Combining situated Cognitive Engineering with a novel testing method in a case study comparing two infusion pump interfaces. (July 2016)
- Record Type:
- Journal Article
- Title:
- Combining situated Cognitive Engineering with a novel testing method in a case study comparing two infusion pump interfaces. (July 2016)
- Main Title:
- Combining situated Cognitive Engineering with a novel testing method in a case study comparing two infusion pump interfaces
- Authors:
- Schnittker, R.
Schmettow, M.
Verhoeven, F.
Schraagen, J.M.C. - Abstract:
- Abstract: We validated the usability of a new infusion pump interface designed with a situated Cognitive Engineering approach by comparing it to a reference interface using a novel testing method employing repeated measurements and process measures, in addition to traditional outcome measures. The sample consisted of 25 nurses who performed eight critical tasks three times. Performance measures consisted of number and type of errors, deviations from a pre-established normative path solution, task completion times, number of keystrokes, mental effort and preferences in use. Results showed that interaction with the new interface resulted in 18% fewer errors, 90% fewer normative path deviations, 42% lower task completion times, 40% fewer keystrokes, 39% lower mental effort and 76% more subjective preferences in use. These outcomes suggest that within the scope of this case study, combining the situated Cognitive Engineering approach with a novel testing method addresses various shortcomings of earlier testing methods. Highlights: Infusion pump interface designed with situated Cognitive Engineering was validated. Usability validation took place by comparison with a reference interface. A novel process tracing technique was used for analysis. Task performance increased with novel interface. Novel method was feasible for validating safety of medical devices.
- Is Part Of:
- Applied ergonomics. Volume 55(2016:Jul.)
- Journal:
- Applied ergonomics
- Issue:
- Volume 55(2016:Jul.)
- Issue Display:
- Volume 55 (2016)
- Year:
- 2016
- Volume:
- 55
- Issue Sort Value:
- 2016-0055-0000-0000
- Page Start:
- 16
- Page End:
- 26
- Publication Date:
- 2016-07
- Subjects:
- Medical device usability testing -- Infusion pump -- Human-machine interaction
Human engineering -- Periodicals
620.82 - Journal URLs:
- http://www.sciencedirect.com/science/journal/00036870 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.apergo.2016.01.004 ↗
- Languages:
- English
- ISSNs:
- 0003-6870
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 1572.500000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 2603.xml