Tip Fold-over in Cochlear Implantation: Case Series. Issue 2 (February 2017)
- Record Type:
- Journal Article
- Title:
- Tip Fold-over in Cochlear Implantation: Case Series. Issue 2 (February 2017)
- Main Title:
- Tip Fold-over in Cochlear Implantation: Case Series
- Authors:
- Zuniga, M. Geraldine
Rivas, Alejandro
Hedley-Williams, Andrea
Gifford, Rene H.
Dwyer, Robert
Dawant, Benoit M.
Sunderhaus, Linsey W.
Hovis, Kristen L.
Wanna, George B.
Noble, Jack H.
Labadie, Robert F. - Abstract:
- Abstract : Objective: To describe the incidence, clinical presentation, and performance of cochlear implant (CI) recipients with tip fold-over. Study Design: Retrospective case series. Setting: Tertiary referral center. Patients: CI recipients who underwent postoperative computed tomography (CT) scanning. Intervention(s): Tip fold-over was identified tomographically using previously validated software that identifies the electrode array. Electrophysiologic testing including spread of excitation or electric field imaging (EFI) was measured on those with fold-over. Main Outcome Measure(s): Location of the fold-over; audiological performance pre and postselective deactivation of fold-over electrodes. Results: Three hundred three ears of 235 CI recipients had postoperative CTs available for review. Six (1.98%) had tip fold-over with 5/6 right-sided ears. Tip fold-over occurred predominantly at 270 degrees and was associated with precurved electrodes (5/6). Patients did not report audiological complaints during initial activation. In one patient, the electrode array remained within the scala tympani with preserved residual hearing despite the fold-over. Spread of excitation supported tip fold-over, but the predictive value was not clear. EFI predicted location of the fold-over with clear predictive value in one patient. At an average follow-up of 11 months, three subjects underwent deactivation of the overlapping electrodes with two of them showing marked audiologicalAbstract : Objective: To describe the incidence, clinical presentation, and performance of cochlear implant (CI) recipients with tip fold-over. Study Design: Retrospective case series. Setting: Tertiary referral center. Patients: CI recipients who underwent postoperative computed tomography (CT) scanning. Intervention(s): Tip fold-over was identified tomographically using previously validated software that identifies the electrode array. Electrophysiologic testing including spread of excitation or electric field imaging (EFI) was measured on those with fold-over. Main Outcome Measure(s): Location of the fold-over; audiological performance pre and postselective deactivation of fold-over electrodes. Results: Three hundred three ears of 235 CI recipients had postoperative CTs available for review. Six (1.98%) had tip fold-over with 5/6 right-sided ears. Tip fold-over occurred predominantly at 270 degrees and was associated with precurved electrodes (5/6). Patients did not report audiological complaints during initial activation. In one patient, the electrode array remained within the scala tympani with preserved residual hearing despite the fold-over. Spread of excitation supported tip fold-over, but the predictive value was not clear. EFI predicted location of the fold-over with clear predictive value in one patient. At an average follow-up of 11 months, three subjects underwent deactivation of the overlapping electrodes with two of them showing marked audiological improvement. Conclusion: In a large academic center with experienced surgeons, tip fold-over occurred at a rate of 1.98% but was not immediately identifiable clinically. CT imaging definitively showed tip fold-over. Deactivating involved electrodes may improve performance possibly avoiding revision surgery. EFI may be highly predictive of tip fold-over and can be run intraoperatively, potentially obviating the need for intraop fluoroscopy. … (more)
- Is Part Of:
- Otology & neurotology. Volume 38:Issue 2(2017)
- Journal:
- Otology & neurotology
- Issue:
- Volume 38:Issue 2(2017)
- Issue Display:
- Volume 38, Issue 2 (2017)
- Year:
- 2017
- Volume:
- 38
- Issue:
- 2
- Issue Sort Value:
- 2017-0038-0002-0000
- Page Start:
- 199
- Page End:
- 206
- Publication Date:
- 2017-02
- Subjects:
- Cochlear implant failure -- Cochlear implant revision -- Electrode tip fold-over
Otology -- Periodicals
Ear -- Diseases -- Periodicals
Skull base -- Surgery -- Periodicals
617.8005 - Journal URLs:
- http://www.otology-neurotology.com ↗
http://journals.lww.com ↗ - DOI:
- 10.1097/MAO.0000000000001283 ↗
- Languages:
- English
- ISSNs:
- 1531-7129
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6313.528000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 2610.xml