- Self‐navigation under non‐steady‐state conditions: Cardiac and respiratory self‐gating of inversion recovery snapshot FLASH acquisitions in mice. Issue 6 (7th January 2016)
- Main Title:
- Self‐navigation under non‐steady‐state conditions: Cardiac and respiratory self‐gating of inversion recovery snapshot FLASH acquisitions in mice
- Authors:
- Winter, Patrick
Kampf, Thomas
Helluy, Xavier
Gutjahr, Fabian T.
Meyer, Cord B.
Bauer, Wolfgang R.
Jakob, Peter M.
Herold, Volker - Abstract:
- Abstract : Purpose: An algorithm is presented to enable cardiac and respiratory self‐gating in combination with Inversion Recovery Look‐Locker read‐outs. Methods: A radial inversion recovery snapshot FLASH sequence was adapted for retrospective cardiac T 1 measurements in mice. Cardiac and respiratory data were extracted from the k‐space center of radial projections and an adapted method for retrospective cardiac synchronization is introduced. Electrocardiogram (ECG) data was acquired concurrently for validation of the proposed self‐gating technique. T 1 maps generated by the proposed technique were compared with maps reconstructed with the ECG reference. Results: Respiratory gating and cardiac trigger points could be obtained for the whole time course of the relaxation dynamic and correlate very well to the ECG signal. T 1 maps reconstructed with the self‐gating technique are in very good agreement with maps reconstructed with the external reference. Conclusion: The proposed method extends "wireless" cardiac MRI to non‐steady‐state inversion recovery measurements. T 1 maps were generated with a quality comparable to ECG based reconstructions. As the method does not rely on an ECG trigger signal it provides easier animal handling. Magn Reson Med 76:1887–1894, 2016. © 2016 International Society for Magnetic Resonance in Medicine
- Is Part Of:
- Magnetic resonance in medicine. Volume 76:Issue 6(2016)
- Journal:
- Magnetic resonance in medicine
- Issue:
- Volume 76:Issue 6(2016)
- Issue Display:
- Volume 76, Issue 6 (2016)
- Year:
- 2016
- Volume:
- 76
- Issue:
- 6
- Issue Sort Value:
- 2016-0076-0006-0000
- Page Start:
- 1887
- Page End:
- 1894
- Publication Date:
- 2016-01-07
- Subjects:
- T1 mapping -- self‐gating -- non‐steady‐state -- cardiac -- retrospective triggering -- cine -- Look‐Locker
Nuclear magnetic resonance -- Periodicals
Electron paramagnetic resonance -- Periodicals
616.07548 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1522-2594 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/mrm.26068 ↗
- Languages:
- English
- ISSNs:
- 0740-3194
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5337.798000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 2097.xml