[OP.6B.03] WRIST BLOOD PRESSURE SELF-MEASUREMENT AT HOME IS UNRELIABLE. RESULTS FROM THE POPULATION-BASED "GROWING OLD WITH LESS DISEASE ENHANCING NEUROFUNCTIONS" (GOLDEN) STUDY. (September 2016)
- Record Type:
- Journal Article
- Title:
- [OP.6B.03] WRIST BLOOD PRESSURE SELF-MEASUREMENT AT HOME IS UNRELIABLE. RESULTS FROM THE POPULATION-BASED "GROWING OLD WITH LESS DISEASE ENHANCING NEUROFUNCTIONS" (GOLDEN) STUDY. (September 2016)
- Main Title:
- [OP.6B.03] WRIST BLOOD PRESSURE SELF-MEASUREMENT AT HOME IS UNRELIABLE. RESULTS FROM THE POPULATION-BASED "GROWING OLD WITH LESS DISEASE ENHANCING NEUROFUNCTIONS" (GOLDEN) STUDY
- Authors:
- Tikhonoff, V.
Casiglia, E.
Palatini, P. - Abstract:
- Abstract : Objective: To investigate the reliability of self-measured home blood pressure (BP) obtained with wrist devices. Design and method: Office and home upper-arm and wrist BP was obtained in 597 representative subjects from general population using two validated devices. Parti-cipants received an individual 20-min training in the office and were asked to repeat the upper-arm and wrist measurements at home for 7 consecutive days using the same procedures. In particular, they were recommended to keep the wrist at heart level during wrist BP measurement. The difference between upper-arm BP and wrist BP measured by the observer in the office was calculated for each individual and compared with the 7-day averaged self-measured upper-arm - wrist BP difference. Anthropometric measures and a complete cognitive assessment were also obtained. Results: In the office, under observer supervision, after adjustment for confounders, wrist BP was lower than upper-arm BP (systolic −2.4%, p < 0.0001; diastolic −0.6%, NS), while at home BP was higher at wrist than at upper-arm (systolic +5.9%, p < 0.0001; diastolic +5.6%, p < 0.0001). Both systolic and diastolic upper-arm - wrist BP differences obtained in office significantly differed from the upper-arm - wrist differences obtained by the participant at home (systolic 3.0 ± 13.7 vs. −6.7 ± 0.1 mmHg, p < 0.0001; diastolic 0.6 ± 9.4 vs. −4.0 ± 6.6 mmHg, p < 0.0001). Office (r = 0.37, p < 0.0001) but not home (r = −0.05, p = 0.9) wristAbstract : Objective: To investigate the reliability of self-measured home blood pressure (BP) obtained with wrist devices. Design and method: Office and home upper-arm and wrist BP was obtained in 597 representative subjects from general population using two validated devices. Parti-cipants received an individual 20-min training in the office and were asked to repeat the upper-arm and wrist measurements at home for 7 consecutive days using the same procedures. In particular, they were recommended to keep the wrist at heart level during wrist BP measurement. The difference between upper-arm BP and wrist BP measured by the observer in the office was calculated for each individual and compared with the 7-day averaged self-measured upper-arm - wrist BP difference. Anthropometric measures and a complete cognitive assessment were also obtained. Results: In the office, under observer supervision, after adjustment for confounders, wrist BP was lower than upper-arm BP (systolic −2.4%, p < 0.0001; diastolic −0.6%, NS), while at home BP was higher at wrist than at upper-arm (systolic +5.9%, p < 0.0001; diastolic +5.6%, p < 0.0001). Both systolic and diastolic upper-arm - wrist BP differences obtained in office significantly differed from the upper-arm - wrist differences obtained by the participant at home (systolic 3.0 ± 13.7 vs. −6.7 ± 0.1 mmHg, p < 0.0001; diastolic 0.6 ± 9.4 vs. −4.0 ± 6.6 mmHg, p < 0.0001). Office (r = 0.37, p < 0.0001) but not home (r = −0.05, p = 0.9) wrist systolic BP correlated with office BP, the two regression coefficients differing significantly (p < 0.0001). Similar results were obtained for diastolic BP. In multivariable regression analyses, using the difference between the upper-arm - wrist BP discrepancies obtained in the office and at home as the dependent variable, forearm length (p < 0.06 for systolic, p < 0.04 for diastolic) and praxic abilities (p < 0.008, p < 0.0001, respectively) were independent determinants of the office-home difference. Conclusions: These data indicate that in spite of adequate training wrist BP measurement at home is unreliable. This is likely due to a cognitive (particularly praxic) deficit leading to incorrect forearm position, a problem that, for hydraulic reasons, is magnified in people with longer forearm. Wrist BP measurement should be discouraged chiefly in subjects presumed to have impaired cognitive function. … (more)
- Is Part Of:
- Journal of hypertension. Volume 34:(2016) Supplement 2
- Journal:
- Journal of hypertension
- Issue:
- Volume 34:(2016) Supplement 2
- Issue Display:
- Volume 34, Issue 2 (2016)
- Year:
- 2016
- Volume:
- 34
- Issue:
- 2
- Issue Sort Value:
- 2016-0034-0002-0000
- Page Start:
- Page End:
- Publication Date:
- 2016-09
- Subjects:
- Hypertension -- Periodicals
Hypertension -- Periodicals
616.132005 - Journal URLs:
- http://firstsearch.oclc.org ↗
http://journals.lww.com/jhypertension/pages/default.aspx ↗
http://ovidsp.ovid.com/ovidweb.cgi?T=JS&NEWS=n&CSC=Y&PAGE=toc&D=yrovft&AN=00004872-000000000-00000 ↗
http://www.jhypertension.com/ ↗
http://journals.lww.com/pages/default.aspx ↗ - DOI:
- 10.1097/01.hjh.0000491516.19347.ac ↗
- Languages:
- English
- ISSNs:
- 1473-5598
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5004.510000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 1507.xml