Standardization of microparticle enumeration across different flow cytometry platforms: results of a multicenter collaborative workshop. (26th November 2016)
- Record Type:
- Journal Article
- Title:
- Standardization of microparticle enumeration across different flow cytometry platforms: results of a multicenter collaborative workshop. (26th November 2016)
- Main Title:
- Standardization of microparticle enumeration across different flow cytometry platforms: results of a multicenter collaborative workshop
- Authors:
- Cointe, S.
Judicone, C.
Robert, S.
Mooberry, M. J.
Poncelet, P.
Wauben, M.
Nieuwland, R.
Key, N. S.
Dignat‐George, F.
Lacroix, R. - Abstract:
- Abstract : Essentials The clinical enumeration of microparticles (MPs) is hampered by a lack of standardization. A new strategy to standardize MP counts by flow cytometry was evaluated in a multicenter study. No difference was found between instruments using forward or side scatter as the trigger parameter. This study demonstrated that beads can be used as a standardization tool for MPs. Click to hear the ISTH Academy's webinar on microvesicles Summary: Background: Microparticles (MPs) are extracellular vesicles resulting from the budding of cellular membranes that have a high potential as emergent biomarkers; however, their clinical relevance is hampered by methodological enumeration concerns and a lack of standardization. Flow cytometry (FCM) remains the most commonly used technique with the best capability to determine the cellular origin of single MPs. However, instruments behave variably depending on which scatter parameter (forward (FSC) or side scatter (SSC)) provides the best resolution to discriminate submicron particles. To overcome this problem, a new approach, based on two sets of selected beads adapted to FSC or SSC‐optimized instruments, was recently proposed to reproducibly enumerate platelet‐derived MP counts among instruments with different optical systems. Objective: The objective was to evaluate this strategy in an international workshop that included 44 laboratories accounting for 52 cytometers of 14 types. Methods/Results: Using resolution capability andAbstract : Essentials The clinical enumeration of microparticles (MPs) is hampered by a lack of standardization. A new strategy to standardize MP counts by flow cytometry was evaluated in a multicenter study. No difference was found between instruments using forward or side scatter as the trigger parameter. This study demonstrated that beads can be used as a standardization tool for MPs. Click to hear the ISTH Academy's webinar on microvesicles Summary: Background: Microparticles (MPs) are extracellular vesicles resulting from the budding of cellular membranes that have a high potential as emergent biomarkers; however, their clinical relevance is hampered by methodological enumeration concerns and a lack of standardization. Flow cytometry (FCM) remains the most commonly used technique with the best capability to determine the cellular origin of single MPs. However, instruments behave variably depending on which scatter parameter (forward (FSC) or side scatter (SSC)) provides the best resolution to discriminate submicron particles. To overcome this problem, a new approach, based on two sets of selected beads adapted to FSC or SSC‐optimized instruments, was recently proposed to reproducibly enumerate platelet‐derived MP counts among instruments with different optical systems. Objective: The objective was to evaluate this strategy in an international workshop that included 44 laboratories accounting for 52 cytometers of 14 types. Methods/Results: Using resolution capability and background noise level as criteria to qualify the instruments, the standardization strategy proved to be compatible with 85% (44/52) of instruments. All instruments correctly ranked the platelet MP (PMP) levels of two platelet‐free plasma samples. The inter‐laboratory variability of PMP counts was 37% and 28% for each sample. No difference was found between instruments using forward or side‐scattered light as the relative sizing parameter. Conclusions: Despite remaining limitations, this study is the first to demonstrate a real potential of bead‐based strategies for standardization of MP enumeration across different FCM platforms. Additional standardization efforts are still mandatory to evaluate MPs' clinical relevance at a multicenter level. … (more)
- Is Part Of:
- Journal of thrombosis and haemostasis. Volume 15:Number 1(2017)
- Journal:
- Journal of thrombosis and haemostasis
- Issue:
- Volume 15:Number 1(2017)
- Issue Display:
- Volume 15, Issue 1 (2017)
- Year:
- 2017
- Volume:
- 15
- Issue:
- 1
- Issue Sort Value:
- 2017-0015-0001-0000
- Page Start:
- 187
- Page End:
- 193
- Publication Date:
- 2016-11-26
- Subjects:
- cell‐derived microparticles -- extracellular vesicles -- flow cytometry -- multicenter study -- standardization
Thrombosis -- Periodicals
Hemostasis -- Periodicals
Blood coagulation disorders -- Periodicals
616.1 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1538-7836 ↗
http://www.blackwellpublishing.com/journals/jth ↗
https://www.sciencedirect.com/journal/journal-of-thrombosis-and-haemostasis ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1111/jth.13514 ↗
- Languages:
- English
- ISSNs:
- 1538-7933
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5069.345000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 1623.xml