Dosimetric comparison of Acuros XB with collapsed cone convolution/superposition and anisotropic analytic algorithm for stereotactic ablative radiotherapy of thoracic spinal metastases. (8th July 2015)
- Record Type:
- Journal Article
- Title:
- Dosimetric comparison of Acuros XB with collapsed cone convolution/superposition and anisotropic analytic algorithm for stereotactic ablative radiotherapy of thoracic spinal metastases. (8th July 2015)
- Main Title:
- Dosimetric comparison of Acuros XB with collapsed cone convolution/superposition and anisotropic analytic algorithm for stereotactic ablative radiotherapy of thoracic spinal metastases
- Authors:
- Zhen, Heming
Hrycushko, Brian
Lee, Huichen
Timmerman, Robert
Pompoš, Arnold
Stojadinovic, Strahinja
Foster, Ryan
Jiang, Steve B.
Solberg, Timothy
Gu, Xuejun - Abstract:
- Abstract : The aim of this study is to compare the recent Eclipse Acuros XB (AXB) dose calculation engine with the Pinnacle collapsed cone convolution/superposition (CCC) dose calculation algorithm and the Eclipse anisotropic analytic algorithm (AAA) for stereotactic ablative radiotherapy (SAbR) treatment planning of thoracic spinal (T‐spine) metastases using IMRT and VMAT delivery techniques. The three commissioned dose engines (CCC, AAA, and AXB) were validated with ion chamber and EBT2 film measurements utilizing a heterogeneous slab‐geometry water phantom and an anthropomorphic phantom. Step‐and‐shoot IMRT and VMAT treatment plans were developed and optimized for eight patients in Pinnacle, following our institutional SAbR protocol for spinal metastases. The CCC algorithm, with heterogeneity corrections, was used for dose calculations. These plans were then exported to Eclipse and recalculated using the AAA and AXB dose calculation algorithms. Various dosimetric parameters calculated with CCC and AAA were compared to that of the AXB calculations. In regions receiving above 50% of prescription dose, the calculated CCC mean dose is 3.1%–4.1% higher than that of AXB calculations for IMRT plans and 2.8%–3.5% higher for VMAT plans, while the calculated AAA mean dose is 1.5%–2.4% lower for IMRT and 1.2%–1.6% lower for VMAT. Statistically significant differences ( p < 0.05 ) were observed for most GTV and PTV indices between the CCC and AXB calculations for IMRT and VMAT, whileAbstract : The aim of this study is to compare the recent Eclipse Acuros XB (AXB) dose calculation engine with the Pinnacle collapsed cone convolution/superposition (CCC) dose calculation algorithm and the Eclipse anisotropic analytic algorithm (AAA) for stereotactic ablative radiotherapy (SAbR) treatment planning of thoracic spinal (T‐spine) metastases using IMRT and VMAT delivery techniques. The three commissioned dose engines (CCC, AAA, and AXB) were validated with ion chamber and EBT2 film measurements utilizing a heterogeneous slab‐geometry water phantom and an anthropomorphic phantom. Step‐and‐shoot IMRT and VMAT treatment plans were developed and optimized for eight patients in Pinnacle, following our institutional SAbR protocol for spinal metastases. The CCC algorithm, with heterogeneity corrections, was used for dose calculations. These plans were then exported to Eclipse and recalculated using the AAA and AXB dose calculation algorithms. Various dosimetric parameters calculated with CCC and AAA were compared to that of the AXB calculations. In regions receiving above 50% of prescription dose, the calculated CCC mean dose is 3.1%–4.1% higher than that of AXB calculations for IMRT plans and 2.8%–3.5% higher for VMAT plans, while the calculated AAA mean dose is 1.5%–2.4% lower for IMRT and 1.2%–1.6% lower for VMAT. Statistically significant differences ( p < 0.05 ) were observed for most GTV and PTV indices between the CCC and AXB calculations for IMRT and VMAT, while differences between the AAA and AXB calculations were not statistically significant. For T‐spine SAbR treatment planning, the CCC calculations give a statistically significant overestimation of target dose compared to AXB. AAA underestimates target dose with no statistical significance compared to AXB. Further study is needed to determine the clinical impact of these findings. PACS number: 87.55.D‐, 87.53.Ly … (more)
- Is Part Of:
- Journal of applied clinical medical physics. Volume 16:Number 4(2015)
- Journal:
- Journal of applied clinical medical physics
- Issue:
- Volume 16:Number 4(2015)
- Issue Display:
- Volume 16, Issue 4 (2015)
- Year:
- 2015
- Volume:
- 16
- Issue:
- 4
- Issue Sort Value:
- 2015-0016-0004-0000
- Page Start:
- 181
- Page End:
- 192
- Publication Date:
- 2015-07-08
- Subjects:
- Acuros -- heterogeneity correction -- stereotactic -- SBRT -- spine radiotherapy
Medical physics -- Periodicals
Clinical medicine -- Periodicals
Health Physics
Clinical Medicine
Electronic journals
Periodicals
Periodicals
Fulltext
Internet Resources
610.153 - Journal URLs:
- http://aapm.onlinelibrary.wiley.com/hub/journal/10.1002/(ISSN)1526-9914/ ↗
http://bibpurl.oclc.org/web/7294 ↗
http://www.jacmp.org/ ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1120/jacmp.v16i4.5493 ↗
- Languages:
- English
- ISSNs:
- 1526-9914
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 1462.xml