A Point‐of‐Care Raman Spectroscopy–Based Device for the Diagnosis of Gout and Pseudogout: Comparison With the Clinical Standard Microscopy. Issue 7 (July 2016)
- Record Type:
- Journal Article
- Title:
- A Point‐of‐Care Raman Spectroscopy–Based Device for the Diagnosis of Gout and Pseudogout: Comparison With the Clinical Standard Microscopy. Issue 7 (July 2016)
- Main Title:
- A Point‐of‐Care Raman Spectroscopy–Based Device for the Diagnosis of Gout and Pseudogout: Comparison With the Clinical Standard Microscopy
- Authors:
- Li, Bolan
Singer, Nora G.
Yeni, Yener N.
Haggins, Donard G.
Barnboym, Emma
Oravec, Daniel
Lewis, Steven
Akkus, Ozan - Abstract:
- Abstract : Objective: To demonstrate the usefulness of a novel medical device based on Raman spectroscopy for the rapid point‐of‐care diagnosis of gout and pseudogout. Methods: A shoebox‐sized point‐of‐care Raman spectroscopy (POCRS) device was developed for use in the diagnosis of gout and pseudogout. The device included a disposable syringe microfiltration kit to collect arthropathic crystals from synovial fluid and a customized automated Raman spectroscopy system to chemically identify crystal species. Diagnosis according to the findings of POCRS was compared with the clinical standard diagnosis based on compensated polarized light microscopy (CPLM) of synovial fluid aspirates collected from symptomatic patients (n = 174). Kappa coefficients were used to measure the agreement between POCRS and CPLM findings. Results: Overall, POCRS and CPLM results were consistent in 89.7% of samples (156 of 174). For the diagnosis of gout, the kappa coefficient for POCRS and CPLM was 0.84 (95% confidence interval [95% CI] 0.75–0.94). For the diagnosis of pseudogout, the kappa coefficient for POCRS and CPLM was 0.61 (95% CI 0.42–0.81). Conclusion: Kappa coefficients indicated that there was excellent agreement between POCRS and CPLM for the diagnosis of gout, with good agreement for the diagnosis of pseudogout. The POCRS device holds the potential to standardize and expedite the time to clinical diagnosis of gout and pseudogout, especially in settings where certified operators trained forAbstract : Objective: To demonstrate the usefulness of a novel medical device based on Raman spectroscopy for the rapid point‐of‐care diagnosis of gout and pseudogout. Methods: A shoebox‐sized point‐of‐care Raman spectroscopy (POCRS) device was developed for use in the diagnosis of gout and pseudogout. The device included a disposable syringe microfiltration kit to collect arthropathic crystals from synovial fluid and a customized automated Raman spectroscopy system to chemically identify crystal species. Diagnosis according to the findings of POCRS was compared with the clinical standard diagnosis based on compensated polarized light microscopy (CPLM) of synovial fluid aspirates collected from symptomatic patients (n = 174). Kappa coefficients were used to measure the agreement between POCRS and CPLM findings. Results: Overall, POCRS and CPLM results were consistent in 89.7% of samples (156 of 174). For the diagnosis of gout, the kappa coefficient for POCRS and CPLM was 0.84 (95% confidence interval [95% CI] 0.75–0.94). For the diagnosis of pseudogout, the kappa coefficient for POCRS and CPLM was 0.61 (95% CI 0.42–0.81). Conclusion: Kappa coefficients indicated that there was excellent agreement between POCRS and CPLM for the diagnosis of gout, with good agreement for the diagnosis of pseudogout. The POCRS device holds the potential to standardize and expedite the time to clinical diagnosis of gout and pseudogout, especially in settings where certified operators trained for CPLM analysis are not available. … (more)
- Is Part Of:
- Arthritis & rheumatology. Volume 68:Issue 7(2016)
- Journal:
- Arthritis & rheumatology
- Issue:
- Volume 68:Issue 7(2016)
- Issue Display:
- Volume 68, Issue 7 (2016)
- Year:
- 2016
- Volume:
- 68
- Issue:
- 7
- Issue Sort Value:
- 2016-0068-0007-0000
- Page Start:
- 1751
- Page End:
- 1757
- Publication Date:
- 2016-07
- Subjects:
- Arthritis -- Periodicals
Rheumatism -- Periodicals
616.72 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)2326-5205 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/art.39638 ↗
- Languages:
- English
- ISSNs:
- 2326-5191
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 1733.820000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 1121.xml