Detection near 1-nm with a laminar-flow, water-based condensation particle counter. Issue 3 (4th March 2017)
- Record Type:
- Journal Article
- Title:
- Detection near 1-nm with a laminar-flow, water-based condensation particle counter. Issue 3 (4th March 2017)
- Main Title:
- Detection near 1-nm with a laminar-flow, water-based condensation particle counter
- Authors:
- Hering, Susanne V.
Lewis, Gregory S.
Spielman, Steven R.
Eiguren-Fernandez, Arantzazu
Kreisberg, Nathan M.
Kuang, Chongai
Attoui, Michel - Abstract:
- ABSTRACT: Presented is a laminar-flow, water-based condensation particle counter capable of particle detection near 1 nm. This instrument employs a three-stage, laminar-flow growth tube with a "moderator" stage that reduces the temperature and water content of the output flow without reducing the peak supersaturation, and makes feasible operation at the large temperature differences necessary for achieving high supersaturations. The instrument has an aerosol flow of 0.3 L/min, and does not use a filtered sheath flow. It is referred to as a "versatile" water condensation particle counter, or vWCPC, as operating temperatures can be adjusted in accordance with the cut-point desired. When operated with wall temperatures of ∼2°C, >90°C, and ∼22°C for the three stages, respectively, the vWCPC detects particles generated from a heated nichrome wire with a 50% efficiency cut-point near 1.6 nm mobility diameter. At these operating temperatures, it also detects 10–20% of large molecular ions formed from passing filtered ambient air through a bipolar ion source. Decreasing the temperature difference between the first two stages, with the first and second stages operated at 10 and 90°C, respectively, essentially eliminates the response to charger ions, and raises the 50% efficiency cut-point for the nichrome wire particles to 1.9 nm mobility diameter. The time response, as measured by rapid removal of an inlet filter, yields a characteristic time constant of 195 ms. Copyright © 2017ABSTRACT: Presented is a laminar-flow, water-based condensation particle counter capable of particle detection near 1 nm. This instrument employs a three-stage, laminar-flow growth tube with a "moderator" stage that reduces the temperature and water content of the output flow without reducing the peak supersaturation, and makes feasible operation at the large temperature differences necessary for achieving high supersaturations. The instrument has an aerosol flow of 0.3 L/min, and does not use a filtered sheath flow. It is referred to as a "versatile" water condensation particle counter, or vWCPC, as operating temperatures can be adjusted in accordance with the cut-point desired. When operated with wall temperatures of ∼2°C, >90°C, and ∼22°C for the three stages, respectively, the vWCPC detects particles generated from a heated nichrome wire with a 50% efficiency cut-point near 1.6 nm mobility diameter. At these operating temperatures, it also detects 10–20% of large molecular ions formed from passing filtered ambient air through a bipolar ion source. Decreasing the temperature difference between the first two stages, with the first and second stages operated at 10 and 90°C, respectively, essentially eliminates the response to charger ions, and raises the 50% efficiency cut-point for the nichrome wire particles to 1.9 nm mobility diameter. The time response, as measured by rapid removal of an inlet filter, yields a characteristic time constant of 195 ms. Copyright © 2017 American Association for Aerosol Research … (more)
- Is Part Of:
- Aerosol science and technology. Volume 51:Issue 3(2017)
- Journal:
- Aerosol science and technology
- Issue:
- Volume 51:Issue 3(2017)
- Issue Display:
- Volume 51, Issue 3 (2017)
- Year:
- 2017
- Volume:
- 51
- Issue:
- 3
- Issue Sort Value:
- 2017-0051-0003-0000
- Page Start:
- 354
- Page End:
- 362
- Publication Date:
- 2017-03-04
- Subjects:
- Jinkun Jiang
Aerosols -- Periodicals
Aerosol Propellants -- Periodicals
Aerosols -- Periodicals
660.294515 - Journal URLs:
- http://www.tandfonline.com/loi/uast20#.VkNQFJUnyig ↗
http://www.tandfonline.com/ ↗ - DOI:
- 10.1080/02786826.2016.1262531 ↗
- Languages:
- English
- ISSNs:
- 0278-6826
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 0729.835400
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 65.xml