Can EMS Providers Provide Appropriate Tidal Volumes in a Simulated Adult-sized Patient with a Pediatric-sized Bag-Valve-Mask?. (2nd January 2017)
- Record Type:
- Journal Article
- Title:
- Can EMS Providers Provide Appropriate Tidal Volumes in a Simulated Adult-sized Patient with a Pediatric-sized Bag-Valve-Mask?. (2nd January 2017)
- Main Title:
- Can EMS Providers Provide Appropriate Tidal Volumes in a Simulated Adult-sized Patient with a Pediatric-sized Bag-Valve-Mask?
- Authors:
- Siegler, Jeffrey
Kroll, Melissa
Wojcik, Susan
Moy, Hawnwan Philip - Abstract:
- Abstract: Introduction: In the prehospital setting, Emergency Medical Services (EMS) professionals rely on providing positive pressure ventilation with a bag-valve-mask (BVM). Multiple emergency medicine and critical care studies have shown that lung-protective ventilation protocols reduce morbidity and mortality. Our primary objective was to determine if a group of EMS professionals could provide ventilations with a smaller BVM that would be sufficient to ventilate patients. Secondary objectives included 1) if the pediatric bag provided volumes similar to lung-protective ventilation in the hospital setting and 2) compare volumes provided to the patient depending on the type of airway (mask, King tube, and intubation).Methods: Using a patient simulator of a head and thorax that was able to record respiratory rate, tidal volume, peak pressure, and minute volume via a laptop computer, participants were asked to ventilate the simulator during six 1-minute ventilation tests. The first scenario was BVM ventilation with an oropharyngeal airway in place ventilating with both an adult- and pediatric-sized BVM, the second scenario had a supraglottic airway and both bags, and the third scenario had an endotracheal tube and both bags. Participants were enrolled in convenience manner while they were on-duty and the research staff was able to travel to their stations. Prior to enrolling, participants were not given any additional training on ventilation skills.Results: We enrolled 50Abstract: Introduction: In the prehospital setting, Emergency Medical Services (EMS) professionals rely on providing positive pressure ventilation with a bag-valve-mask (BVM). Multiple emergency medicine and critical care studies have shown that lung-protective ventilation protocols reduce morbidity and mortality. Our primary objective was to determine if a group of EMS professionals could provide ventilations with a smaller BVM that would be sufficient to ventilate patients. Secondary objectives included 1) if the pediatric bag provided volumes similar to lung-protective ventilation in the hospital setting and 2) compare volumes provided to the patient depending on the type of airway (mask, King tube, and intubation).Methods: Using a patient simulator of a head and thorax that was able to record respiratory rate, tidal volume, peak pressure, and minute volume via a laptop computer, participants were asked to ventilate the simulator during six 1-minute ventilation tests. The first scenario was BVM ventilation with an oropharyngeal airway in place ventilating with both an adult- and pediatric-sized BVM, the second scenario had a supraglottic airway and both bags, and the third scenario had an endotracheal tube and both bags. Participants were enrolled in convenience manner while they were on-duty and the research staff was able to travel to their stations. Prior to enrolling, participants were not given any additional training on ventilation skills.Results: We enrolled 50 providers from a large, busy, urban fire-based EMS agency with 14.96 (SD = 9.92) mean years of experience. Only 1.5% of all breaths delivered with the pediatric BVM during the ventilation scenarios were below the recommended tidal volume. A greater percentage of breaths delivered in the recommended range occurred when the pediatric BVM was used (17.5% vs 5.1%, p < 0.001). Median volumes for each scenario were 570.5mL, 664.0mL, 663.0mL for the pediatric BMV and 796.0mL, 994.5mL, 981.5mL for the adult BVM. In all three categories of airway devices, the pediatric BVM provided lower median tidal volumes (p < 0.001).Conclusion: The study suggests that ventilating an adult patient is possible with a smaller, pediatric-sized BVM. The tidal volumes recorded with the pediatric BVM were more consistent with lung-protective ventilation volumes. … (more)
- Is Part Of:
- Prehospital emergency care. Volume 21:Number 1(2017)
- Journal:
- Prehospital emergency care
- Issue:
- Volume 21:Number 1(2017)
- Issue Display:
- Volume 21, Issue 1 (2017)
- Year:
- 2017
- Volume:
- 21
- Issue:
- 1
- Issue Sort Value:
- 2017-0021-0001-0000
- Page Start:
- 74
- Page End:
- 78
- Publication Date:
- 2017-01-02
- Subjects:
- BVM -- ventilation -- tidal volume -- simulation -- lung protection
362.18 - Journal URLs:
- http://informahealthcare.com/loi/pec ↗
http://informahealthcare.com ↗ - DOI:
- 10.1080/10903127.2016.1227003 ↗
- Languages:
- English
- ISSNs:
- 1090-3127
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6605.917000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 2071.xml