The effectiveness of pulse oximetry sonification enhanced with tremolo and brightness for distinguishing clinically important oxygen saturation ranges: a laboratory study. (15th March 2016)
- Record Type:
- Journal Article
- Title:
- The effectiveness of pulse oximetry sonification enhanced with tremolo and brightness for distinguishing clinically important oxygen saturation ranges: a laboratory study. (15th March 2016)
- Main Title:
- The effectiveness of pulse oximetry sonification enhanced with tremolo and brightness for distinguishing clinically important oxygen saturation ranges: a laboratory study
- Authors:
- Paterson, E.
Sanderson, P. M.
Paterson, N. A. B.
Liu, D.
Loeb, R. G. - Abstract:
- Summary: Our study examined the effectiveness of pulse oximetry sonification enhanced with acoustic tremolo and brightness to help listeners differentiate clinically relevant oxygen saturation ranges. In a series of trials lasting 30 s each, 76 undergraduate participants identified final oxygen saturation range (Target: 100% to 97%; Low: 96% to 90%; Critical: 89% and below), and detected threshold transitions into and out of the target range using conventional sonification (n = 38) or enhanced sonification (n = 38). Median (IQR [range]) accuracy for range identification with the conventional sonification was 80 (70–85 [45–95])%, whereas with the enhanced sonification it was 100 (99–100 [80–100])%; p < 0.001. Accuracy for detecting threshold transitions with the conventional sonification was 60 (50–75 [30–95])%, but with the enhanced sonification it was 100 (95–100 [75–100]%; p < 0.001. Participants can identify clinically meaningful oxygen saturation ranges and detect threshold transitions more accurately with enhanced sonification than with conventional sonification.
- Is Part Of:
- Anaesthesia. Volume 71:Number 5(2016:May)
- Journal:
- Anaesthesia
- Issue:
- Volume 71:Number 5(2016:May)
- Issue Display:
- Volume 71, Issue 5 (2016)
- Year:
- 2016
- Volume:
- 71
- Issue:
- 5
- Issue Sort Value:
- 2016-0071-0005-0000
- Page Start:
- 565
- Page End:
- 572
- Publication Date:
- 2016-03-15
- Subjects:
- auditory alarms -- auditory displays -- patient monitoring -- pulse oximetry -- sonification
Anesthesia -- Periodicals
617.96 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1365-2044 ↗
http://onlinelibrary.wiley.com/ ↗
http://www.aagbi.org/publications ↗ - DOI:
- 10.1111/anae.13424 ↗
- Languages:
- English
- ISSNs:
- 0003-2409
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 0859.900000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 1799.xml