at two separate academic institutions. The device is disassembled allowing the delivery sheath to serve as the introducer. It is advanced to the RA with re‐introduction of the device allowing for deployment in the standard fashion. Results: Overall, 10 patients underwent secundum ASD closure via this novel delivery technique. There were six females (60%) and the average age at time of the procedure was 4.2 years ± 1.6 years (range 2–6 years). The majority of patients had an isolated secundum ASD (70%) with the primary indication for closure being right ventricular volume overload (90%). All patients had successful placement of a GORE ® CARDIOFORM ® Septal Occluder (GCSO) with no or trivial residual shunt. No patients had vascular complications related to the procedure. Conclusions: We describe a novel technique for ASD closure using the GCSO delivery sheath as the access sheath, which reduces the vascular access size requirement by 25%, thus addressing one common limitation for percutaneous device closure of a secundum ASD in young patients. © 2016 Wiley Periodicals, Inc. … (more)
- Is Part Of:
- Catheterization and cardiovascular interventions. Volume 89:Issue 7(2017)
- Journal:
- Catheterization and cardiovascular interventions
- Issue:
- Volume 89:Issue 7(2017)
- Issue Display:
- Volume 89, Issue 7 (2017)
- Year:
- 2017
- Volume:
- 89
- Issue:
- 7
- Issue Sort Value:
- 2017-0089-0007-0000
- Page Start:
- 1232
- Page End:
- 1238
- Publication Date:
- 2016-11-29
- Subjects:
- endovascular procedures -- septal occluder device
Heart -- Diseases -- Diagnosis -- Periodicals
Cardiac catheterization -- Periodicals
616.1207572 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1522-726X ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/ccd.26874 ↗
- Languages:
- English
- ISSNs:
- 1522-1946
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3092.992000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 1836.xml