Design and development of a dedicated portable gamma camera system for intra-operative imaging. Issue 7 (July 2016)
- Record Type:
- Journal Article
- Title:
- Design and development of a dedicated portable gamma camera system for intra-operative imaging. Issue 7 (July 2016)
- Main Title:
- Design and development of a dedicated portable gamma camera system for intra-operative imaging
- Authors:
- Kaviani, Sanaz
Zeraatkar, Navid
Sajedi, Salar
Akbarzadeh, Afshin
Gorjizadeh, Nahid
Farahani, Mohammad Hossein
Teimourian, Behnoosh
Ghafarian, Pardis
Sabet, Hamid
Ay, Mohammad Reza - Abstract:
- Highlights: We developed a high performance portable gamma camera for radio-guided surgery. Results show 41 kcps for 20% count loss and the energy resolution was 20.6%. The developed system can be used for radiopharmaceutical guided surgery. The system sensitivity was measured as 142 cps/MBq on the collimator surface. Abstract: Purpose: We developed a high performance portable gamma camera platform dedicated to identification of sentinel lymph nodes (SLNs) and radio-guided surgery for cancer patients. In this work, we present the performance characteristics of SURGEOSIGHT-I, the first version of this platform that can intra-operatively provide high-resolution images of the surveyed areas. Methods: At the heart of this camera, there is a 43 × 43 array of pixelated sodium-activated cesium iodide (CsI(Na)) scintillation crystal with 1 × 1 mm 2 pixel size and 5 mm thickness coupled to a Hamamatsu H8500 flat-panel multi-anode (64 channels) photomultiplier tube. The probe is equipped with a hexagonal parallel-hole lead collimator with 1.2 mm holes. The detector, collimator, and the associated front-end electronics are encapsulated in a common housing referred to as head. Results: Our results show a count rate of ∼41 kcps for 20% count loss. The extrinsic energy resolution was measured as 20.6% at 140 keV. The spatial resolution and the sensitivity of the system on the collimator surface was measured as 2.2 mm and 142 cps/MBq, respectively. In addition, the integral andHighlights: We developed a high performance portable gamma camera for radio-guided surgery. Results show 41 kcps for 20% count loss and the energy resolution was 20.6%. The developed system can be used for radiopharmaceutical guided surgery. The system sensitivity was measured as 142 cps/MBq on the collimator surface. Abstract: Purpose: We developed a high performance portable gamma camera platform dedicated to identification of sentinel lymph nodes (SLNs) and radio-guided surgery for cancer patients. In this work, we present the performance characteristics of SURGEOSIGHT-I, the first version of this platform that can intra-operatively provide high-resolution images of the surveyed areas. Methods: At the heart of this camera, there is a 43 × 43 array of pixelated sodium-activated cesium iodide (CsI(Na)) scintillation crystal with 1 × 1 mm 2 pixel size and 5 mm thickness coupled to a Hamamatsu H8500 flat-panel multi-anode (64 channels) photomultiplier tube. The probe is equipped with a hexagonal parallel-hole lead collimator with 1.2 mm holes. The detector, collimator, and the associated front-end electronics are encapsulated in a common housing referred to as head. Results: Our results show a count rate of ∼41 kcps for 20% count loss. The extrinsic energy resolution was measured as 20.6% at 140 keV. The spatial resolution and the sensitivity of the system on the collimator surface was measured as 2.2 mm and 142 cps/MBq, respectively. In addition, the integral and differential uniformity, after uniformity correction, in useful field-of-view (UFOV) were measured 4.5% and 4.6%, respectively. Conclusions: This system can be used for a number of clinical applications including SLN biopsy and radiopharmaceutical-guided surgery. … (more)
- Is Part Of:
- Physica medica. Volume 32:Issue 7(2016)
- Journal:
- Physica medica
- Issue:
- Volume 32:Issue 7(2016)
- Issue Display:
- Volume 32, Issue 7 (2016)
- Year:
- 2016
- Volume:
- 32
- Issue:
- 7
- Issue Sort Value:
- 2016-0032-0007-0000
- Page Start:
- 889
- Page End:
- 897
- Publication Date:
- 2016-07
- Subjects:
- Sentinel lymph node (SLN) -- Lymphoscintigraphy -- SURGEOSIGHT-I -- Breast cancer -- Intra-operative gamma camera
Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2016.06.004 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 517.xml