Collision indicator charts for gantry‐couch position combinations for Siemens ONCOR and Elekta Infinity linacs. (6th September 2013)
- Record Type:
- Journal Article
- Title:
- Collision indicator charts for gantry‐couch position combinations for Siemens ONCOR and Elekta Infinity linacs. (6th September 2013)
- Main Title:
- Collision indicator charts for gantry‐couch position combinations for Siemens ONCOR and Elekta Infinity linacs
- Authors:
- Becker, Stewart J.
Culberson, Wes
Flynn, Ryan T. - Abstract:
- Abstract : Noncoplanar radiation fields from a linear accelerator can be used to deliver radiation dose distributions that are superior to those delivered using coplanar radiation fields. Noncoplanar radiation field arrangements are especially valuable when delivering stereotactic body radiation therapy (SBRT). Noncoplanar radiation fields, however, are geometrically more challenging to deliver than coplanar radiation fields, and are associated with a greater risk of collisions between the gantry, treatment couch, and patient. Knowledge of which treatment couch offset, treatment couch angle, and gantry angle combinations provide a collision‐free radiotherapy delivery is useful in the treatment planning process, as the risk of requiring replanning due to improperly selected treatment parameters can be minimized. Such tables are by default specific to the linear accelerator make and model used for treatment. In this work a set of plots is presented indicating which combination of treatment couch lateral offsets (‐10 cm to 10 cm), couch angles (270° to 90°), and gantry angles (0° to 360°), will result in collision‐free radiation delivery using Siemens ONCOR linear accelerators equipped with a 160‐leaf multileaf collimator and a 550 TxT treatment table, and a Elekta Infinity linear accelerator with an MLCi2 and Elekta iBEAM evo Couchtop EP. The patient was assumed to have a width of 50 cm and a height of 25 cm. PACS numbers: 87.55.‐x, 87.56.‐v
- Is Part Of:
- Journal of applied clinical medical physics. Volume 14:Number 5(2013)
- Journal:
- Journal of applied clinical medical physics
- Issue:
- Volume 14:Number 5(2013)
- Issue Display:
- Volume 14, Issue 5 (2013)
- Year:
- 2013
- Volume:
- 14
- Issue:
- 5
- Issue Sort Value:
- 2013-0014-0005-0000
- Page Start:
- 278
- Page End:
- 283
- Publication Date:
- 2013-09-06
- Subjects:
- collision -- treatment planning -- gantry -- couch -- noncoplanar -- Siemens ONCOR -- Elekta Infinity
Medical physics -- Periodicals
Clinical medicine -- Periodicals
Health Physics
Clinical Medicine
Electronic journals
Periodicals
Periodicals
Fulltext
Internet Resources
610.153 - Journal URLs:
- http://aapm.onlinelibrary.wiley.com/hub/journal/10.1002/(ISSN)1526-9914/ ↗
http://bibpurl.oclc.org/web/7294 ↗
http://www.jacmp.org/ ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1120/jacmp.v14i5.4355 ↗
- Languages:
- English
- ISSNs:
- 1526-9914
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 2119.xml