MD‐Logic overnight type 1 diabetes control in home settings: A multicentre, multinational, single blind randomized trial. Issue 4 (19th January 2017)
- Record Type:
- Journal Article
- Title:
- MD‐Logic overnight type 1 diabetes control in home settings: A multicentre, multinational, single blind randomized trial. Issue 4 (19th January 2017)
- Main Title:
- MD‐Logic overnight type 1 diabetes control in home settings: A multicentre, multinational, single blind randomized trial
- Authors:
- Nimri, Revital
Bratina, Natasa
Kordonouri, Olga
Avbelj Stefanija, Magdalena
Fath, Maryam
Biester, Torben
Muller, Ido
Atlas, Eran
Miller, Shahar
Fogel, Aviel
Phillip, Moshe
Danne, Thomas
Battelino, Tadej - Abstract:
- Abstract : Aims: To evaluate the safety, efficacy and need for remote monitoring of the MD‐Logic closed‐loop system during short‐term overnight use at home. Methods: Seventy‐five patients (38 male; aged 10‐54 years; average A1c, 7.8% ± 0.7%, 61.8 ± 7.2 mmol/mol) were enrolled from 3 clinical sites. Patients were randomly assigned to participate in 2 overnight crossover periods, each including 4 consecutive nights, 1 under closed‐loop control and 1 under sensor‐augmented pump (SAP) therapy in the patient's home. Both study arms were supervised using a remote‐monitoring system in a blinded manner. Primary endpoints were time spent with glucose levels below 70 mg/dL and percentage of nights in which mean overnight glucose levels were within 90 to 140 mg/dL. Results: The median [interquartile range] percentage of time spent in hypoglycaemia was significantly lower on nights when MD‐Logic was used, compared to SAP therapy (2.07 [0, 4.78] and 2.6 [0, 10.34], respectively; P = .004) and the percentage of individual nights with a mean overnight glucose level in target was significantly greater (75 [42, 75] and 50 [25, 75], respectively; P = .008). The time spent in target range was increased by a median of 28% ( P = .001), with the same amount of insulin (10.69 [7.28, 13.94] and 10.41[6.9, 14.07], respectively; P = .087). The remote monitoring triggered calls for hypoglycaemia at twice the rate during SAP therapy compared to closed‐loop control (62 and 29, respectively; PAbstract : Aims: To evaluate the safety, efficacy and need for remote monitoring of the MD‐Logic closed‐loop system during short‐term overnight use at home. Methods: Seventy‐five patients (38 male; aged 10‐54 years; average A1c, 7.8% ± 0.7%, 61.8 ± 7.2 mmol/mol) were enrolled from 3 clinical sites. Patients were randomly assigned to participate in 2 overnight crossover periods, each including 4 consecutive nights, 1 under closed‐loop control and 1 under sensor‐augmented pump (SAP) therapy in the patient's home. Both study arms were supervised using a remote‐monitoring system in a blinded manner. Primary endpoints were time spent with glucose levels below 70 mg/dL and percentage of nights in which mean overnight glucose levels were within 90 to 140 mg/dL. Results: The median [interquartile range] percentage of time spent in hypoglycaemia was significantly lower on nights when MD‐Logic was used, compared to SAP therapy (2.07 [0, 4.78] and 2.6 [0, 10.34], respectively; P = .004) and the percentage of individual nights with a mean overnight glucose level in target was significantly greater (75 [42, 75] and 50 [25, 75], respectively; P = .008). The time spent in target range was increased by a median of 28% ( P = .001), with the same amount of insulin (10.69 [7.28, 13.94] and 10.41[6.9, 14.07], respectively; P = .087). The remote monitoring triggered calls for hypoglycaemia at twice the rate during SAP therapy compared to closed‐loop control (62 and 29, respectively; P = .002). Conclusions: The MD‐Logic system demonstrated a safe and efficient profile during overnight use by children, adolescents and adults with type 1 diabetes and, therefore, provides an effective means of mitigating the risk of nocturnal hypoglycaemia. … (more)
- Is Part Of:
- Diabetes, obesity & metabolism. Volume 19:Issue 4(2017)
- Journal:
- Diabetes, obesity & metabolism
- Issue:
- Volume 19:Issue 4(2017)
- Issue Display:
- Volume 19, Issue 4 (2017)
- Year:
- 2017
- Volume:
- 19
- Issue:
- 4
- Issue Sort Value:
- 2017-0019-0004-0000
- Page Start:
- 553
- Page End:
- 561
- Publication Date:
- 2017-01-19
- Subjects:
- closed‐loop -- MD‐Logic -- nocturnal hypoglycaemia -- remote‐monitoring -- type 1
Diabetes -- Periodicals
Obesity -- Periodicals
Metabolism -- Disorders -- Periodicals
Clinical pharmacology -- Periodicals
616.462 - Journal URLs:
- http://www.blackwellpublishing.com/journal.asp?ref=1462-8902&site=1 ↗
http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1463-1326 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1111/dom.12852 ↗
- Languages:
- English
- ISSNs:
- 1462-8902
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3579.601970
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 2278.xml