Exploration and comparison of in vitro eye irritation tests with the ISO standard in vivo rabbit test for the evaluation of the ocular irritancy of contact lenses. (December 2016)
- Record Type:
- Journal Article
- Title:
- Exploration and comparison of in vitro eye irritation tests with the ISO standard in vivo rabbit test for the evaluation of the ocular irritancy of contact lenses. (December 2016)
- Main Title:
- Exploration and comparison of in vitro eye irritation tests with the ISO standard in vivo rabbit test for the evaluation of the ocular irritancy of contact lenses
- Authors:
- Yun, Jun-Won
Hailian, Quan
Na, Yirang
Kang, Byeong-Cheol
Yoon, Jung-Hee
Cho, Eun-Young
Lee, Miri
Kim, Da-eun
Bae, SeungJin
Seok, Seung Hyeok
Lim, Kyung-Min - Abstract:
- Abstract: In an effort to explore the use of alternative methods to animal testing for the evaluation of the ocular irritancy of medical devices, we evaluated representative contact lenses with the bovine corneal opacity and permeability test (BCOP) and an in vitro eye irritation test using the three-dimensionally-reconstructed human corneal epithelium (RhCE) models, EpiOcular™ and MCTT HCE™. In addition, we compared the obtained results with the ISO standard in vivo rabbit eye irritation test (ISO10993-10). Along with the positive controls (benzalkonium chloride, BAK, 0.02, 0.2, and 1%), the extracts of 4 representative contact lenses (soft, disposable, hard, and colored lenses) and 2 reference lenses (dye-eluting and BAK-coated lenses) were tested. All the lenses, except for the BAK-coated lens, were determined non-irritants in all test methods, while the positive controls yielded relevant results. More importantly, BCOP, EpiOcular™, and MCTT HCE™ yielded a consistent decision for all the tested samples, with the exception of 0.2% BAK in BCOP, for which no prediction could be made. Overall, all the in vitro tests correlated well with the in vivo rabbit eye irritation test, and furthermore, the combination of in vitro tests as a tiered testing strategy was able to produce results similar to those seen in vivo . These observations suggest that such methods can be used as alternative assays to replace the conventional in vivo test method in the evaluation of the ocularAbstract: In an effort to explore the use of alternative methods to animal testing for the evaluation of the ocular irritancy of medical devices, we evaluated representative contact lenses with the bovine corneal opacity and permeability test (BCOP) and an in vitro eye irritation test using the three-dimensionally-reconstructed human corneal epithelium (RhCE) models, EpiOcular™ and MCTT HCE™. In addition, we compared the obtained results with the ISO standard in vivo rabbit eye irritation test (ISO10993-10). Along with the positive controls (benzalkonium chloride, BAK, 0.02, 0.2, and 1%), the extracts of 4 representative contact lenses (soft, disposable, hard, and colored lenses) and 2 reference lenses (dye-eluting and BAK-coated lenses) were tested. All the lenses, except for the BAK-coated lens, were determined non-irritants in all test methods, while the positive controls yielded relevant results. More importantly, BCOP, EpiOcular™, and MCTT HCE™ yielded a consistent decision for all the tested samples, with the exception of 0.2% BAK in BCOP, for which no prediction could be made. Overall, all the in vitro tests correlated well with the in vivo rabbit eye irritation test, and furthermore, the combination of in vitro tests as a tiered testing strategy was able to produce results similar to those seen in vivo . These observations suggest that such methods can be used as alternative assays to replace the conventional in vivo test method in the evaluation of the ocular irritancy of ophthalmic medical devices, although further study is necessary. Highlights: Here, we did in vitro eye irritation tests using RhCE and BCOP for contact lens extracts. Positive controls and 4 representative contact lens were tested in vitro and in vivo . In vitro tests produced comparable results to in vivo and combination of BCOP and RhCE yielded the most reliable results. … (more)
- Is Part Of:
- Toxicology in vitro. Volume 37(2017)
- Journal:
- Toxicology in vitro
- Issue:
- Volume 37(2017)
- Issue Display:
- Volume 37, Issue 2017 (2017)
- Year:
- 2017
- Volume:
- 37
- Issue:
- 2017
- Issue Sort Value:
- 2017-0037-2017-0000
- Page Start:
- 79
- Page End:
- 87
- Publication Date:
- 2016-12
- Subjects:
- Eye irritation -- Animal alternative -- Reconstructed human corneal epithelium, RhCE model -- BCOP -- Contact lens -- ISO10993-10 -- Medical device
Toxicity testing -- In vitro -- Periodicals
Toxicology -- Periodicals
615.9 - Journal URLs:
- http://www.sciencedirect.com/science/journal/08872333 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.tiv.2016.09.001 ↗
- Languages:
- English
- ISSNs:
- 0887-2333
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 8873.043400
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 831.xml