Analysing the effects of cold, normal, and warm digits on transmittance pulse oximetry. (April 2016)
- Record Type:
- Journal Article
- Title:
- Analysing the effects of cold, normal, and warm digits on transmittance pulse oximetry. (April 2016)
- Main Title:
- Analysing the effects of cold, normal, and warm digits on transmittance pulse oximetry
- Authors:
- Khan, Musabbir
Pretty, Christopher G.
Amies, Alexander C.
Elliott, Rodney
Chiew, Yeong Shiong
Shaw, Geoffrey M.
Chase, J. Geoffrey - Abstract:
- Highlights: Investigated the effects of temperature on pulse oximeter signal quality in 20 healthy adults. Signal quality may vary at normal site temperature, and is subject dependant. Cold site temperature significantly degrades signal quality in general. Warm site temperature significantly improves signal quality in general. Warm site temperature (about 33 °C) should be maintained for reliable pulse oximetry. Abstract: Non-invasive estimation of arterial oxygen saturation (SpO2 ) and heart rate using pulse oximeters is widely used in hospitals. Pulse oximeters rely on photoplethysmographic (PPG) signals from a peripherally placed optical sensor. However, pulse oximeters can be less accurate if the sensor site is relatively cold. This research investigates the effects on PPG signal quality of local site temperatures for 20 healthy adult volunteers (24.5 ± 4.1 years of age). Raw PPG data, composed of Infrared (IR) and Red (RD) signals, was obtained from a transmittance finger probe using a custom pulse oximeter (PO) system. Three tests were performed with the subject's hand surface temperature maintained at baseline (29 ± 2 °C), cold (19 ± 2 °C), and warm (33 ± 2 °C) conditions. Median root mean square (RMS) of PPG signal during the Cold test dropped by 54.0% for IR and 30.6% for RD from the baseline values. In contrast, the PPG RMS increased by 64.4% and 60.2% for RD and IR, respectively, during the Warm test. Mean PPG pulse amplitudes decreased by 59.5% for IR and 46.1%Highlights: Investigated the effects of temperature on pulse oximeter signal quality in 20 healthy adults. Signal quality may vary at normal site temperature, and is subject dependant. Cold site temperature significantly degrades signal quality in general. Warm site temperature significantly improves signal quality in general. Warm site temperature (about 33 °C) should be maintained for reliable pulse oximetry. Abstract: Non-invasive estimation of arterial oxygen saturation (SpO2 ) and heart rate using pulse oximeters is widely used in hospitals. Pulse oximeters rely on photoplethysmographic (PPG) signals from a peripherally placed optical sensor. However, pulse oximeters can be less accurate if the sensor site is relatively cold. This research investigates the effects on PPG signal quality of local site temperatures for 20 healthy adult volunteers (24.5 ± 4.1 years of age). Raw PPG data, composed of Infrared (IR) and Red (RD) signals, was obtained from a transmittance finger probe using a custom pulse oximeter (PO) system. Three tests were performed with the subject's hand surface temperature maintained at baseline (29 ± 2 °C), cold (19 ± 2 °C), and warm (33 ± 2 °C) conditions. Median root mean square (RMS) of PPG signal during the Cold test dropped by 54.0% for IR and 30.6% for RD from the baseline values. In contrast, the PPG RMS increased by 64.4% and 60.2% for RD and IR, respectively, during the Warm test. Mean PPG pulse amplitudes decreased by 59.5% for IR and 46.1% for RD in the cold test when compared to baseline, but improved by 70.1% for IR and 59.0% for RD in the warm test. This improvement of up to 4× in signal quality during the warm condition was associated with a closer match (median difference of 1.5%) between the SpO2 values estimated by the PO system and a commercial pulse oximeter. The differences measured in RMS and mean amplitudes for the three tests were statistically significant ( p < 0.001). Overall, warm temperatures significantly improve PPG signal quality and SpO2 estimation accuracy. Sensor site temperature is recommended to be maintained near 33 °C for reliable transmittance pulse oximetry. … (more)
- Is Part Of:
- Biomedical signal processing and control. Volume 26(2016)
- Journal:
- Biomedical signal processing and control
- Issue:
- Volume 26(2016)
- Issue Display:
- Volume 26, Issue 2016 (2016)
- Year:
- 2016
- Volume:
- 26
- Issue:
- 2016
- Issue Sort Value:
- 2016-0026-2016-0000
- Page Start:
- 34
- Page End:
- 41
- Publication Date:
- 2016-04
- Subjects:
- Pulse oximetry -- Non-invasive -- Photoplethysmograph -- Arterial blood oxygen saturation -- Signal processing -- Sensor -- Thermocouple -- Perfusion -- Vasoconstriction -- Vasodilation
Signal processing -- Periodicals
Biomedical engineering -- Periodicals
Signal Processing, Computer-Assisted -- Periodicals
Image Processing, Computer-Assisted -- Periodicals
Biomedical Engineering -- Periodicals
610.28 - Journal URLs:
- http://www.sciencedirect.com/science/journal/17468094 ↗
http://www.elsevier.com/journals ↗
http://www.sciencedirect.com/science?_ob=PublicationURL&_tockey=%23TOC%2329675%232006%23999989998%23626449%23FLA%23&_cdi=29675&_pubType=J&_auth=y&_acct=C000045259&_version=1&_urlVersion=0&_userid=836873&md5=664b5cf9a57fc91971a17faf20c32ec1 ↗ - DOI:
- 10.1016/j.bspc.2015.12.006 ↗
- Languages:
- English
- ISSNs:
- 1746-8094
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 2087.880400
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 883.xml