Validation and reliability of two activity monitors for energy expenditure assessment. Issue 1 (January 2016)
- Record Type:
- Journal Article
- Title:
- Validation and reliability of two activity monitors for energy expenditure assessment. Issue 1 (January 2016)
- Main Title:
- Validation and reliability of two activity monitors for energy expenditure assessment
- Authors:
- Brazeau, Anne-Sophie
Beaudoin, Nadia
Bélisle, Virginie
Messier, Virginie
Karelis, Antony D.
Rabasa-Lhoret, Rémi - Abstract:
- Abstract: Objectives: This study explores the reliability and validity of the SenseWear Armband (SWA) and Actical (ACT) for free-living total energy expenditure, and energy expenditure during rest and light-to-moderate exercises (walking, ergocycling). Design: Participants wore the 2 devices during 7 days (free-living) and then participated to 3 days of testing in our laboratory. Methods: SWA and ACT estimates of total energy expenditure was compared to the doubly labeled water technique (7 days), and energy expenditure during rest (60 min), treadmill (45 min; intensities ∼22% to ∼41% VO2peak ) and ergocycling (45 min; ∼50% VO2peak ) were compared to indirect calorimetry over the following 3 days. Paired T -tests and intra-class correlation coefficient (ICC) with 95% confidence interval (CI95 ) were computed. Results: Twenty adults were recruited (BMI 23.1 ± 2.3 kg/m 2 ). Compared to doubly labelled water, SWA overestimated energy expenditure by 94 kcal/d (±319; P = 0.2) and ACT underestimated by −244 kcal/d (±258; P = 0.001). Energy expenditure during rest (SWA 210 ± 116, ACT 124 ± 133 kcal/d; p < 0.05) and treadmill (according on intensity: SWA 54 ± 46 to 67 ± 38, ACT 68 ± 25 to 84 ± 40 kcal; p < 0.05) were overestimated and underestimated during ergocycling (SWA −93 ± 65, ACT −269 ± 111 kcal; p < 0.05) compared to indirect calorimetry. High ICC were observed at rest (SWA 0.994 CI95 0.987–0.997; ACT 0.998 CI95 0.996–0.999) and during ergocycling (SWA 0.941 CI95Abstract: Objectives: This study explores the reliability and validity of the SenseWear Armband (SWA) and Actical (ACT) for free-living total energy expenditure, and energy expenditure during rest and light-to-moderate exercises (walking, ergocycling). Design: Participants wore the 2 devices during 7 days (free-living) and then participated to 3 days of testing in our laboratory. Methods: SWA and ACT estimates of total energy expenditure was compared to the doubly labeled water technique (7 days), and energy expenditure during rest (60 min), treadmill (45 min; intensities ∼22% to ∼41% VO2peak ) and ergocycling (45 min; ∼50% VO2peak ) were compared to indirect calorimetry over the following 3 days. Paired T -tests and intra-class correlation coefficient (ICC) with 95% confidence interval (CI95 ) were computed. Results: Twenty adults were recruited (BMI 23.1 ± 2.3 kg/m 2 ). Compared to doubly labelled water, SWA overestimated energy expenditure by 94 kcal/d (±319; P = 0.2) and ACT underestimated by −244 kcal/d (±258; P = 0.001). Energy expenditure during rest (SWA 210 ± 116, ACT 124 ± 133 kcal/d; p < 0.05) and treadmill (according on intensity: SWA 54 ± 46 to 67 ± 38, ACT 68 ± 25 to 84 ± 40 kcal; p < 0.05) were overestimated and underestimated during ergocycling (SWA −93 ± 65, ACT −269 ± 111 kcal; p < 0.05) compared to indirect calorimetry. High ICC were observed at rest (SWA 0.994 CI95 0.987–0.997; ACT 0.998 CI95 0.996–0.999) and during ergocycling (SWA 0.941 CI95 0.873–0.975; ACT 0.854 CI95 0.687–0.939). Conclusion: Acceptable estimation of total energy expenditure was observed with the SWA. Both devices were reliable but not accurate for energy expenditure's estimations during rest and for specific exercises. … (more)
- Is Part Of:
- Journal of science and medicine in sport. Volume 19:Issue 1(2016:Jan.)
- Journal:
- Journal of science and medicine in sport
- Issue:
- Volume 19:Issue 1(2016:Jan.)
- Issue Display:
- Volume 19, Issue 1 (2016)
- Year:
- 2016
- Volume:
- 19
- Issue:
- 1
- Issue Sort Value:
- 2016-0019-0001-0000
- Page Start:
- 46
- Page End:
- 50
- Publication Date:
- 2016-01
- Subjects:
- Test–retest -- Activity monitoring -- Accelerometer -- Motion sensors -- Doubly labeled water
Sports sciences -- Periodicals
Sports medicine -- Periodicals
Exercise -- Physiological aspects -- Periodicals
Sports -- physiology -- Periodicals
Sports Medicine -- Periodicals
Sportgeneeskunde
617.102705 - Journal URLs:
- http://www.sciencedirect.com/science/journal/14402440 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.jsams.2014.11.001 ↗
- Languages:
- English
- ISSNs:
- 1440-2440
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5054.840000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 1076.xml