Automatic T2* determination for quantification of iron load in heart and liver: a comparison between automatic inline Maximum Likelihood Estimate and the truncation and offset methods. (16th October 2015)
- Record Type:
- Journal Article
- Title:
- Automatic T2* determination for quantification of iron load in heart and liver: a comparison between automatic inline Maximum Likelihood Estimate and the truncation and offset methods. (16th October 2015)
- Main Title:
- Automatic T2* determination for quantification of iron load in heart and liver: a comparison between automatic inline Maximum Likelihood Estimate and the truncation and offset methods
- Authors:
- Hedström, Erik
Voigt, Tobias
Greil, Gerald
Schaeffter, Tobias
Nagel, Eike - Abstract:
- Summary: Purpose: To validate ironload T2* by automatic inline Maximum Likelihood Estimate (MLE) with k‐space Rician noise correction, against the manual and automated truncation, as well as offset methods, in phantoms and in heart and liver in patients. Methods: Twenty‐five patients and an iron‐oxide phantom were scanned at 1.5T using 2 multi‐echo gradient‐echo sequences. All parameters were identical (voxel 2–3 × 2–3 × 10 mm 3, 10 echoes, TR = 26 ms, FA = 20°, BW = 833 Hz, SENSE = 2) except for TE (cardiac: TE1 = 2·5 ms, ΔTE = 2·5 ms; liver: TE1 = 1·2 ms, ΔTE = 1·5 ms). Phantoms were scanned at 1 and 32 signal averages (NSA), with NSA32 representing low‐noise reference. Results: Phantoms: MLE showed low variability between NSA1 and NSA32 (0·02 ± 0·29 ms, CI ±0·21 ms). Between methods, no difference was shown (MLE versus all: <0·31 ms, CI < ±0·35 ms). Patients: No differences were found between methods in heart (MLE versus all: <−0·22 ms, CI < ±0·75 ms) or liver (MLE versus all: <0·12 ms, CI < ±0·26 ms). Conclusions: The automatic inline MLE method is comparable to the general reference standards for determining cardiac and liver T2* for ironload in man. An automatic inline method may simplify ironload assessment, particularly in centres seeing fewer cases.
- Is Part Of:
- Clinical physiology and functional imaging. Volume 37:Number 3(2017:May)
- Journal:
- Clinical physiology and functional imaging
- Issue:
- Volume 37:Number 3(2017:May)
- Issue Display:
- Volume 37, Issue 3 (2017)
- Year:
- 2017
- Volume:
- 37
- Issue:
- 3
- Issue Sort Value:
- 2017-0037-0003-0000
- Page Start:
- 299
- Page End:
- 304
- Publication Date:
- 2015-10-16
- Subjects:
- algorithm -- automated -- heart -- ironload -- liver -- validation
Physiology, Pathological -- Periodicals
Diagnostic imaging -- Periodicals
612 - Journal URLs:
- http://www.blackwell-synergy.com/servlet/useragent?func=showIssues&code=cpf ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1111/cpf.12303 ↗
- Languages:
- English
- ISSNs:
- 1475-0961
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3286.333520
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File: