P30. Experience feedback of a radiation protection incident in radiotherapy: Irradiation of a pregnant patient. (December 2016)
- Record Type:
- Journal Article
- Title:
- P30. Experience feedback of a radiation protection incident in radiotherapy: Irradiation of a pregnant patient. (December 2016)
- Main Title:
- P30. Experience feedback of a radiation protection incident in radiotherapy: Irradiation of a pregnant patient
- Authors:
- Batalla, A.
Menard, B.
Gery, B. - Abstract:
- Abstract : Introduction: A woman, treated by radiotherapy in the Centre Francois Baclesse in Caen for a head and neck cancer was proving pregnant (3rd month of pregnancy during the treatment).The pregnancy was revealed after the completion of the treatment (30 fractions – 2 Gy / fraction).In order to know the dose received by the fetus, measurements have been achieved with identical conditions as for the treatment. Materials and methods: Irradiation technique:VMAT on a Clinac iX (Varian) linac.2 arcs of 307 UM and 278 UM with 6 MV photons.Phantom:Compact polystyrene to simulate the body and head and neck cylindrical phantom to simulate the irradiated volume.The ionization chamber was positioned at 45 cm of the field edge.This distance was measured by the medical doctor on the patient.The measurements were achieved with the same parameters recorded on the "record and verify" software (Mosaiq – Elekta).The measurement cumulated 5 repetitions to improve the accuracy. Results: For 1 fraction:0.83 mGy For the whole treatment (30 fractions):24.9 mGy Discussion: 1/The dose far from the irradiation field has 3 components[1] -Photons scattered by the irradiated volume of the patient (internal) -Photons scattered by the irradiated materials (couch) and the walls of the treatment vault -Leakage radiation from the linac 2/The 1st component of the dose (scattered photons inside the patient) is very low because of the large distance between the field edge and the measuring point.In theAbstract : Introduction: A woman, treated by radiotherapy in the Centre Francois Baclesse in Caen for a head and neck cancer was proving pregnant (3rd month of pregnancy during the treatment).The pregnancy was revealed after the completion of the treatment (30 fractions – 2 Gy / fraction).In order to know the dose received by the fetus, measurements have been achieved with identical conditions as for the treatment. Materials and methods: Irradiation technique:VMAT on a Clinac iX (Varian) linac.2 arcs of 307 UM and 278 UM with 6 MV photons.Phantom:Compact polystyrene to simulate the body and head and neck cylindrical phantom to simulate the irradiated volume.The ionization chamber was positioned at 45 cm of the field edge.This distance was measured by the medical doctor on the patient.The measurements were achieved with the same parameters recorded on the "record and verify" software (Mosaiq – Elekta).The measurement cumulated 5 repetitions to improve the accuracy. Results: For 1 fraction:0.83 mGy For the whole treatment (30 fractions):24.9 mGy Discussion: 1/The dose far from the irradiation field has 3 components[1] -Photons scattered by the irradiated volume of the patient (internal) -Photons scattered by the irradiated materials (couch) and the walls of the treatment vault -Leakage radiation from the linac 2/The 1st component of the dose (scattered photons inside the patient) is very low because of the large distance between the field edge and the measuring point.In the ICRP report #84 (Pregnancy and medical radiation)[2] off axis values of dose are given for cobalt 60:approximately 0.25% at 45 cm but for linacs this off axis dose is lower 2–5 times (depending on beam energy).Our measurements (0.04%) are so, in good agreement with the ICRP #84 data.In the literature, several authors reported equivalent fetal doses for head and neck irradiations[3] [4] [7] .For irradiated volumes closer to the uterus (Hodgkin desease), the doses are obviously much greater[5] [6] . Conclusion: The value of the effective dose measured ( ≈ 25 mSv) is superior to the upper legal limit for the fetus (1 mSv). In consequence, this case was declared as an incident to the French Nuclear Safety Authority (ASN) but it is assumed not to be critical for the children: – Malformations have a threshold of 100–200 mGy or higher and are typically associated with central nervous system problems[1] – A fetal dose of 100 mGy has a small individual risk of radiation-induced cancer. There is over a 99% chance that the exposed fetus will NOT develop childhood cancer or leukaemia[2] … (more)
- Is Part Of:
- Physica medica. Volume 32(2016)Supplement 4
- Journal:
- Physica medica
- Issue:
- Volume 32(2016)Supplement 4
- Issue Display:
- Volume 32, Issue 4 (2016)
- Year:
- 2016
- Volume:
- 32
- Issue:
- 4
- Issue Sort Value:
- 2016-0032-0004-0000
- Page Start:
- 379
- Page End:
- 380
- Publication Date:
- 2016-12
- Subjects:
- Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2016.11.042 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 29.xml