A Comparative Review of Osseointegration Failure Between Osseointegrated Bone Conduction Device Models in Pediatric Patients. Issue 3 (March 2016)
- Record Type:
- Journal Article
- Title:
- A Comparative Review of Osseointegration Failure Between Osseointegrated Bone Conduction Device Models in Pediatric Patients. Issue 3 (March 2016)
- Main Title:
- A Comparative Review of Osseointegration Failure Between Osseointegrated Bone Conduction Device Models in Pediatric Patients
- Authors:
- Nelson, Kurt L.
Cox, Matthew D.
Richter, Gresham T.
Dornhoffer, John L. - Abstract:
- Abstract : Objective: Compare rates of osseointegration failure (OIF) between system 2 (200 series) and system 3 (BI300) osseointegrated bone conduction device (OCBD) models. Study Design: Case series with chart review. Setting: Tertiary referral center. Patients: Pediatric patients who were implanted with either system 2 or system 3 OCBD between May 2009 and July 2014 at Arkansas Children's Hospital. Interventions: Implantation with either system 2 or system 3 OCBD using 3- and 4-mm implants with 5.5- and 8.5-mm abutments (system 2) or 6- and 9-mm abutments (system 3). Main Outcome Measures: OIF with resulting loss of flange-fixture. Results: OIF occurred in 28.6% (14/49) of implanted fixtures in system 2 compared with 8.0% (2/25) of implanted fixtures in system 3 ( p = 0.042). The average time to complication for system 2 was 13 months compared with 3 months for system 3. The causes of OIF in system 2 were trauma (n = 4), chronic soft tissue disease (n = 5), or idiopathic in nature (n = 5). System 3 failures were both because of trauma. Using the combined cohorts, the average age of patients who experienced OIF was 8.7 years, compared with 11.4 years in patients without OIF ( p = 0.047). There was no difference in OIF among 3-mm and 4-mm screws ( p = 0.876), linear and flap technique ( p = 0.375), or surgeons ( p = 0.211). Conclusion: System 3 OCBD showed a significantly reduced rate of OIF compared with system 2 in our study. Modification to the system 2 implantAbstract : Objective: Compare rates of osseointegration failure (OIF) between system 2 (200 series) and system 3 (BI300) osseointegrated bone conduction device (OCBD) models. Study Design: Case series with chart review. Setting: Tertiary referral center. Patients: Pediatric patients who were implanted with either system 2 or system 3 OCBD between May 2009 and July 2014 at Arkansas Children's Hospital. Interventions: Implantation with either system 2 or system 3 OCBD using 3- and 4-mm implants with 5.5- and 8.5-mm abutments (system 2) or 6- and 9-mm abutments (system 3). Main Outcome Measures: OIF with resulting loss of flange-fixture. Results: OIF occurred in 28.6% (14/49) of implanted fixtures in system 2 compared with 8.0% (2/25) of implanted fixtures in system 3 ( p = 0.042). The average time to complication for system 2 was 13 months compared with 3 months for system 3. The causes of OIF in system 2 were trauma (n = 4), chronic soft tissue disease (n = 5), or idiopathic in nature (n = 5). System 3 failures were both because of trauma. Using the combined cohorts, the average age of patients who experienced OIF was 8.7 years, compared with 11.4 years in patients without OIF ( p = 0.047). There was no difference in OIF among 3-mm and 4-mm screws ( p = 0.876), linear and flap technique ( p = 0.375), or surgeons ( p = 0.211). Conclusion: System 3 OCBD showed a significantly reduced rate of OIF compared with system 2 in our study. Modification to the system 2 implant resulted in improved rate of OIF. Younger age was independently associated with a higher rate of OIF when evaluating both systems. … (more)
- Is Part Of:
- Otology & neurotology. Volume 37:Issue 3(2016)
- Journal:
- Otology & neurotology
- Issue:
- Volume 37:Issue 3(2016)
- Issue Display:
- Volume 37, Issue 3 (2016)
- Year:
- 2016
- Volume:
- 37
- Issue:
- 3
- Issue Sort Value:
- 2016-0037-0003-0000
- Page Start:
- 276
- Page End:
- 280
- Publication Date:
- 2016-03
- Subjects:
- BAHA -- BAHAD -- BAI -- Bone-anchored implant -- Conductive hearing loss -- Deafness -- Implanted hearing aid -- OCBD -- Osseointegrated bone conduction device -- Osseointegration -- Sensorineural hearing loss -- Single-sided
Otology -- Periodicals
Ear -- Diseases -- Periodicals
Skull base -- Surgery -- Periodicals
617.8005 - Journal URLs:
- http://www.otology-neurotology.com ↗
http://journals.lww.com ↗ - DOI:
- 10.1097/MAO.0000000000000970 ↗
- Languages:
- English
- ISSNs:
- 1531-7129
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6313.528000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 850.xml