Fracture at the Modular Junction of a Cementless Revision Hip System: A Case Report. Issue 2 (April 2016)
- Record Type:
- Journal Article
- Title:
- Fracture at the Modular Junction of a Cementless Revision Hip System: A Case Report. Issue 2 (April 2016)
- Main Title:
- Fracture at the Modular Junction of a Cementless Revision Hip System
- Authors:
- Duncan, Stephen T.
Hayes, Christopher B.
Nunley, Ryan M. - Abstract:
- Abstract : Case: Multiple cases of modular femoral component failure involving fatigue at the modular stem junction have been reported following revision total hip arthroplasty. In the present report, we describe what we believe to be the first case of catastrophic failure involving the Biomet Arcos revision system. The failure involved a young patient with an elevated body mass index and proximal bone deficiency. Conclusion: The possibility of this failure mechanism must be weighed against the potential advantages of modular cementless stems when planning for complex hip arthroplasty, especially in the young, active patient.
- Is Part Of:
- JBJS case connector. Volume 6:Issue 2(2016)
- Journal:
- JBJS case connector
- Issue:
- Volume 6:Issue 2(2016)
- Issue Display:
- Volume 6, Issue 2 (2016)
- Year:
- 2016
- Volume:
- 6
- Issue:
- 2
- Issue Sort Value:
- 2016-0006-0002-0000
- Page Start:
- Page End:
- Publication Date:
- 2016-04
- Subjects:
- Orthopedic surgery -- Case studies -- Periodicals
Orthopedic surgery -- Periodicals
Orthopedics
Orthopedic Procedures
Orthopedic surgery
Periodicals
Electronic journals
Case studies
Periodicals
616.7 - Journal URLs:
- http://caseconnector.jbjs.org/ ↗
http://journals.lww.com/pages/default.aspx ↗ - DOI:
- 10.2106/JBJS.CC.15.00089 ↗
- Languages:
- English
- ISSNs:
- 2160-3251
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4663.437600
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 2053.xml