Angular stable lateral plating is a valid alternative to conventional plate fixation in the proximal phalanx. A biomechanical study. Issue 5 (June 2015)
- Record Type:
- Journal Article
- Title:
- Angular stable lateral plating is a valid alternative to conventional plate fixation in the proximal phalanx. A biomechanical study. Issue 5 (June 2015)
- Main Title:
- Angular stable lateral plating is a valid alternative to conventional plate fixation in the proximal phalanx. A biomechanical study
- Authors:
- Shanmugam, R.
Ernst, M.
Stoffel, K.
Fischer, M.F.
Wahl, D.
Richards, R.G.
Gueorguiev, B. - Abstract:
- Abstract: Background: Dorsal plating is commonly used in proximal phalanx fractures but it bears the risk of interfering with the extensor apparatus. In this study, dorsal and lateral plating fixation methods are compared to assess biomechanical differences using conventional 1.5 mm non-locking plates and novel 1.3 mm lateral locking plates. Methods: Twenty-four fresh frozen human cadaveric proximal phalanges were equally divided into four groups. An osteotomy was set at the proximal metaphyseal–diaphyseal junction and fixed with either dorsal (group A) or lateral (group B) plating using a 1.5 mm non-locking plate, or lateral plating with a novel 1.3 mm locking plate with bicortical (group C) or unicortical (group D) screws. The specimens were loaded in axial, dorsovolar and mediolateral direction to assess fixation stiffness followed by a cyclic destructive test in dorsovolar loading direction. Findings: Axial stiffness was highest in group D (mean 321.02, SEM 21.47 N/mm) with a significant difference between groups D and B ( P = 0.033). Locking plates (groups C and D) were stiffer than non-locking plates under mediolateral loading ( P = 0.007), no significant differences were noted under dorsovolar loading. Furthermore, no significant differences were observed under cyclic loading to failure between any of the study groups. Interpretation: No considerable biomechanical advantage of using a conventional 1.5 mm dorsal non-locking plate was identified over the novel 1.3 mmAbstract: Background: Dorsal plating is commonly used in proximal phalanx fractures but it bears the risk of interfering with the extensor apparatus. In this study, dorsal and lateral plating fixation methods are compared to assess biomechanical differences using conventional 1.5 mm non-locking plates and novel 1.3 mm lateral locking plates. Methods: Twenty-four fresh frozen human cadaveric proximal phalanges were equally divided into four groups. An osteotomy was set at the proximal metaphyseal–diaphyseal junction and fixed with either dorsal (group A) or lateral (group B) plating using a 1.5 mm non-locking plate, or lateral plating with a novel 1.3 mm locking plate with bicortical (group C) or unicortical (group D) screws. The specimens were loaded in axial, dorsovolar and mediolateral direction to assess fixation stiffness followed by a cyclic destructive test in dorsovolar loading direction. Findings: Axial stiffness was highest in group D (mean 321.02, SEM 21.47 N/mm) with a significant difference between groups D and B ( P = 0.033). Locking plates (groups C and D) were stiffer than non-locking plates under mediolateral loading ( P = 0.007), no significant differences were noted under dorsovolar loading. Furthermore, no significant differences were observed under cyclic loading to failure between any of the study groups. Interpretation: No considerable biomechanical advantage of using a conventional 1.5 mm dorsal non-locking plate was identified over the novel 1.3 mm lateral locking plate in the treatment of proximal phalanx fractures. Since the novel low-profile plate is less disruptive to the extensor mechanism, it should be considered as a valid alternative. Highlights: Superior biomechanical performance of newly developed lateral locking plate No advantage of conventional dorsal over lateral plating using a locking implant Sufficient stability using unicortical screws for proximal phalanx plating … (more)
- Is Part Of:
- Clinical biomechanics. Volume 30:Issue 5(2015)
- Journal:
- Clinical biomechanics
- Issue:
- Volume 30:Issue 5(2015)
- Issue Display:
- Volume 30, Issue 5 (2015)
- Year:
- 2015
- Volume:
- 30
- Issue:
- 5
- Issue Sort Value:
- 2015-0030-0005-0000
- Page Start:
- 405
- Page End:
- 410
- Publication Date:
- 2015-06
- Subjects:
- Proximal phalanx -- Lateral plating -- Fracture fixation
Biomechanics -- Periodicals
Osteopathic medicine -- Periodicals
Biomechanics -- Periodicals
Osteopathic Medicine -- Periodicals
612.76 - Journal URLs:
- http://www.sciencedirect.com/science/journal/02680033 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.clinbiomech.2015.03.019 ↗
- Languages:
- English
- ISSNs:
- 0268-0033
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3286.262800
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 830.xml