Transapical implantation of a self‐expandable aortic valve prosthesis utilizing a novel designed positioning element. Issue 1 (26th January 2016)
- Record Type:
- Journal Article
- Title:
- Transapical implantation of a self‐expandable aortic valve prosthesis utilizing a novel designed positioning element. Issue 1 (26th January 2016)
- Main Title:
- Transapical implantation of a self‐expandable aortic valve prosthesis utilizing a novel designed positioning element
- Authors:
- Liu, Xiaopeng
Tang, Yue
Luo, Fuliang
Tian, Yi
Li, Kai
Sun, Jiakang
Jia, Liujun
Wang, Wei - Abstract:
- Abstract : Objectives: To evaluate a new transapical system which utilizes a novel designed positioning element and a two‐step positioning mechanism for easy and accurate implantation of transcatheter valves. Background: Transcatheter aortic valve implantation is an important treatment option for non‐surgical patients with severe aortic stenosis. However, accurate placement of the transcatheter valve remains challenging. Methods: Self‐expandable aortic valve prosthesis with a flexibly connected, annulus‐like positioning element was implanted through a transapical approach in 12 pigs. The positioning element was separated and can be released independent of the valve prosthesis. During valve implantation, firstly, the positioning element was unsheathed and fixed into the aortic sinus. Then, the prosthetic valve was guided to an anatomically oriented position and deployed. Six animals were followed up to 180 days. Results: With the help of the positioning element, all 12 valves were successfully deployed at the anticipated site. The valve release procedure took an average of 7.3 ± 2.5 min. The mean transvalvular pressure gradient was 2.8 ± 1.1 mm Hg at valve deployment. Of the six chronic animals, the mean transvalvular pressure gradient was 3.0 ± 1.0 mm Hg on day 7, and 2.9 ± 1.6 mm Hg on day 180 ( P = 0.91). No migration, embolization, or coronary obstruction was observed during surgery and at necropsy. Pathological examination showed anatomically correct positioning of theAbstract : Objectives: To evaluate a new transapical system which utilizes a novel designed positioning element and a two‐step positioning mechanism for easy and accurate implantation of transcatheter valves. Background: Transcatheter aortic valve implantation is an important treatment option for non‐surgical patients with severe aortic stenosis. However, accurate placement of the transcatheter valve remains challenging. Methods: Self‐expandable aortic valve prosthesis with a flexibly connected, annulus‐like positioning element was implanted through a transapical approach in 12 pigs. The positioning element was separated and can be released independent of the valve prosthesis. During valve implantation, firstly, the positioning element was unsheathed and fixed into the aortic sinus. Then, the prosthetic valve was guided to an anatomically oriented position and deployed. Six animals were followed up to 180 days. Results: With the help of the positioning element, all 12 valves were successfully deployed at the anticipated site. The valve release procedure took an average of 7.3 ± 2.5 min. The mean transvalvular pressure gradient was 2.8 ± 1.1 mm Hg at valve deployment. Of the six chronic animals, the mean transvalvular pressure gradient was 3.0 ± 1.0 mm Hg on day 7, and 2.9 ± 1.6 mm Hg on day 180 ( P = 0.91). No migration, embolization, or coronary obstruction was observed during surgery and at necropsy. Pathological examination showed anatomically correct positioning of the prosthetic valve without signs of thrombosis or calcification. Conclusions: In this study, we confirmed the feasibility of the J‐Valve transapical system for transapical implantation through a two‐step process. Satisfactory hemodynamic and pathological performance during a follow‐up of 180 days was demonstrated. © 2016 Wiley Periodicals, Inc. … (more)
- Is Part Of:
- Catheterization and cardiovascular interventions. Volume 89:Issue 1(2017)
- Journal:
- Catheterization and cardiovascular interventions
- Issue:
- Volume 89:Issue 1(2017)
- Issue Display:
- Volume 89, Issue 1 (2017)
- Year:
- 2017
- Volume:
- 89
- Issue:
- 1
- Issue Sort Value:
- 2017-0089-0001-0000
- Page Start:
- E30
- Page End:
- E37
- Publication Date:
- 2016-01-26
- Subjects:
- transapical aortic valve implantation -- anatomically correct positioning -- two‐step positioning mechanism -- preclinical evaluation
Heart -- Diseases -- Diagnosis -- Periodicals
Cardiac catheterization -- Periodicals
616.1207572 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1522-726X ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/ccd.26429 ↗
- Languages:
- English
- ISSNs:
- 1522-1946
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3092.992000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 759.xml