Radiation dose among different cardiac and vascular invasive procedures: The RODEO study. (1st August 2017)
- Record Type:
- Journal Article
- Title:
- Radiation dose among different cardiac and vascular invasive procedures: The RODEO study. (1st August 2017)
- Main Title:
- Radiation dose among different cardiac and vascular invasive procedures: The RODEO study
- Authors:
- Sciahbasi, Alessandro
Ferrante, Giuseppe
Fischetti, Dionigi
Miklin, Daniel Jonathan
Sarandrea, Alessandro
Schirripa, Valentina
Guarracini, Fabrizio
Arya, Arash
Loreni, Giorgio
Bruni, Antonio
Rigattieri, Stefano
Marini, Massimiliano
Vagnarelli, Simone
Scevola, Germano
Reimers, Bernhard
Hindricks, Gerhard
Cohen, Mauricio - Abstract:
- Abstract: Background: Interventional radiology and cardiology procedures contribute significantly to the collective dose of radiation exposure from medical interventions. Recent and dedicated studies comparing directly these procedures in term of patient radiation exposure are lacking. Our aim was to compare radiation exposure among different interventional procedures performed under fluoroscopic guidance. Methods: The RODEO study (NCT: 02972736) is an international observational retrospective multicenter study enrolling all patients undergoing diagnostic or interventional procedures performed by different interventional operators (i.e. radiologists, interventional cardiologists or electrophysiologists) in 6 centers, without exclusion criteria. The primary end-point of the study was the comparison of dose area product (DAP) among interventional cardiology, electrophysiology or interventional radiology procedures. Results: A total of 17, 711 procedures were included in the study: 13, 522 interventional cardiology, 2352 electrophysiology and 1864 interventional radiology procedures. The highest DAP values were observed for interventional radiology procedures (74 Gy ∗ cm 2 [Interquartile range 27–178 Gy ∗ cm 2 ]), followed by interventional cardiology (40 Gy ∗ cm 2 [22–78 Gy ∗ cm 2 ]) and electrophysiology procedures (13 Gy ∗ cm 2 [4–44 Gy ∗ cm 2 ], p < 0.0001). In term of specific procedures, the highest DAP values were observed in structural or valvular cardiac proceduresAbstract: Background: Interventional radiology and cardiology procedures contribute significantly to the collective dose of radiation exposure from medical interventions. Recent and dedicated studies comparing directly these procedures in term of patient radiation exposure are lacking. Our aim was to compare radiation exposure among different interventional procedures performed under fluoroscopic guidance. Methods: The RODEO study (NCT: 02972736) is an international observational retrospective multicenter study enrolling all patients undergoing diagnostic or interventional procedures performed by different interventional operators (i.e. radiologists, interventional cardiologists or electrophysiologists) in 6 centers, without exclusion criteria. The primary end-point of the study was the comparison of dose area product (DAP) among interventional cardiology, electrophysiology or interventional radiology procedures. Results: A total of 17, 711 procedures were included in the study: 13, 522 interventional cardiology, 2352 electrophysiology and 1864 interventional radiology procedures. The highest DAP values were observed for interventional radiology procedures (74 Gy ∗ cm 2 [Interquartile range 27–178 Gy ∗ cm 2 ]), followed by interventional cardiology (40 Gy ∗ cm 2 [22–78 Gy ∗ cm 2 ]) and electrophysiology procedures (13 Gy ∗ cm 2 [4–44 Gy ∗ cm 2 ], p < 0.0001). In term of specific procedures, the highest DAP values were observed in structural or valvular cardiac procedures (134 Gy ∗ cm 2 [51–260 Gy ∗ cm 2 ]) whereas the lowest DAP values in pacemaker insertion (11 Gy ∗ cm 2 [4–28 Gy ∗ cm 2 ]). Conclusion: In this large multicenter study, the highest radiation exposure was observed in procedures performed by interventional radiologists. However, among specific procedures, structural or valvular cardiac procedures were associated with the highest radiation exposure. … (more)
- Is Part Of:
- International journal of cardiology. Volume 240(2017)
- Journal:
- International journal of cardiology
- Issue:
- Volume 240(2017)
- Issue Display:
- Volume 240, Issue 2017 (2017)
- Year:
- 2017
- Volume:
- 240
- Issue:
- 2017
- Issue Sort Value:
- 2017-0240-2017-0000
- Page Start:
- 92
- Page End:
- 96
- Publication Date:
- 2017-08-01
- Subjects:
- AICD automatic implantable cardioverter defibrillator -- ANOVA analysis of variance -- CA coronary angiography -- DAP dose area product -- ED effective dose -- PCI percutaneous coronary interventions -- PM pacemakers -- TAVI transcatheter aortic valve implantation -- TIPS transjugular intrahepatic porto-systemic shunt
Radiation -- PCI -- Cardiology -- Electrophysiology -- Radiology
Cardiology -- Periodicals
Electronic journals
616.12 - Journal URLs:
- http://www.clinicalkey.com/dura/browse/journalIssue/01675273 ↗
http://www.sciencedirect.com/science/journal/01675273 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.ijcard.2017.03.031 ↗
- Languages:
- English
- ISSNs:
- 0167-5273
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4542.158000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 1312.xml