A validated methodology for patient specific computational modeling of self-expandable transcatheter aortic valve implantation. Issue 13 (6th September 2016)
- Record Type:
- Journal Article
- Title:
- A validated methodology for patient specific computational modeling of self-expandable transcatheter aortic valve implantation. Issue 13 (6th September 2016)
- Main Title:
- A validated methodology for patient specific computational modeling of self-expandable transcatheter aortic valve implantation
- Authors:
- Bosmans, Bart
Famaey, Nele
Verhoelst, Eva
Bosmans, Johan
Vander Sloten, Jos - Abstract:
- Abstract: Leakage of blood alongside the implant is a relatively frequent and life-limiting complication after transcatheter aortic valve implantation. The aim of this study is to develop and validate a workflow to simulate the implantation prior to the intervention. Based on the simulation outcome, the amount of leakage is estimated in order to evaluate the risk of a severe complication. A finite element model of the stent implantation in 10 patients was created based on a pre-operative computed tomography scan. All 10 patients also received a follow-up computed tomography scan, after the implantation. This scan was used to extract the deformed geometry of the stent and the position of the calcifications for validation of the simulation results. The maximal average perimeter difference between the simulated stent and the post-operative stent is 2.9±2.1 mm, and occurs at the bottom of the device. The sensitivity of the simulation to the soft tissue material parameters and aortic root wall thickness was tested. The maximal diameter deviation of 6% occurred when the thickness of the aortic root was doubled. The result of the leakage analysis based on the distance between the simulated stent and the surrounding aortic root corresponded well when no regurgitation was observed. The developed tools have the potential to reduce the occurrence and severity of leakage by providing the clinician with additional information prior to the intervention. The simulated geometry andAbstract: Leakage of blood alongside the implant is a relatively frequent and life-limiting complication after transcatheter aortic valve implantation. The aim of this study is to develop and validate a workflow to simulate the implantation prior to the intervention. Based on the simulation outcome, the amount of leakage is estimated in order to evaluate the risk of a severe complication. A finite element model of the stent implantation in 10 patients was created based on a pre-operative computed tomography scan. All 10 patients also received a follow-up computed tomography scan, after the implantation. This scan was used to extract the deformed geometry of the stent and the position of the calcifications for validation of the simulation results. The maximal average perimeter difference between the simulated stent and the post-operative stent is 2.9±2.1 mm, and occurs at the bottom of the device. The sensitivity of the simulation to the soft tissue material parameters and aortic root wall thickness was tested. The maximal diameter deviation of 6% occurred when the thickness of the aortic root was doubled. The result of the leakage analysis based on the distance between the simulated stent and the surrounding aortic root corresponded well when no regurgitation was observed. The developed tools have the potential to reduce the occurrence and severity of leakage by providing the clinician with additional information prior to the intervention. The simulated geometry and estimated leakage can help decide on the best implant type, size and position before treatment. … (more)
- Is Part Of:
- Journal of biomechanics. Volume 49:Issue 13(2016)
- Journal:
- Journal of biomechanics
- Issue:
- Volume 49:Issue 13(2016)
- Issue Display:
- Volume 49, Issue 13 (2016)
- Year:
- 2016
- Volume:
- 49
- Issue:
- 13
- Issue Sort Value:
- 2016-0049-0013-0000
- Page Start:
- 2824
- Page End:
- 2830
- Publication Date:
- 2016-09-06
- Subjects:
- Patient-specific -- Finite element simulation -- Transcatheter aortic valve implantation -- Computed tomography
Animal mechanics -- Periodicals
Biomechanics -- Periodicals
Biomechanics -- Periodicals
Mécanique animale -- Périodiques
Biomécanique -- Périodiques
Electronic journals
571.4305 - Journal URLs:
- http://www.sciencedirect.com/science/journal/00219290 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/00219290 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/00219290 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.jbiomech.2016.06.024 ↗
- Languages:
- English
- ISSNs:
- 0021-9290
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4953.600000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 1860.xml