Investigation of the response variability of ionization chambers for the standard transfer of SIR-Spheres®. (March 2016)
- Record Type:
- Journal Article
- Title:
- Investigation of the response variability of ionization chambers for the standard transfer of SIR-Spheres®. (March 2016)
- Main Title:
- Investigation of the response variability of ionization chambers for the standard transfer of SIR-Spheres®
- Authors:
- Thiam, C.
Bobin, C.
Lourenço, V.
Chisté, V.
Amiot, M.-N.
Mougeot, X.
Lacour, D.
Rigoulay, F.
Ferreux, L. - Abstract:
- Abstract: The present paper addresses the calibration of well-type ionization chambers (ICs) used at LNE-LNHB as standard transfer instruments to calibrate hospitals in the case of SIR-Spheres ® 90 Y resin microspheres (Sirtex, Australia). Developed for interventional oncology, this radiopharmaceutical is directly injected in the liver for cancer treatment via a selective internal radiation therapy. The present work was carried out in the framework of the European project "Metrology for molecular radiotherapy" (MetroMRT). As commonly performed in radionuclide metrology for radiopharmaceuticals, the objective is to ensure the metrological traceability of SIR-Spheres ® to hospitals. Preceding studies were focused on primary measurements of SIR-Spheres ® based on the TDCR (Triple to Double Coincidence Ratio) method, applied after the dissolution of the 90 Y-labeled resin microspheres. As 90 Y is a high-energy β - -emitter, the IC response strongly depends on the transport of electrons in the radioactive solution and surroundings (vial, chamber liners and materials). The variability of the IC-response due to the geometry dependence is investigated by means of measurements and Monte Carlo simulations in the case of a Vinten IC. The aim of the present study was also to propose a reliable uncertainty for ICs calibrations for the standard transfer of SIR-Spheres ® to hospitals. Highlights: LNE-LNHB ionization chambers (ICs) have been calibrated with SIR-Spheres ® . The variabilityAbstract: The present paper addresses the calibration of well-type ionization chambers (ICs) used at LNE-LNHB as standard transfer instruments to calibrate hospitals in the case of SIR-Spheres ® 90 Y resin microspheres (Sirtex, Australia). Developed for interventional oncology, this radiopharmaceutical is directly injected in the liver for cancer treatment via a selective internal radiation therapy. The present work was carried out in the framework of the European project "Metrology for molecular radiotherapy" (MetroMRT). As commonly performed in radionuclide metrology for radiopharmaceuticals, the objective is to ensure the metrological traceability of SIR-Spheres ® to hospitals. Preceding studies were focused on primary measurements of SIR-Spheres ® based on the TDCR (Triple to Double Coincidence Ratio) method, applied after the dissolution of the 90 Y-labeled resin microspheres. As 90 Y is a high-energy β - -emitter, the IC response strongly depends on the transport of electrons in the radioactive solution and surroundings (vial, chamber liners and materials). The variability of the IC-response due to the geometry dependence is investigated by means of measurements and Monte Carlo simulations in the case of a Vinten IC. The aim of the present study was also to propose a reliable uncertainty for ICs calibrations for the standard transfer of SIR-Spheres ® to hospitals. Highlights: LNE-LNHB ionization chambers (ICs) have been calibrated with SIR-Spheres ® . The variability of ICs responses due to the Sirtex vial geometry was investigated. Systematic underestimations of ICs responses for pure β - -emitters were also addressed. A strong correlation between ICs responses and empty-vials masses was observed. A reliable uncertainty for the standard transfer of SIR-Spheres ® is proposed. … (more)
- Is Part Of:
- Applied radiation and isotopes. Volume 109(2016:Mar.)
- Journal:
- Applied radiation and isotopes
- Issue:
- Volume 109(2016:Mar.)
- Issue Display:
- Volume 109 (2016)
- Year:
- 2016
- Volume:
- 109
- Issue Sort Value:
- 2016-0109-0000-0000
- Page Start:
- 231
- Page End:
- 235
- Publication Date:
- 2016-03
- Subjects:
- Radionuclide metrology -- SIR-Spheres® standardization -- 90Y-labeled resin microspheres -- Ionization chamber calibration -- Monte Carlo simulation
Radiology -- Periodicals
Radiation -- Industrial applications -- Periodicals
Nuclear chemistry -- Periodicals
Internet resource
Periodical
660.298 - Journal URLs:
- http://www.sciencedirect.com/science/journal/09698043 ↗
http://catalog.hathitrust.org/api/volumes/oclc/27456684.html ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.apradiso.2015.11.056 ↗
- Languages:
- English
- ISSNs:
- 0969-8043
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 1576.565000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 707.xml