A benchmarking method to evaluate the accuracy of a commercial proton monte carlo pencil beam scanning treatment planning system. (2nd February 2017)
- Record Type:
- Journal Article
- Title:
- A benchmarking method to evaluate the accuracy of a commercial proton monte carlo pencil beam scanning treatment planning system. (2nd February 2017)
- Main Title:
- A benchmarking method to evaluate the accuracy of a commercial proton monte carlo pencil beam scanning treatment planning system
- Authors:
- Lin, Liyong
Huang, Sheng
Kang, Minglei
Hiltunen, Petri
Vanderstraeten, Reynald
Lindberg, Jari
Siljamaki, Sami
Wareing, Todd
Davis, Ian
Barnett, Allen
McGhee, John
Simone, Charles B.
Solberg, Timothy D.
McDonough, James E.
Ainsley, Christopher - Abstract:
- Abstract: AcurosPT is a Monte Carlo algorithm in the Eclipse 13.7 treatment planning system, which is designed to provide rapid and accurate dose calculations for proton therapy. Computational run‐time in minimized by simplifying or eliminating less significant physics processes. In this article, the accuracy of AcurosPT was benchmarked against both measurement and an independent MC calculation, TOPAS. Such a method can be applied to any new MC calculation for the detection of potential inaccuracies. To validate multiple Coulomb scattering (MCS) which affects primary beam broadening, single spot profiles in a Solidwater ® phantom were compared for beams of five selected proton energies between AcurosPT, measurement and TOPAS. The spot Gaussian sigma in AcurosPT was found to increase faster with depth than both measurement and TOPAS, suggesting that the MCS algorithm in AcurosPT overestimates the scattering effect. To validate AcurosPT modeling of the halo component beyond primary beam broadening, field size factors (FSF) were compared for multi‐spot profiles measured in a water phantom. The FSF for small field sizes were found to disagree with measurement, with the disagreement increasing with depth. Conversely, TOPAS simulations of the same FSF consistently agreed with measurement to within 1.5%. The disagreement in absolute dose between AcurosPT and measurement was smaller than 2% at the mid‐range depth of multi‐energy beams. While AcurosPT calculates acceptable doseAbstract: AcurosPT is a Monte Carlo algorithm in the Eclipse 13.7 treatment planning system, which is designed to provide rapid and accurate dose calculations for proton therapy. Computational run‐time in minimized by simplifying or eliminating less significant physics processes. In this article, the accuracy of AcurosPT was benchmarked against both measurement and an independent MC calculation, TOPAS. Such a method can be applied to any new MC calculation for the detection of potential inaccuracies. To validate multiple Coulomb scattering (MCS) which affects primary beam broadening, single spot profiles in a Solidwater ® phantom were compared for beams of five selected proton energies between AcurosPT, measurement and TOPAS. The spot Gaussian sigma in AcurosPT was found to increase faster with depth than both measurement and TOPAS, suggesting that the MCS algorithm in AcurosPT overestimates the scattering effect. To validate AcurosPT modeling of the halo component beyond primary beam broadening, field size factors (FSF) were compared for multi‐spot profiles measured in a water phantom. The FSF for small field sizes were found to disagree with measurement, with the disagreement increasing with depth. Conversely, TOPAS simulations of the same FSF consistently agreed with measurement to within 1.5%. The disagreement in absolute dose between AcurosPT and measurement was smaller than 2% at the mid‐range depth of multi‐energy beams. While AcurosPT calculates acceptable dose distributions for typical clinical beams, users are cautioned of potentially larger errors at distal depths due to overestimated MCS and halo implementation. … (more)
- Is Part Of:
- Journal of applied clinical medical physics. Volume 18:Number 2(2017)
- Journal:
- Journal of applied clinical medical physics
- Issue:
- Volume 18:Number 2(2017)
- Issue Display:
- Volume 18, Issue 2 (2017)
- Year:
- 2017
- Volume:
- 18
- Issue:
- 2
- Issue Sort Value:
- 2017-0018-0002-0000
- Page Start:
- 44
- Page End:
- 49
- Publication Date:
- 2017-02-02
- Subjects:
- AcurosPT -- commissioning -- Monte Carlo dose calculation -- pencil beam scanning -- proton therapy
Medical physics -- Periodicals
Clinical medicine -- Periodicals
Health Physics
Clinical Medicine
Electronic journals
Periodicals
Periodicals
Fulltext
Internet Resources
610.153 - Journal URLs:
- http://aapm.onlinelibrary.wiley.com/hub/journal/10.1002/(ISSN)1526-9914/ ↗
http://bibpurl.oclc.org/web/7294 ↗
http://www.jacmp.org/ ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/acm2.12043 ↗
- Languages:
- English
- ISSNs:
- 1526-9914
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 773.xml