Development of a four-dimensional Monte Carlo dose calculation system for real-time tumor-tracking irradiation with a gimbaled X-ray head. (March 2017)
- Record Type:
- Journal Article
- Title:
- Development of a four-dimensional Monte Carlo dose calculation system for real-time tumor-tracking irradiation with a gimbaled X-ray head. (March 2017)
- Main Title:
- Development of a four-dimensional Monte Carlo dose calculation system for real-time tumor-tracking irradiation with a gimbaled X-ray head
- Authors:
- Ishihara, Yoshitomo
Nakamura, Mitsuhiro
Miyabe, Yuki
Mukumoto, Nobutaka
Matsuo, Yukinori
Sawada, Akira
Kokubo, Masaki
Mizowaki, Takashi
Hiraoka, Masahiro - Abstract:
- Highlights: A 4D dose calculation system for real-time tumor tracking (RTTT) was developed. A 6-MV photon beam delivered by the Vero4DRT was simulated using EGSnrc. Differences between the simulated and measured doses were less than 3% for RTTT. Target coverage in RTTT plans was almost identical to that of the non-RTTT plans. Our system helps predicting intrafraction variations in the dose distributions. Abstract: Purpose: To develop a four-dimensional (4D) dose calculation system for real-time tumor tracking (RTTT) irradiation by the Vero4DRT. Methods: First, a 6-MV photon beam delivered by the Vero4DRT was simulated using EGSnrc. A moving phantom position was directly measured by a laser displacement gauge. The pan and tilt angles, monitor units, and the indexing time indicating the phantom position were also extracted from a log file. Next, phase space data at any angle were created from both the log file and particle data under the dynamic multileaf collimator. Irradiation both with and without RTTT, with the phantom moving, were simulated using several treatment field sizes. Each was compared with the corresponding measurement using films. Finally, dose calculation for each computed tomography dataset of 10 respiratory phases with the X-ray head rotated was performed to simulate the RTTT irradiation (4D plan) for lung, liver, and pancreatic cancer patients. Dose-volume histograms of the 4D plan were compared with those calculated on the single reference respiratoryHighlights: A 4D dose calculation system for real-time tumor tracking (RTTT) was developed. A 6-MV photon beam delivered by the Vero4DRT was simulated using EGSnrc. Differences between the simulated and measured doses were less than 3% for RTTT. Target coverage in RTTT plans was almost identical to that of the non-RTTT plans. Our system helps predicting intrafraction variations in the dose distributions. Abstract: Purpose: To develop a four-dimensional (4D) dose calculation system for real-time tumor tracking (RTTT) irradiation by the Vero4DRT. Methods: First, a 6-MV photon beam delivered by the Vero4DRT was simulated using EGSnrc. A moving phantom position was directly measured by a laser displacement gauge. The pan and tilt angles, monitor units, and the indexing time indicating the phantom position were also extracted from a log file. Next, phase space data at any angle were created from both the log file and particle data under the dynamic multileaf collimator. Irradiation both with and without RTTT, with the phantom moving, were simulated using several treatment field sizes. Each was compared with the corresponding measurement using films. Finally, dose calculation for each computed tomography dataset of 10 respiratory phases with the X-ray head rotated was performed to simulate the RTTT irradiation (4D plan) for lung, liver, and pancreatic cancer patients. Dose-volume histograms of the 4D plan were compared with those calculated on the single reference respiratory phase without the gimbal rotation [three-dimensional (3D) plan]. Results: Differences between the simulated and measured doses were less than 3% for RTTT irradiation in most areas, except the high-dose gradient. For clinical cases, the target coverage in 4D plans was almost identical to that of the 3D plans. However, the doses to organs at risk in the 4D plans varied at intermediate- and low-dose levels. Conclusions: Our proposed system has acceptable accuracy for RTTT irradiation in the Vero4DRT and is capable of simulating clinical RTTT plans. … (more)
- Is Part Of:
- Physica medica. Volume 35(2017)
- Journal:
- Physica medica
- Issue:
- Volume 35(2017)
- Issue Display:
- Volume 35, Issue 2017 (2017)
- Year:
- 2017
- Volume:
- 35
- Issue:
- 2017
- Issue Sort Value:
- 2017-0035-2017-0000
- Page Start:
- 59
- Page End:
- 65
- Publication Date:
- 2017-03
- Subjects:
- Monte Carlo -- 4D dose calculation -- Real-time tumor-tracking irradiation -- SBRT -- IMRT
Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2017.02.004 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 206.xml