Experimental investigation on minimum frame rate requirements of high-speed videoendoscopy for clinical voice assessment. (March 2015)
- Record Type:
- Journal Article
- Title:
- Experimental investigation on minimum frame rate requirements of high-speed videoendoscopy for clinical voice assessment. (March 2015)
- Main Title:
- Experimental investigation on minimum frame rate requirements of high-speed videoendoscopy for clinical voice assessment
- Authors:
- Deliyski, Dimitar D.
Powell, Maria E.G.
Zacharias, Stephanie R.C.
Gerlach, Terri Treman
de Alarcon, Alessandro - Abstract:
- Highlights: Voice assessment was studied using laryngeal imaging via high-speed videoendoscopy Frame rates impact the assessment of clinical vocal fold vibratory features Glottal edge, mucosal wave, aperiodicity and contact are most sensitive to frame rate The recommended frame rates for laryngeal imaging should be 8000 frames per second The minimum frame rates for laryngeal imaging should be 4000 frames per second Abstract: This study investigated the impact of high-speed videoendoscopy (HSV) frame rates on the assessment of nine clinically relevant vocal-fold vibratory features. Fourteen adult patients with voice disorder and 14 adult normal controls were recorded using monochromatic rigid HSV at a rate of 16, 000 frames per second (fps) and spatial resolution of 639 × 639 pixels. The 16, 000-fps data were downsampled to 16 other rate denominations. Using paired comparisons design, nine common clinical vibratory features were visually compared between the downsampled and the original images. Three raters reported the thresholds at which: (1) a detectable difference between the two videos was first noticed, and (2) differences between the two videos would result in a change of clinical rating. Results indicated that glottal edge, mucosal wave magnitude and extent, aperiodicity, contact and loss of contact of the vocal folds were the vibratory features most sensitive to frame rate. Of these vibratory features, the glottal edge was selected for further analysis, due to itsHighlights: Voice assessment was studied using laryngeal imaging via high-speed videoendoscopy Frame rates impact the assessment of clinical vocal fold vibratory features Glottal edge, mucosal wave, aperiodicity and contact are most sensitive to frame rate The recommended frame rates for laryngeal imaging should be 8000 frames per second The minimum frame rates for laryngeal imaging should be 4000 frames per second Abstract: This study investigated the impact of high-speed videoendoscopy (HSV) frame rates on the assessment of nine clinically relevant vocal-fold vibratory features. Fourteen adult patients with voice disorder and 14 adult normal controls were recorded using monochromatic rigid HSV at a rate of 16, 000 frames per second (fps) and spatial resolution of 639 × 639 pixels. The 16, 000-fps data were downsampled to 16 other rate denominations. Using paired comparisons design, nine common clinical vibratory features were visually compared between the downsampled and the original images. Three raters reported the thresholds at which: (1) a detectable difference between the two videos was first noticed, and (2) differences between the two videos would result in a change of clinical rating. Results indicated that glottal edge, mucosal wave magnitude and extent, aperiodicity, contact and loss of contact of the vocal folds were the vibratory features most sensitive to frame rate. Of these vibratory features, the glottal edge was selected for further analysis, due to its higher rating reliability, universal prevalence and consistent definition. Rates of 8000 fps were found to be free from visually perceivable feature degradation, and for rates of 5333 fps, degradation was minimal. For rates of 4000 fps and higher, clinical assessments of glottal edge were not affected. Rates of 2000 fps changed the clinical ratings in over 16% of the samples, which could lead to inaccurate functional assessment. … (more)
- Is Part Of:
- Biomedical signal processing and control. Volume 17(2015)
- Journal:
- Biomedical signal processing and control
- Issue:
- Volume 17(2015)
- Issue Display:
- Volume 17, Issue 2015 (2015)
- Year:
- 2015
- Volume:
- 17
- Issue:
- 2015
- Issue Sort Value:
- 2015-0017-2015-0000
- Page Start:
- 21
- Page End:
- 28
- Publication Date:
- 2015-03
- Subjects:
- High-speed videoendoscopy -- Frame rate -- Clinical voice assessment -- Laryngeal imaging
Signal processing -- Periodicals
Biomedical engineering -- Periodicals
Signal Processing, Computer-Assisted -- Periodicals
Image Processing, Computer-Assisted -- Periodicals
Biomedical Engineering -- Periodicals
610.28 - Journal URLs:
- http://www.sciencedirect.com/science/journal/17468094 ↗
http://www.elsevier.com/journals ↗
http://www.sciencedirect.com/science?_ob=PublicationURL&_tockey=%23TOC%2329675%232006%23999989998%23626449%23FLA%23&_cdi=29675&_pubType=J&_auth=y&_acct=C000045259&_version=1&_urlVersion=0&_userid=836873&md5=664b5cf9a57fc91971a17faf20c32ec1 ↗ - DOI:
- 10.1016/j.bspc.2014.11.007 ↗
- Languages:
- English
- ISSNs:
- 1746-8094
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 2087.880400
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 570.xml