A novel magnetic-driven tissue retraction device for minimally invasive surgery. Issue 1 (2nd January 2017)
- Record Type:
- Journal Article
- Title:
- A novel magnetic-driven tissue retraction device for minimally invasive surgery. Issue 1 (2nd January 2017)
- Main Title:
- A novel magnetic-driven tissue retraction device for minimally invasive surgery
- Authors:
- Brancadoro, Margherita
Tognarelli, Selene
Ciuti, Gastone
Menciassi, Arianna - Abstract:
- Abstract: Introduction: The purpose of this work is to design and validate an innovative magnetic-based device for tissue retraction for minimally invasive surgery. Material and methods: An intra-abdominal magnetic module is coupled with an extracorporeal system for establishing a stable attraction, and consequently a reliable tissue retraction. Once the retractor has been inserted into the abdomen, tissue retraction is not constrained by a fixed access port, thus guaranteeing a more flexible, safer and less invasive operation. The intra-abdominal unit is composed of an axial permanent magnet attached to a stainless-steel non-magnetic alligator clip by a traditional suturing thread. A miniaturized mechanism to adjust the length of the suturing thread for lengthening or shortening the distance between the tissue grasper and the internal magnetic unit is included. A multiphysics approach assured a dedicated design that thoroughly fulfills anatomical, physiological and engineering constraints. Results: System functionalities were demonstrated both in in-vitro and ex-vivo conditions, reaching good results and promising outcomes in terms of effectiveness and maneuverability. The retractor was successfully tested in an animal model, carrying out a whole retraction procedure. Conclusion: The proposed retraction system resulted to be intuitive, reliable, robust and easy to use, representing a suitable device for MIS procedures.
- Is Part Of:
- Minimally invasive therapy & allied technologies. Volume 26:Issue 1(2017)
- Journal:
- Minimally invasive therapy & allied technologies
- Issue:
- Volume 26:Issue 1(2017)
- Issue Display:
- Volume 26, Issue 1 (2017)
- Year:
- 2017
- Volume:
- 26
- Issue:
- 1
- Issue Sort Value:
- 2017-0026-0001-0000
- Page Start:
- 7
- Page End:
- 14
- Publication Date:
- 2017-01-02
- Subjects:
- Magnetic surgical device -- FEM simulation -- MIS -- retraction -- surgical tool
Endoscopy -- Periodicals
Interventional radiology -- Periodicals
Endoscopic surgery -- Periodicals
617.05 - Journal URLs:
- http://informahealthcare.com/loi/mit ↗
http://informahealthcare.com ↗ - DOI:
- 10.1080/13645706.2016.1234488 ↗
- Languages:
- English
- ISSNs:
- 1364-5706
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5797.714000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 2253.xml