A Simplified (Screening) Approach to Check the Calibration Status of the Goldmann Applanation Tonometer. Issue 10 (October 2016)
- Record Type:
- Journal Article
- Title:
- A Simplified (Screening) Approach to Check the Calibration Status of the Goldmann Applanation Tonometer. Issue 10 (October 2016)
- Main Title:
- A Simplified (Screening) Approach to Check the Calibration Status of the Goldmann Applanation Tonometer
- Authors:
- Choudhari, Nikhil S.
Ramavath, Suresh
Rekha, Gunturu
Rao, Aparna
Rao, Harsha L.
Senthil, Sirisha
Garudadri, Chandra Sekhar - Abstract:
- Abstract : Background: To investigate whether a lower-end calibration error check alone can identify a Goldmann applanation tonometer (GAT) with acceptable calibration error as per Asia-Pacific Glaucoma Society (APGS) and World Glaucoma Association (WGA) guidelines. Methods: This was a multicentric prospective study in a tertiary eye-care institutional setting. Seventy-six slit-lamp–mounted Haag-Streit GATs (Model AT 900 C/M) were included in the study. All GATs were checked for calibration error at 0, 20, and 60 mm Hg testing levels as recommended by the manufacturer. The ability of 2 screening criteria (calibration error within ±1 or ±2 mm Hg, respectively, at the 0 mm Hg testing level without using the calibration error check weight bar) to identify a tonometer within acceptable calibration limits defined by APGS (calibration error within ±2, ±3, and ±4 mm Hg at 0, 20, and 60 mm Hg testing levels, respectively) and WGA (calibration error within ±1 mm Hg at all testing levels) was determined. Main Outcome Measure: The negative likelihood ratio of the screening criteria Results: Five (6.6%) and 42 (55%) GATs were outside of the APGS and the WGA calibration limits, respectively. Negative likelihood ratios of the first and the second screening criteria at the APGS definition of calibration error tolerance were 0.11 and 0.25, respectively, and at that of the WGA definition of calibration error tolerance were 0.38 and 0.90, respectively. Conclusions: The screening criteria ofAbstract : Background: To investigate whether a lower-end calibration error check alone can identify a Goldmann applanation tonometer (GAT) with acceptable calibration error as per Asia-Pacific Glaucoma Society (APGS) and World Glaucoma Association (WGA) guidelines. Methods: This was a multicentric prospective study in a tertiary eye-care institutional setting. Seventy-six slit-lamp–mounted Haag-Streit GATs (Model AT 900 C/M) were included in the study. All GATs were checked for calibration error at 0, 20, and 60 mm Hg testing levels as recommended by the manufacturer. The ability of 2 screening criteria (calibration error within ±1 or ±2 mm Hg, respectively, at the 0 mm Hg testing level without using the calibration error check weight bar) to identify a tonometer within acceptable calibration limits defined by APGS (calibration error within ±2, ±3, and ±4 mm Hg at 0, 20, and 60 mm Hg testing levels, respectively) and WGA (calibration error within ±1 mm Hg at all testing levels) was determined. Main Outcome Measure: The negative likelihood ratio of the screening criteria Results: Five (6.6%) and 42 (55%) GATs were outside of the APGS and the WGA calibration limits, respectively. Negative likelihood ratios of the first and the second screening criteria at the APGS definition of calibration error tolerance were 0.11 and 0.25, respectively, and at that of the WGA definition of calibration error tolerance were 0.38 and 0.90, respectively. Conclusions: The screening criteria of calibration error within ±1 mm Hg at the 0 mm Hg testing level seem to be clinically useful to detect GAT with an acceptable calibration error as per the APGS guideline. … (more)
- Is Part Of:
- Journal of glaucoma. Volume 25:Issue 10(2016)
- Journal:
- Journal of glaucoma
- Issue:
- Volume 25:Issue 10(2016)
- Issue Display:
- Volume 25, Issue 10 (2016)
- Year:
- 2016
- Volume:
- 25
- Issue:
- 10
- Issue Sort Value:
- 2016-0025-0010-0000
- Page Start:
- Page End:
- Publication Date:
- 2016-10
- Subjects:
- Goldmann applanation tonometer -- calibration error -- screening -- diagnostic ability of a test
Glaucoma -- Periodicals
617.741005 - Journal URLs:
- http://ovidsp.ovid.com/ovidweb.cgi?T=JS&NEWS=n&CSC=Y&PAGE=toc&D=yrovft&AN=00061198-000000000-00000 ↗
http://www.glaucomajournal.com ↗
http://journals.lww.com/glaucomajournal/pages/default.aspx ↗
http://journals.lww.com ↗ - DOI:
- 10.1097/IJG.0000000000000426 ↗
- Languages:
- English
- ISSNs:
- 1057-0829
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4996.230000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 2686.xml